A medical device-grade T1 and ECV phantom for global T1 mapping quality assurance—the T1 Mapping and ECV Standardization in cardiovascular magnetic resonance (T1MES) program

BackgroundT1 mapping and extracellular volume (ECV) have the potential to guide patient care and serve as surrogate end-points in clinical trials, but measurements differ between cardiovascular magnetic resonance (CMR) scanners and pulse sequences. To help deliver T1 mapping to global clinical care, we developed a phantom-based quality assurance (QA) system for verification of measurement stability over time at individual sites, with further aims of generalization of results across sites, vendor systems, software versions and imaging sequences. We thus created T1MES: The T1 Mapping and ECV Standardization Program.MethodsA design collaboration consisting of a specialist MRI small-medium enterprise, clinicians, physicists and national metrology institutes was formed. A phantom was designed covering clinically relevant ranges of T1 and T2 in blood and myocardium, pre and post-contrast, for 1.5 T and 3 T. Reproducible mass manufacture was established. The device received regulatory clearance by the Food and Drug Administration (FDA) and Conformité Européene (CE) marking.ResultsThe T1MES phantom is an agarose gel-based phantom using nickel chloride as the paramagnetic relaxation modifier. It was reproducibly specified and mass-produced with a rigorously repeatable process. Each phantom contains nine differently-doped agarose gel tubes embedded in a gel/beads matrix. Phantoms were free of air bubbles and susceptibility artifacts at both field strengths and T1 maps were free from off-resonance artifacts. The incorporation of high-density polyethylene beads in the main gel fill was effective at flattening the B1 field. T1 and T2 values measured in T1MES showed coefficients of variation of 1 % or less between repeat scans indicating good short-term reproducibility. Temperature dependency experiments confirmed that over the range 15–30 °C the short-T1 tubes were more stable with temperature than the long-T1 tubes. A batch of 69 phantoms was mass-produced with random sampling of ten of these showing coefficients of variations for T1 of 0.64 ± 0.45 % and 0.49 ± 0.34 % at 1.5 T and 3 T respectively.ConclusionThe T1MES program has developed a T1 mapping phantom to CE/FDA manufacturing standards. An initial 69 phantoms with a multi-vendor user manual are now being scanned fortnightly in centers worldwide. Future results will explore T1 mapping sequences, platform performance, stability and the potential for standardization.

[1]  Diego Hernando,et al.  Robust water/fat separation in the presence of large field inhomogeneities using a graph cut algorithm , 2009, Magnetic resonance in medicine.

[2]  David A Bluemke,et al.  Modified Look‐Locker T1 evaluation using Bloch simulations: Human and phantom validation , 2013, Magnetic resonance in medicine.

[3]  D. Pennell,et al.  Longitudinal stability of gel T1 MRI Phantoms for quality assurance of T1 mapping , 2015, Journal of Cardiovascular Magnetic Resonance.

[4]  D. Bluemke,et al.  Modified look-locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scanners , 2013, Journal of Cardiovascular Magnetic Resonance.

[5]  P. Kellman,et al.  T1 and extracellular volume mapping in the heart: estimation of error maps and the influence of noise on precision , 2013, Journal of Cardiovascular Magnetic Resonance.

[6]  S. H. Koenig,et al.  Relaxation of solvent protons by paramagnetic ions and its dependence on magnetic field and chemical environment: implications for NMR imaging. , 1985, Magnetic resonance in medicine.

[7]  K. Briley-Saebo,et al.  Phantom standards with temperature- and field-independent relaxation rates for magnetic resonance imaging. , 1998, Investigative radiology.

[8]  Paul S. Tofts,et al.  Standing Waves in Uniform Water Phantoms , 1994 .

[9]  P. W. Kuchel,et al.  Magnetic susceptibility: solutions, emulsions, and cells , 2003, q-bio/0601030.

[10]  Stefan Neubauer,et al.  T1 measurements in the human myocardium: The effects of magnetization transfer on the SASHA and MOLLI sequences , 2013, Magnetic resonance in medicine.

[11]  Peter Kellman,et al.  Influence of Off-resonance in myocardial T1-mapping using SSFP based MOLLI method , 2013, Journal of Cardiovascular Magnetic Resonance.

[12]  H. Carr,et al.  DIFFUSION AND NUCLEAR SPIN RELAXATION IN WATER , 1958 .

[13]  M. Robson,et al.  Myocardial T1 mapping and extracellular volume quantification: a Society for Cardiovascular Magnetic Resonance (SCMR) and CMR Working Group of the European Society of Cardiology consensus statement , 2013, Journal of Cardiovascular Magnetic Resonance.

[14]  Paul S. Tofts,et al.  QA: Quality Assurance, Accuracy, Precision and Phantoms , 2004 .

[15]  S. Sjöberg,et al.  Nickel-Doped Agarose Gel Phantoms in MR Imaging , 1991 .

[16]  D. Freed Dependence on chain length of NMR relaxation times in mixtures of alkanes. , 2007, The Journal of chemical physics.

[17]  M Tsunoda,et al.  Development of a phantom compatible for MRI and hyperthermia using carrageenan gel—relationship between T1 and T2 values and NaCl concentration , 2004, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[18]  P. Fatouros,et al.  An MRI phantom material for quantitative relaxometry , 1987, Magnetic resonance in medicine.

[19]  Andreas Greiser,et al.  Optimization and validation of a fully‐integrated pulse sequence for modified look‐locker inversion‐recovery (MOLLI) T1 mapping of the heart , 2007, Journal of magnetic resonance imaging : JMRI.

[20]  L. Axel,et al.  Agarose as a tissue equivalent phantom material for NMR imaging. , 1986, Magnetic resonance imaging.

[21]  Andrew G. Webb,et al.  Optimizing the point spread function in phase‐encoded magnetic resonance microscopy , 2004 .

[22]  P. Kellman,et al.  T1-mapping in the heart: accuracy and precision , 2014, Journal of Cardiovascular Magnetic Resonance.

[23]  P S Tofts,et al.  A low dielectric constant customized phantom design to measure RF coil nonuniformity. , 1997, Magnetic resonance imaging.