Initial evaluation of robotic technology for microsurgical vasovasostomy.

PURPOSE Conventional microscope assisted vasovasostomy (MAVV) is a technically difficult procedure that is most successful in the hands of well-trained microsurgeons. Robotics may help surgeons overcome the microsurgical challenges of tremor, limited dexterity, miniaturized instrumentation and use of fine suture. We determine the feasibility of a robotic assisted vasovasostomy (RAVV) and compare performance measures with those of conventional MAVV. MATERIALS AND METHODS One surgeon performed 10 vasovasostomies with a modified 1-layer technique and 9-zero suture on fresh human vas specimens using the robot in 5 RAVV cases and standard microsurgical instrumentation in 5 MAVV cases. Pre-specified performance measures and adverse haptic events (broken sutures, bent needles or loose stitches) were recorded. Patency was evaluated by instilling saline through the anastomoses. RESULTS Mean operating time and number of adverse haptic events were higher for RAVV than for MAVV (84 vs 38 minutes, p = 0.01; 2.4 vs 0.0 events, p = 0.03). The number of needle passes required for the 6 full-thickness stitches was similar in both groups (16.8 vs 15.2 passes, p = 0.55). Although no tremor occurred during RAVV, minimal to moderate amounts occurred during MAVV. Minimal fatigue was noted for both groups. Patency was confirmed in all 10 operations. CONCLUSIONS Use of RAVV in this human ex vivo vas model was feasible. While RAVV took longer to perform and was associated with adverse haptic events, elimination of tremor and comparable patency rates suggest that it may be a viable surgical alternative for microsurgical vasovasostomy.

[1]  Louis R. Kavoussi,et al.  Remote monitoring and assistance during laparoscopy , 2001 .

[2]  A. Berrebi,et al.  Endoscopic coronary artery bypass grafting with the aid of robotic assisted instruments. , 1999, The Journal of thoracic and cardiovascular surgery.

[3]  T. Falcone,et al.  Robotic-assisted laparoscopic microsurgical tubal anastomosis: a human pilot study. , 2000, Fertility and sterility.

[4]  M. Degueldre,et al.  Robotically assisted laparoscopic microsurgical tubal reanastomosis: a feasibility study. , 2000, Fertility and sterility.

[5]  A Darzi,et al.  ‘Urobotics’: robots in urology , 2001, BJU international.

[6]  R. Link,et al.  Indications for pelvic lymphadenectomy in prostate cancer. , 2001, The Urologic clinics of North America.

[7]  R A Buckingham,et al.  Robots in operating theatres , 1995, BMJ.

[8]  Ashutosh Tewari,et al.  Technique of da Vinci robot-assisted anatomic radical prostatectomy. , 2002, Urology.

[9]  I. Sharlip,et al.  Results of 1,469 microsurgical vasectomy reversals by the Vasovasostomy Study Group. , 1991, Journal of urological nursing.

[10]  J C Bowersox,et al.  Microvascular Anastomoses Performed in Rats Using a Microsurgical Telemanipulator , 2000, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[11]  D. Stoianovici,et al.  Robotics in urologic surgery. , 1997, Urology.

[12]  L R Kavoussi,et al.  Telesurgery. Remote monitoring and assistance during laparoscopy. , 2001, The Urologic clinics of North America.

[13]  I J Constable,et al.  Robotic ocular ultramicrosurgery. , 1998, Australian and New Zealand journal of ophthalmology.

[14]  H. Das,et al.  Robot-assisted Microsurgery: A Feasibility Study in the Rat , 2001, Neurosurgery.

[15]  W. Bargar,et al.  Primary and Revision Total Hip Replacement Using the Robodoc® System , 1998, Clinical orthopaedics and related research.

[16]  H H Chen,et al.  Development of a robotic navigation system for neurosurgery. , 1999, Studies in health technology and informatics.

[17]  M. Siemionow,et al.  Robotic Assistance in Microsurgery , 2000, Journal of reconstructive microsurgery.

[18]  Ashutosh Tewari,et al.  Prospective comparison of radical retropubic prostatectomy and robot-assisted anatomic prostatectomy: the Vattikuti Urology Institute experience. , 2002, Urology.