A method for fast multislice T1 measurement: Feasibility studies on phantoms, young children, and children with canavan's disease

We have developed a multislice protocol for quantitative T1 measurements in which the processing time and the acquisition time are under 2 minutes each for a complete brain study of 15 slices. An echoplanar, inversion‐recovery image sequence is designed to collect data suitable for analysis using a linear regression algorithm. The precision is approximately twice the noise to signal ratio of the images. The accuracy of the protocol is better than 1% for T1 in the range 0–2 seconds and deviates slightly for longer T1 values. The protocol is insensitive to B1 field values. If needed, the data can be postprocessed using a slow, nonlinear algorithm to give an accuracy of less than 1% and a precision of approximately the noise to signal ratio throughout a range of T1 values from 0 to 4 seconds. J. Magn. Reson. Imaging 2000;11:360–367. © 2000 Wiley‐Liss, Inc.

[1]  P. V. van Dijk,et al.  Restoration of Signal Polarity in a Set of Inversion Recovery NMR Images , 1984, IEEE Transactions on Medical Imaging.

[2]  A. Bax,et al.  Tip-angle-reduced T1 imaging , 1986 .

[3]  A. Crawley,et al.  A comparison of one‐shot and recovery methods in T1 imaging , 1988, Magnetic resonance in medicine.

[4]  J. E. Glynn,et al.  Numerical Recipes: The Art of Scientific Computing , 1989 .

[5]  P. Mansfield,et al.  High‐speed multislice T1 mapping using inversion‐recovery echo‐planar imaging , 1990, Magnetic resonance in medicine.

[6]  R. Henkelman,et al.  Practical Implementation and Optimization of One‐shot T1 imaging , 1991, Magnetic resonance in medicine.

[7]  P. Gowland,et al.  Accurate measurement of T1 in vivo in less than 3 seconds using echo‐planar imaging , 1993, Magnetic resonance in medicine.

[8]  S G Kim,et al.  Accurate T1 determination from inversion recovery images: Application to human brain at 4 Tesla , 1994, Magnetic resonance in medicine.

[9]  F S Prato,et al.  A novel fast T1‐mapping method , 1994, Journal of magnetic resonance imaging : JMRI.

[10]  R. Steen,et al.  Precise and accurate measurement of proton T1 in human brain in vivo: Validation and preliminary clinical application , 1994, Journal of magnetic resonance imaging : JMRI.

[11]  W. Reddick,et al.  Statistical error mapping for reliable quantitative T1 imaging , 1996, Journal of magnetic resonance imaging : JMRI.

[12]  W E Reddick,et al.  Age-related changes in the pediatric brain: quantitative MR evidence of maturational changes during adolescence. , 1997, AJNR. American journal of neuroradiology.

[13]  W E Reddick,et al.  Correction of errors caused by imperfect inversion pulses in MR imaging measurement of T1 relaxation times. , 1998, Magnetic resonance imaging.

[14]  P Mansfield,et al.  Optimization of the ultrafast Look-Locker echo-planar imaging T1 mapping sequence. , 1998, Magnetic resonance imaging.

[15]  Rudolf Stollberger,et al.  Fast multislice T1 imaging using a PACE technique. , 1999 .

[16]  P. Kingsley,et al.  Methods of measuring spin-lattice ( T 1 ) relaxation times: an annotated bibliography , 1999 .

[17]  P. Kingsley,et al.  Signal intensities and T 1 calculations in multiple-echo sequences with imperfect pulses , 1999 .