Adaptiive finite element modeling of long‐term polyethylene wear in total hip arthroplasty

Adaptive remeshing capability was addded to an existing sliding‐distance‐coupled finite element model of polyethylene wear in total hip arthroplasty. This augmentation allowed earlier postoperative wear simulation to be extended to the clinically more significant long‐term regimen (as long as 20 years). Loads and femoral head excursions were taken from a physically validated gait analysis model of a patient with an instrumented total hip replacement. For otherwise identical 22, 28, and 32 mm components, the least volumetric wear but the most linear wear occurred for the 22 mm head. When the polyethylene thickness in a 22 mm component was reduced to the same as that in a 32 mm component, the volumetric wear rate for the 22 mm component was still much less than that for the larger component, indicating that sliding distance (head size). rather than polyethylene liner thickness, was primarily responsible for the difference in rates, A “28 mm” series, for which head sizes were varied across the range of currently accepted industrial tolerances, showed that although initial wear rates were greatest for the least congruent articulations, the long‐term volumetric wear was nearly the same, regardless of initial clearance.

[1]  Duncan Dowson,et al.  Laboratory wear tests and clinical observations of the penetration of femoral heads into acetabular cups in total replacement hip joints: II: A microscopical study of the surfaces of Charnley polyethylene acetabular sockets , 1985 .

[2]  B. Morrey,et al.  Effect of femoral head size on wear of the polyethylene acetabular component. , 1990, The Journal of bone and joint surgery. American volume.

[3]  J. C. Simo,et al.  Numerical instabilities in bone remodeling simulations: the advantages of a node-based finite element approach. , 1995, Journal of biomechanics.

[4]  Kurt M. Marshek,et al.  Discretization Pressure-Wear Theory for Bodies in Sliding Contact , 1989 .

[5]  R C Johnston,et al.  The outcome of Charnley total hip arthroplasty with cement after a minimum twenty-year follow-up. The results of one surgeon. , 1993, The Journal of bone and joint surgery. American volume.

[6]  J. Archard Contact and Rubbing of Flat Surfaces , 1953 .

[7]  John Charnley,et al.  Low Friction Arthroplasty of the Hip , 1979 .

[8]  D Dowson,et al.  Tribology of Total Artificial Joints , 1991, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[9]  B. K. Vaughn,et al.  Aseptic loosening in total hip arthroplasty secondary to osteolysis induced by wear debris from titanium-alloy modular femoral heads. , 1989, The Journal of bone and joint surgery. American volume.

[10]  J. Galante,et al.  Contact pressure dependence of wear rates of ultra high molecular weight polyethylene. , 1979, Journal of biomedical materials research.

[11]  R. Brand,et al.  Comparison of hip force calculations and measurements in the same patient. , 1994, The Journal of arthroplasty.

[12]  T. Brown,et al.  A sliding-distance-coupled finite element formulation for polyethylene wear in total hip arthroplasty. , 1996, Journal of biomechanics.

[13]  J. Kabo,et al.  In vivo wear of polyethylene acetabular components. , 1993, The Journal of bone and joint surgery. British volume.

[14]  John Charnley,et al.  Low Friction Arthroplasty of the Hip: Theory and Practice , 1978 .