3-D Velocity and Volume Flow Measurement $In~Vivo$ Using Speckle Decorrelation and 2-D High-Frame-Rate Contrast-Enhanced Ultrasound

Being able to measure 3-D flow velocity and volumetric flow rate effectively in the cardiovascular system is valuable but remains a significant challenge in both clinical practice and research. Currently, there has not been an effective and practical solution to the measurement of volume flow using ultrasound imaging systems due to challenges in existing 3-D imaging techniques and high system cost. In this study, a new technique for quantifying volumetric flow rate from the cross-sectional imaging plane of the blood vessel was developed by using speckle decorrelation (SDC), 2-D high-frame-rate imaging with a standard 1-D array transducer, microbubble contrast agents, and ultrasound imaging velocimetry (UIV). Through SDC analysis of microbubble signals acquired with a very high frame rate and by using UIV to estimate the two in-plane flow velocity components, the third and out-of-plane velocity component can be obtained over time and integrated to estimate volume flow. The proposed technique was evaluated on a wall-less flow phantom in both steady and pulsatile flow. UIV in the longitudinal direction was conducted as a reference. The influences of frame rate, mechanical index (MI), orientation of imaging plane, and compounding on velocity estimation were also studied. In addition, an in vivo trial on the abdominal aorta of a rabbit was conducted. The results show that the new system can estimate volume flow with an averaged error of 3.65% ± 2.37% at a flow rate of 360 mL/min and a peak velocity of 0.45 m/s, and an error of 5.03% ± 2.73% at a flow rate of 723 mL/min and a peak velocity of 0.8 m/s. The accuracy of the flow velocity and volumetric flow rate estimation directly depend on the imaging frame rate. With a frame rate of 6000 Hz, a velocity up to 0.8 m/s can be correctly estimated. A higher mechanical index (MI = 0.42) is shown to produce greater errors (up to 21.78±0.49%, compared to 3.65±2.37% at MI = 0.19). An in vivo trial, where velocities up to 1 m/s were correctly measured, demonstrated the potential of the technique in clinical applications.

[1]  C L de Korte,et al.  Echo decorrelation from displacement gradients in elasticity and velocity estimation , 1999, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[2]  M. Fink,et al.  Time-Resolved Pulsed Elastography with Ultrafast Ultrasonic Imaging , 1999, Ultrasonic imaging.

[3]  William F Weitzel,et al.  Quantification of Ultrasound Correlation‐Based Flow Velocity Mapping and Edge Velocity Gradient Measurement , 2013, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[4]  B. Sigel,et al.  Red cell aggregation as a cause of blood-flow echogenicity. , 1983, Radiology.

[5]  J. Brian Fowlkes,et al.  Determination of scan-plane motion using speckle decorrelation: Theoretical considerations and initial test , 1997, Int. J. Imaging Syst. Technol..

[6]  Richard W Prager,et al.  Sensorless freehand 3-D ultrasound using regression of the echo intensity. , 2003, Ultrasound in medicine & biology.

[7]  J. Brian Fowlkes,et al.  Three-dimensional flow vectors from rf ultrasound signals , 2002, SPIE Medical Imaging.

[8]  Kristoffer Lindskov Hansen,et al.  Ultrasonic 3-D Vector Flow Method for Quantitative In Vivo Peak Velocity and Flow Rate Estimation , 2017, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[9]  P Vallance,et al.  Measuring forearm blood flow and interpreting the responses to drugs and mediators. , 1995, Hypertension.

[10]  Chee Hau Leow,et al.  Flow Velocity Mapping Using Contrast Enhanced High-Frame-Rate Plane Wave Ultrasound and Image Tracking: Methods and Initial in Vitro and in Vivo Evaluation. , 2015, Ultrasound in medicine & biology.

[11]  Richard W Prager,et al.  Speckle classification for sensorless freehand 3-D ultrasound. , 2005, Ultrasound in medicine & biology.

[12]  Meng-Xing Tang,et al.  Effects of nonlinear propagation in ultrasound contrast agent imaging. , 2010, Ultrasound in medicine & biology.

[13]  D O Cosgrove,et al.  Quantitative contrast-enhanced ultrasound imaging: a review of sources of variability , 2011, Interface Focus.

[14]  Charles Tremblay-Darveau,et al.  Combined perfusion and doppler imaging using plane-wave nonlinear detection and microbubble contrast agents , 2014, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[15]  Richard W Prager,et al.  Sensorless reconstruction of unconstrained freehand 3D ultrasound data. , 2007, Ultrasound in medicine & biology.

[16]  Meng-Xing Tang,et al.  The influence of gas saturation on microbubble stability. , 2012, Ultrasound in medicine & biology.

[17]  C. Tegeler,et al.  Volume Flow Rate of Common Carotid Artery Measured by Doppler Method and Color Velocity Imaging Quantification (CVI‐Q) , 2006, Journal of neuroimaging : official journal of the American Society of Neuroimaging.

[18]  Damien Garcia,et al.  Ultrasound Vector Flow Imaging—Part I: Sequential Systems , 2016, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[19]  Charlie Demené,et al.  Spatiotemporal Clutter Filtering of Ultrafast Ultrasound Data Highly Increases Doppler and fUltrasound Sensitivity , 2015, IEEE Transactions on Medical Imaging.

[20]  S. Anderson,et al.  The aging kidney: physiological changes. , 2010, Advances in chronic kidney disease.

[21]  Chee Hau Leow,et al.  Spatio-Temporal Flow and Wall Shear Stress Mapping Based on Incoherent Ensemble-Correlation of Ultrafast Contrast Enhanced Ultrasound Images. , 2018, Ultrasound in medicine & biology.

[22]  Bastien Denarie,et al.  Coherent Plane Wave Compounding for Very High Frame Rate Ultrasonography of Rapidly Moving Targets , 2013, IEEE Transactions on Medical Imaging.

[23]  Hairong Zheng,et al.  Ultrasonic particle image velocimetry for improved flow gradient imaging: algorithms, methodology and validation , 2010, Physics in medicine and biology.

[24]  Charles Tremblay-Darveau,et al.  Concepts and Tradeoffs in Velocity Estimation With Plane-Wave Contrast-Enhanced Doppler , 2016, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[25]  Damien Garcia,et al.  Ultrasound Vector Flow Imaging: II: Parallel Systems. , 2016, IEEE transactions on ultrasonics, ferroelectrics, and frequency control.

[26]  B. Friemel,et al.  Speckle decorrelation due to two-dimensional flow gradients , 1998, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[27]  M. Lockhart,et al.  Hemodialysis arteriovenous fistula maturity: US evaluation. , 2002, Radiology.

[28]  J M Rubin,et al.  Volume flow measurement using Doppler and grey-scale decorrelation. , 2001, Ultrasound in medicine & biology.

[29]  G. Cloutier,et al.  The effects of hematocrit, shear rate, and turbulence on ultrasonic Doppler spectrum from blood , 1992, IEEE Transactions on Biomedical Engineering.

[30]  J. Jensen,et al.  A new method for estimation of velocity vectors , 1998, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[31]  Matthias Bo Stuart,et al.  3-D Vector Flow Estimation With Row–Column-Addressed Arrays , 2016, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[32]  R. F. Wagner,et al.  Statistical properties of radio-frequency and envelope-detected signals with applications to medical ultrasound. , 1987, Journal of the Optical Society of America. A, Optics and image science.

[33]  Jacinta E Browne,et al.  Development of a vessel-mimicking material for use in anatomically realistic Doppler flow phantoms. , 2011, Ultrasound in medicine & biology.

[34]  Mickael Tanter,et al.  Out-of-plane Doppler imaging based on ultrafast plane wave imaging , 2015, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[35]  P. Carson,et al.  Speckle decorrelation flow measurement with B-mode US of contrast agent flow in a phantom and in rabbit kidney. , 1999, Radiology.

[36]  R. F. Wagner,et al.  Statistics of Speckle in Ultrasound B-Scans , 1983, IEEE Transactions on Sonics and Ultrasonics.

[37]  A F van der Steen,et al.  Decorrelation characteristics of transverse blood flow along an intravascular array catheter: effects of aggregation of red blood cells. , 2001, Ultrasound in medicine & biology.

[38]  Allen R. Tannenbaum,et al.  Localizing Region-Based Active Contours , 2008, IEEE Transactions on Image Processing.

[39]  R. Adrian Particle-Imaging Techniques for Experimental Fluid Mechanics , 1991 .

[40]  Paul A Dayton,et al.  Formulation and acoustic studies of a new phase-shift agent for diagnostic and therapeutic ultrasound. , 2011, Langmuir : the ACS journal of surfaces and colloids.

[41]  P. Carson,et al.  Automated three-dimensional US frame positioning computed from elevational speckle decorrelation. , 1998, Radiology.

[42]  John M. Reid,et al.  Scattering of Ultrasound by Blood , 1976, IEEE Transactions on Biomedical Engineering.

[43]  N Bom,et al.  Blood flow imaging and volume flow quantitation with intravascular ultrasound. , 1998, Ultrasound in medicine & biology.

[44]  Peter R Hoskins,et al.  Simulation and validation of arterial ultrasound imaging and blood flow. , 2008, Ultrasound in medicine & biology.

[45]  Nigel L. Bush,et al.  Freehand Elasticity Imaging Using Speckle Decorrelation Rate , 1996 .

[46]  K. Beach,et al.  Cross-beam vector Doppler ultrasound for angle-independent velocity measurements. , 2000, Ultrasound in medicine & biology.

[47]  M. Fink,et al.  Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastography , 2009, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.