REPEATABILITY OF IOP MEASUREMENT WITH THE NON‐CONTACT TONOMETER*

&NA; Fifty readings were taken with the non‐contact tonometer on 16 men with normal IOP. There was a decrease in IOP on successive measurements of about 1 mm Hg. The median of any three successive readings was always within ±2 mm Hg of the average of all 50 measurements.