A positron camera detector design with cross-coupled scintillators and quadrant sharing photomultipliers

A new positron camera scintillation detector design is proposed. This design uses discrete scintillators which are tightly optically coupled (glued) together to form a block. The coupling interfaces between crystals have different degrees of polish depending on the interface location within the block. Each block has the same dimension as that of the photomultipliers (PMTs) used, but the blocks and the PMT array are offset in both X-Y directions. The advantages of the design are: (a) higher detection efficiency with almost 100% packing fraction; (b) very uniform signal pulse-heights even for very small crystals at the edge of the block; (c) decoding which is less dependent on the depth of interaction; (d) significant decrease in the number of PMTs by up to 75%; and (e) with the smallest PMT (1.0 cm) currently available, spatial resolution in the range of 1-2 mm is possible. >

[1]  M. Bergstrom,et al.  Design Studtes of Two Possible Detector Blocks for High Resolution Positron Emission Tomography of the Brain , 1987, IEEE Transactions on Nuclear Science.

[2]  N. Nohara,et al.  A quad BGO detector and its timing and positioning discrimination for positron computed tomography , 1982 .

[3]  R. Nutt,et al.  An improved multicrystal 2-D BGO detector for PET , 1991, Conference Record of the 1991 IEEE Nuclear Science Symposium and Medical Imaging Conference.

[4]  M Dahlbom,et al.  An evaluation of a two-dimensional array detector for high resolution PET. , 1988, IEEE transactions on medical imaging.

[5]  M. Dahlbom,et al.  An Investigation Of Methods Of Pileup Rejection For 2-D Array Detectors Employed In High Resolution Pet , 1990, 1990 IEEE Nuclear Science Symposium Conference Record.

[6]  R. Nutt,et al.  A Multicrystal Two Dimensional BGO Detector System for Positron Emission Tomography , 1986, IEEE Transactions on Nuclear Science.

[7]  L. Feinendegen,et al.  A preliminary evaluation of a positron camera system using weighted decoding of individual crystals , 1988 .

[8]  Edward J. Hoffman,et al.  An investigation of methods of pileup rejection for 2-D array detectors employed in high resolution PET , 1991 .

[9]  Magnus Dahlbom,et al.  Detector, shielding and geometric design factors for a high-resolution PET system , 1990 .

[10]  S. Holte,et al.  Evaluation of the Karolinska new positron camera system; the Scanditronix PC2048-15B , 1990 .

[11]  F. Daghighian,et al.  Quality control in PET systems employing 2-D modular detectors , 1989 .