The laws of electrical stimulation of cardiac tissue

Classic description of the criteria for cardiac stimulation is by the strength-duration curve which may be presented for current, energy, or charge. Minima for each of these occur at different waveform durations. The fundamental mechanism by which the stimulus is effective is altering the polarization potential of the cell membranes in the heart. Partial depolarization of these cells initiates a cascade leading to complete depolarization which in turn triggers cardiac cell contraction. Active propagation of the depolarization from cell to cell is the mechanism for activation of the entire heart when stimulation is used for pacemaking. For depolarization with much stronger shocks, this depolarization mechanism in some cells is combined with hyperpolarization in other cells. The hyperpolarization causes anodal block of propagation. Two methods are used to study cardiac stimulation, the threshold concept and the probability concept, and each method has advantages and limitations. Many variables affect the capability of an electrical shock to stimulate the heart to a desired response. These include the current waveform the characteristics of the electrodes used to inject the current into the tissue, the disorder being treated, the mass of cells required to be effected, metabolic variables (such as temperature or pH of the the tissue), drugs, and the geometric arrangements of electrodes and body organs. Of particular engineering interest are different effectiveness of different electrical waveforms, and the electrode characteristics of half-cell potentials, polarization potentials and materials. Equivalent circuits are presented for the complex and nonlinear, interactive behavior of the electrical shock, electrodes, and tissues.

[1]  B PELESKA,et al.  Cardiac Arrhythmias Following Condenser Discharges and Their Dependence Upon Strength of Current and Phases of Cardiac Cycle , 1963, Circulation research.

[2]  L. Geddes,et al.  Comparative efficacy of damped sine wave and square wave current for transchest ventricular defibrillation in animals. , 1978, Medical instrumentation.

[3]  H. Fricke,et al.  XXXIII. The theory of electrolytic polarization , 1932 .

[4]  L A Geddes,et al.  Strength-duration curves for trapezoidal waveforms of various tilts for transchest defibrillation in animals. , 1978, Medical instrumentation.

[5]  L. Geddes,et al.  Electrical dose for ventricular defibrillation of large and small animals using precordial electrodes. , 1974, The Journal of clinical investigation.

[6]  Gurvich Nl,et al.  Restoration of heart rhythm during fibrillation by a condenser discharge. , 1947 .

[7]  Schoenberg Aa,et al.  Development of standard test methods for evaluating defibrillation recovery characteristics of disposable ECG electrodes. , 1979 .

[8]  L. Geddes,et al.  Energy dose for ventricular defibrillation of children. , 1976, Pediatrics.

[9]  Robert Kennedy L. Chemical analysis of an uncommon species of zeolite , 1803 .

[10]  E. Warburg,et al.  Ueber die Polarisationscapacität des Platins , 1901 .

[11]  John G. Webster,et al.  Defibrillation Recovery Curves for Different Electrode Materials , 1980, IEEE Transactions on Biomedical Engineering.

[12]  B. Lown,et al.  New method for terminating cardiac arrhythmias. Use of synchronized capacitor discharge. , 1986, JAMA.

[13]  P. Zoll,et al.  External noninvasive electric stimulation of the heart. , 1981, Critical care medicine.

[14]  J. Schuder,et al.  Contour graph for relating per cent success in achieving ventricular defibrillation to duration, current, and energy content of shock. , 1979, American heart journal.

[15]  L. Geddes,et al.  The electrical dose for ventricular defibrillation with electrodes applied directly to the heart , 1974 .

[16]  E. Warburg,et al.  Ueber das Verhalten sogenannter unpolarisirbarer Elektroden gegen Wechselstrom , 1899 .

[17]  D P Zipes,et al.  Termination of ventricular fibrillation in dogs by depolarizing a critical amount of myocardium. , 1975, The American journal of cardiology.

[18]  G. I. Thomas,et al.  DC pulse defibrillation. , 1966, The Journal of thoracic and cardiovascular surgery.

[19]  John C. Schuder,et al.  Transthoracic Ventricular Defibrillation with Triangular and Trapezoidal Waveforms , 1966 .

[20]  L. Geddes,et al.  The Strength-Duration Curve , 1985, IEEE Transactions on Biomedical Engineering.