Assessment of Prosthesis Alignment after Revision Total Knee Arthroplasty Using EOS 2D and 3D Imaging: A Reliability Study

Introduction A new low-dose X-ray device, called EOS, has been introduced for determining lower-limb alignment in 2D and 3D. Reliability has not yet been assessed when using EOS on lower limbs containing a knee prosthesis. Therefore purpose of this study was to determine intraobserver and interobserver reliability of EOS 2D and 3D knee prosthesis alignment measurements after revision total knee arthroplasty (rTKA). Methods Forty anteroposterior and lateral images of 37 rTKA patients were included. Two observers independently performed measurements on these images twice. Varus/valgus angles were measured in 2D (VV2D) and 3D (VV3D). Intraclass correlation coefficients and the Bland and Altman method were used to determine reliability. T-tests were used to test potential differences. Results Intraobserver and interobserver reliability were excellent for VV2D and VV3D. No significant difference or bias between the first and second measurements or the two observers was found. A significant mean and absolute difference of respectively 1.00° and 1.61° existed between 2D and 3D measurements. Conclusions EOS provides reliable varus/valgus measurements in 2D and 3D for the alignment of the knee joint with a knee prosthesis. However, significant differences exist between varus/valgus measurements in 2D and 3D.

[1]  P. Stratford Reliability: consistency or differentiating among subjects? , 1989, Physical therapy.

[2]  Peter F. M. Choong,et al.  Does accurate anatomical alignment result in better function and quality of life? Comparing conventional and computer-assisted total knee arthroplasty. , 2009, The Journal of arthroplasty.

[3]  Karen Sloan,et al.  Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. , 2009, The Journal of arthroplasty.

[4]  C. Delin,et al.  Evaluation of a new low-dose biplanar system to assess lower-limb alignment in 3D: a phantom study , 2012, Skeletal Radiology.

[5]  Daisuke Ogino,et al.  Revision total knee arthroplasty: 1990 through 2002. A review of the Finnish arthroplasty registry. , 2006, The Journal of bone and joint surgery. American volume.

[6]  J. Lonner,et al.  Effect of Rotation and Knee Flexion on Radiographic Alignment in Total Knee Arthroplasties , 1996, Clinical orthopaedics and related research.

[7]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[8]  Heleen Beckerman,et al.  Smallest real difference, a link between reproducibility and responsiveness , 2001, Quality of Life Research.

[9]  B. Guenouna,et al.  Reliability of a new method for lower-extremity measurements based on stereoradiographic three-dimensional reconstruction , 2012 .

[10]  M. Tunyogi-Csapó,et al.  Accuracy and reliability of coronal and sagittal spinal curvature data based on patient-specific three-dimensional models created by the EOS 2D/3D imaging system. , 2012, The spine journal : official journal of the North American Spine Society.

[11]  P D Bezemer,et al.  REPRODUCIBILITY AND RESPONSIVENESS OF EVALUATIVE OUTCOME MEASURES , 2001, International Journal of Technology Assessment in Health Care.

[12]  W Skalli,et al.  Reproducibility of Measuring the Shape and Three-Dimensional Position of Cervical Vertebrae in Upright Position Using the EOS Stereoradiography System , 2007, Spine.

[13]  R. Morris,et al.  Coronal alignment after total knee replacement. , 1991, The Journal of bone and joint surgery. British volume.

[14]  J. Lazenneca,et al.  Pelvis and total hip arthroplasty acetabular component orientations in sitting and standing positions: Measurements reproductibility with EOS imaging system versus conventional radiographies , 2011 .

[15]  J. Verhaar,et al.  Pitfalls in Determining Knee Alignment – A Radiographic Cadaver Study , 2010, The journal of knee surgery.

[16]  Jacob Cohen A Coefficient of Agreement for Nominal Scales , 1960 .

[17]  Eric Nectoux,et al.  Three-dimensional measurements of the lower extremity in children and adolescents using a low-dose biplanar X-ray device , 2012, European Radiology.

[18]  Catherine Adamsbaum,et al.  Evaluation of a new low-dose digital X-ray device: first dosimetric and clinical results in children , 1998, Pediatric Radiology.

[19]  K. Swanson,et al.  Does Axial Limb Rotation Affect the Alignment Measurements in Deformed Limbs? , 2000, Clinical orthopaedics and related research.

[20]  Georges Charpak,et al.  Une nouvelle imagerie Ostéo-Articulaire basse dose en position debout : le système EOS , 2005 .

[21]  J. R. Landis,et al.  The measurement of observer agreement for categorical data. , 1977, Biometrics.

[22]  Fabien Billuart,et al.  Shoulder bony landmarks location using the EOS® low-dose stereoradiography system: a reproducibility study , 2009, Surgical and Radiologic Anatomy.

[23]  J. Fleiss The design and analysis of clinical experiments , 1987 .

[24]  H. Hosalkar,et al.  EOS imaging of the human pelvis: reliability, validity, and controlled comparison with radiography. , 2013, The Journal of bone and joint surgery. American volume.

[25]  S. Kurtz,et al.  Reason for Revision Influences Early Patient Outcomes After Aseptic Knee Revision , 2012, Clinical Orthopaedics and Related Research.

[26]  M. Ritter,et al.  Postoperative alignment of total knee replacement. Its effect on survival. , 1994, Clinical orthopaedics and related research.

[27]  Wafa Skalli,et al.  Angle Measurement Reproducibility Using EOSThree-Dimensional Reconstructions in Adolescent Idiopathic Scoliosis Treated by Posterior Instrumentation , 2011, Spine.

[28]  A. Beckett,et al.  AKUFO AND IBARAPA. , 1965, Lancet.

[29]  Meir Liebergall,et al.  Revision total knee arthroplasty. , 2003, The Israel Medical Association journal : IMAJ.

[30]  S. Ostermeier,et al.  Effect of limb rotation on radiographic alignment in total knee arthroplasties , 2010, Archives of Orthopaedic and Trauma Surgery.

[31]  M. Stokes,et al.  Reliability of assessment tools in rehabilitation: an illustration of appropriate statistical analyses , 1998, Clinical rehabilitation.

[32]  G. Sebag,et al.  Reliability of cervical lordosis and global sagittal spinal balance measurements in adolescent idiopathic scoliosis , 2013, European Spine Journal.