Elastic image registration to quantify 3-D regional myocardial deformation from volumetric ultrasound: experimental validation in an animal model.

Although real-time 3-D echocardiography has the potential to allow more accurate assessment of global and regional ventricular dynamics compared with more traditional 2-D ultrasound examinations, it still requires rigorous testing and validation should it break through as a standard examination in routine clinical practice. However, only a limited number of studies have validated 3-D strain algorithms in an in vivo experimental setting. The aim of the present study, therefore, was to validate a registration-based strain estimation methodology in an animal model. Volumetric images were acquired in 14 open-chest sheep instrumented with ultrasonic microcrystals. Radial strain (ɛRR), longitudinal strain (ɛLL) and circumferential strain (ɛCC) were estimated during different stages: at rest, during reduced and increased cardiac inotropy induced by esmolol and dobutamine infusion, respectively, and during acute ischemia. Agreement between image-based and microcrystal-based strain estimates was evaluated by their linear correlation, indicating that all strain components could be estimated with acceptable accuracy (r = 0.69 for ɛRR, r = 0.64 for ɛLL and r = 0.62 for ɛCC). These findings are comparable to the performance of the current state-of-the-art commercial 3-D speckle tracking methods. Furthermore, shape of the strain curves, timing of peak values and location of dysfunctional regions were identified well. Whether 3-D elastic registration performs better than 3-D block matching-based methodologies still remains to be proven.

[1]  Michael Unser,et al.  Fast parametric elastic image registration , 2003, IEEE Trans. Image Process..

[2]  L. Barbot,et al.  Evaluation of global left ventricular systolic function using three-dimensional echocardiography speckle-tracking strain parameters. , 2012, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[3]  Piet Claus,et al.  On the calculation of principle curvatures of the left-ventricular surfaces , 2008, 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[4]  Shunichi Homma,et al.  Region-based endocardium tracking on real-time three-dimensional ultrasound. , 2009, Ultrasound in medicine & biology.

[5]  J. Bogaert,et al.  Regional nonuniformity of normal adult human left ventricle. , 2001, American journal of physiology. Heart and circulatory physiology.

[6]  Jorge Nocedal,et al.  A Limited Memory Algorithm for Bound Constrained Optimization , 1995, SIAM J. Sci. Comput..

[7]  Brecht Heyde,et al.  Current state of three-dimensional myocardial strain estimation using echocardiography. , 2013, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[8]  Nico de Jong,et al.  Left ventricular border tracking using cardiac motion models and optical flow. , 2011, Ultrasound in medicine & biology.

[9]  P. Wouters,et al.  Three-dimensional myocardial strain estimation from volumetric ultrasound: Experimental validation in an animal model , 2011, 2011 IEEE International Ultrasonics Symposium.

[10]  Frederik Maes,et al.  Elastic Image Registration Versus Speckle Tracking for 2-D Myocardial Motion Estimation: A Direct Comparison In Vivo , 2013, IEEE Transactions on Medical Imaging.

[11]  Charles Davidson,et al.  Correspondence between the 17-segment model and coronary arterial anatomy using contrast-enhanced cardiac magnetic resonance imaging. , 2008, JACC. Cardiovascular imaging.

[12]  Thomas H Marwick,et al.  Will standardization make strain a standard measurement? , 2012, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[13]  Bart Bijnens,et al.  Identification of acutely ischemic myocardium using ultrasonic strain measurements. A clinical study in patients undergoing coronary angioplasty. , 2003, Journal of the American College of Cardiology.

[14]  L. Weinert,et al.  Reproducibility and inter-vendor variability of left ventricular deformation measurements by three-dimensional speckle-tracking echocardiography. , 2011, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[15]  Daniel Rueckert,et al.  Nonrigid registration using free-form deformations: application to breast MR images , 1999, IEEE Transactions on Medical Imaging.

[16]  Paul Suetens,et al.  Three-Dimensional Cardiac Strain Estimation Using Spatio–Temporal Elastic Registration of Ultrasound Images: A Feasibility Study , 2008, IEEE Transactions on Medical Imaging.

[17]  Jan Flusser,et al.  Image registration methods: a survey , 2003, Image Vis. Comput..

[18]  M. Cerqueira,et al.  Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. , 2002, Circulation.

[19]  Richard G P Lopata,et al.  Three-dimensional cardiac strain imaging in healthy children using RF-data. , 2011, Ultrasound in medicine & biology.

[20]  Yang Wang,et al.  Prediction Based Collaborative Trackers (PCT): A Robust and Accurate Approach Toward 3D Medical Object Tracking , 2011, IEEE Transactions on Medical Imaging.

[21]  Alejandro F. Frangi,et al.  Temporal diffeomorphic free-form deformation: Application to motion and strain estimation from 3D echocardiography , 2012, Medical Image Anal..

[22]  Michael Unser,et al.  Spatio-temporal nonrigid registration for ultrasound cardiac motion estimation , 2005, IEEE Transactions on Medical Imaging.

[23]  Piet Claus,et al.  Experimental assessment of a new research tool for the estimation of two-dimensional myocardial strain. , 2006, Ultrasound in medicine & biology.

[24]  Tomoko Ishizu,et al.  Endocardial surface area tracking for assessment of regional LV wall deformation with 3D speckle tracking imaging. , 2011, JACC. Cardiovascular imaging.

[25]  Michael J Ackerman,et al.  Engineering and algorithm design for an image processing Api: a technical report on ITK--the Insight Toolkit. , 2002, Studies in health technology and informatics.

[26]  Hon Fai Choi,et al.  Regional cardiac motion and strain estimation in three-dimensional echocardiography: a validation study in thick-walled univentricular phantoms , 2012, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[27]  Victor Mor-Avi,et al.  Current and evolving echocardiographic techniques for the quantitative evaluation of cardiac mechanics: ASE/EAE consensus statement on methodology and indications endorsed by the Japanese Society of Echocardiography. , 2011, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[28]  F Maffessanti,et al.  Quantitative evaluation of regional left ventricular function using three-dimensional speckle tracking echocardiography , 2009, 2009 36th Annual Computers in Cardiology Conference (CinC).

[29]  Andriy Myronenko,et al.  Maximum Likelihood Motion Estimation in 3D Echocardiography through Non-rigid Registration in Spherical Coordinates , 2009, FIMH.

[30]  Piet Claus,et al.  Investigating Cardiac Function Using Motion and Deformation Analysis in the Setting of Coronary Artery Disease , 2007, Circulation.

[31]  Tomoko Ishizu,et al.  Validation of 3-Dimensional Speckle Tracking Imaging to Quantify Regional Myocardial Deformation , 2009, Circulation. Cardiovascular imaging.

[32]  James S. Duncan,et al.  Estimation of 3D left ventricular deformation from echocardiography , 2001, Medical Image Anal..

[33]  Hervé Delingette,et al.  Computational and Physical Phantom Setups for the Second Cardiac Motion Analysis Challenge (cMAC2) , 2012, STACOM.

[34]  Jan D’hooge,et al.  Comparison of a new methodology for the assessment of 3D myocardial strain from volumetric ultrasound with 2D speckle tracking , 2011, The International Journal of Cardiovascular Imaging.

[35]  Laurent D. Cohen,et al.  Robust image registration based on a Partition of Unity Finite Element Method , 2008, 2008 5th IEEE International Symposium on Biomedical Imaging: From Nano to Macro.

[36]  H. Torp,et al.  3-D speckle tracking for assessment of regional left ventricular function. , 2009, Ultrasound in medicine & biology.

[37]  P. Théroux,et al.  Regional Myocardial function in the conscious dog during acute coronary occlusion and responses to morphine, propranolol, nitroglycerin, and lidocaine. , 1976 .