Microscopic asperity contact and deformation of ultrahigh molecular weight polyethylene bearing surfaces

Abstract The effect of the roughness and topography of ultrahigh molecular weight polyethylene (UHMWPE) bearing surfaces on the microscopic contact mechanics with a metallic counterface was investigated in the present study. Both simple sinusoidal roughness forms, with a wide range of amplitudes and wavelengths, and real surface topographies, measured before and after wear testing in a simple pin-on-plate machine, were considered in the theoretical analysis. The finite difference method was used to solve the microscopic contact between the rough UHMWPE bearing surface and a smooth hard counterface. The fast Fourier transform (FFT) was used to cope with the large number of mesh points required to represent the surface topography of the UHMWPE bearing surface. It was found that only isolated asperity contacts occurred under physiological loading, and the real contact area was only a small fraction of the nominal contact area. Consequently, the average contact pressure experienced at the articulating surfaces was significantly higher than the nominal contact pressure. Furthermore, it was shown that the majority of asperities on the worn UHMWPE pin were deformed in the elastic region, and consideration of the plastic deformation only resulted in a negligible increase in the predicted asperity contact area. Microscopic asperity contact and deformation mechanisms may play an important role in the understanding of the wear mechanisms of UHMWPE bearing surfaces.

[1]  A. Wang,et al.  Lubrication and wear of ultra-high molecular weight polyethylene in total joint replacements , 1998 .

[2]  A. Edidin,et al.  Deconvolution of surface topology for quantification of initial wear in highly cross-linked acetabular components for THA. , 2002, Journal of biomedical materials research.

[3]  A. Wang,et al.  Effect of contact stress on friction and wear of ultra-high molecular weight polyethylene in total hip replacement , 2001, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[4]  K. Johnson,et al.  The contact of elastic regular wavy surfaces , 1985 .

[5]  Duncan Dowson,et al.  Macroscopic and microscopic wear mechanisms in ultra-high molecular weight polyethylene , 1993 .

[6]  J. Fisher,et al.  Biological reactions to wear debris in total joint replacement , 2000, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[7]  W. Harris,et al.  Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept. , 2001, The Journal of arthroplasty.

[8]  B. Wroblewski,et al.  A comparison of the wear and debris generation of GUR 1120 (compression moulded) and GUR 4150HP (ram extruded) ultra high molecular weight polyethylene. , 1999, Bio-medical materials and engineering.

[9]  J D DesJardins,et al.  Effects of in vitro wear of machined and molded UHMWPE tibial inserts on TKR kinematics. , 2001, Journal of biomedical materials research.

[10]  W. Harris,et al.  Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE). , 1999, Biomaterials.

[11]  D L Bartel,et al.  The effect of conformity and plastic thickness on contact stresses in metal-backed plastic implants. , 1985, Journal of biomechanical engineering.

[12]  John Fisher,et al.  The influence of contact stress on the wear of UHMWPE for total replacement hip prostheses , 1995 .

[13]  T D Brown,et al.  The Frank Stinchfield Award. 3-Dimensional sliding/contact computational simulation of total hip wear. , 1996, Clinical orthopaedics and related research.

[14]  A. Edidin,et al.  Surface morphology and wear mechanisms of four clinically relevant biomaterials after hip simulator testing. , 2000, Journal of biomedical materials research.

[15]  I. Pinder Surface Topography of Retrieved PCA Acetabular Liners: Proposal of a Novel Wear Mechanism , 1998 .

[16]  Hironobu Oonishi,et al.  Micro-wear patterns on UHMWPE tibial inserts in total knee joint simulation. , 2002, Journal of biomedical materials research.

[17]  John J. Callaghan,et al.  Current Concepts Review - Wear in Total Hip and Knee Replacements* , 1999 .

[18]  Q. Wang,et al.  A versatile method of discrete convolution and FFT (DC-FFT) for contact analyses , 2000 .

[19]  T. Boland,et al.  Nanotribology of CoCr–UHMWPE TJR prosthesis using atomic force microscopy , 2002 .

[20]  I. Clarke,et al.  Microwear phenomena of ultrahigh molecular weight polyethylene cups and debris morphology related to gamma radiation dose in simulator study. , 2001, Journal of biomedical materials research.