Three dimensional shape optimization of total knee replacements for reduced wear

This paper presents the design optimization of a total knee replacement (TKR) using a parametric three-dimensional finite element (FE) model, considering wear of the ultra high molecular weight polyethylene (UHMWPE) insert. A framework has been developed to generate three-dimensional FE models of the femoral component and UHMWPE insert of a TKR design in a batch mode process, then simulate an ISO standard TKR wear test. A modified version of Archard’s wear model calculates abrasive wear as a function of contact pressure, sliding distance and an experimentally determined wear factor. The UHMWPE wear was reduced by modifying the contact geometry of both components in the frontal and sagittal planes. Wear was reduced by 18.5%, from 55.248 to 45.013 mm3 per year by reducing the radii of curvature of the femoral condyles in the sagittal planes, increasing the radii in the frontal plane, and reducing conformity between the implant components.

[1]  M. A. R. Freeman,et al.  The production and biology of polyethylene wear debris , 1978, Archives of orthopaedic and traumatic surgery.

[2]  D. Dowson,et al.  The effect of sliding velocity on the friction and wear of UHMWPE for use in total artificial joints , 1994 .

[3]  T D Brown,et al.  The Frank Stinchfield Award. 3-Dimensional sliding/contact computational simulation of total hip wear. , 1996, Clinical orthopaedics and related research.

[4]  Benjamin J Fregly,et al.  Computational wear prediction of a total knee replacement from in vivo kinematics. , 2005, Journal of biomechanics.

[5]  K. Johnson Contact Mechanics: Frontmatter , 1985 .

[6]  Javad Dargahi,et al.  Optimization of the geometry of total knee implant in the sagittal plane using FEA. , 2003, Bio-medical materials and engineering.

[7]  Thomas D. Brown,et al.  Adaptiive finite element modeling of long‐term polyethylene wear in total hip arthroplasty , 1996 .

[8]  A Amirfazli,et al.  A parametric analysis of fixation post shape in tibial knee prostheses. , 2005, Medical engineering & physics.

[9]  C. J. Bell,et al.  The influence of design, materials and kinematics on the in vitro wear of total knee replacements. , 2005, Journal of biomechanics.

[10]  Liam Blunt,et al.  A metrology solution for the orthopaedic industry , 2005 .

[11]  Kurt M. Marshek,et al.  Discretization Pressure-Wear Theory for Bodies in Sliding Contact , 1989 .

[12]  J. Bryant,et al.  A quantitative technique for reporting surface degradation patterns of UHMWPE components of retrieved total knee replacements. , 1995, Journal of applied biomaterials : an official journal of the Society for Biomaterials.

[13]  Jason P. Halloran,et al.  Explicit finite element modeling of total knee replacement mechanics. , 2005, Journal of biomechanics.

[14]  Robert W. Klein,et al.  The effect of sagittal conformity on knee wear , 2003 .

[15]  J. Archard,et al.  The wear of metals under unlubricated conditions , 1956, Proceedings of the Royal Society of London. Series A. Mathematical and Physical Sciences.

[16]  P. Walker,et al.  Computer model to predict subsurface damage in tibial inserts of total knees , 1998, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[17]  Mark Taylor,et al.  Comparison of long-term numerical and experimental total knee replacement wear during simulated gait loading. , 2007, Journal of biomechanics.

[18]  Benjamin J Fregly,et al.  Predicting knee replacement damage in a simulator machine using a computational model with a consistent wear factor. , 2008, Journal of biomechanical engineering.

[19]  P. Sharkey,et al.  Why Are Total Knee Arthroplasties Failing Today? , 2002 .

[20]  Benjamin J Fregly,et al.  Experimental evaluation of an elastic foundation model to predict contact pressures in knee replacements. , 2003, Journal of biomechanics.

[21]  T. Brown,et al.  A sliding-distance-coupled finite element formulation for polyethylene wear in total hip arthroplasty. , 1996, Journal of biomechanics.

[22]  M Beaugonin,et al.  Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis. , 2002, Journal of biomechanics.

[23]  V. L. Giddings,et al.  Total Knee Replacement Polyethylene Stresses During Loading in a Knee Simulator , 2001 .

[24]  T D Brown,et al.  Finite element analysis of acetabular wear. Validation, and backing and fixation effects. , 1997, Clinical orthopaedics and related research.

[25]  P S Walker,et al.  Optimization of the bearing surface geometry of total knees. , 1994, Journal of biomechanics.