CHARACTERIZATION OF A MULTILEAF COLLIMATOR SYSTEM

Abstract Commissioning measurements for a multileaf collimator installed on a dual energy accelerator with 6 and 15 MV photons are described. Detailed dosimetric characterization of the multileaf collimator is a requirement for modeling the collimator with treatment planning software. Measurements include a determination of the penumbra width, leaf transmission, between-leaf leakage, and localization of the leaf ends and sides. Standard radiographic film was used for the penumbra measurements, and separate experiments using radiochromic film and thermoluminescent dosimeters were performed to verify that distortions of the dose distribution at an edge due to changing energy sensitivity of silver bromide film are negligible. Films were analyzed with a scanning laser densitometer with a 210 micron spot. Little change in the penumbra edge distribution was noted for different positions of a leaf in the field. Experiments localizing the physical end of the leaves showed less than 1 mm deviation from the 50% decrement line. This small difference is attributed to the shaped end on the leaves. One side of a single leaf corresponded to the 50% decrement line, but the opposite face was aligned with a lower value. This difference is due to the tongue and groove used to decrease between-leaf leakage. For both energies, approximately 2% of photons incident on the multileaf collimator are transmitted and an additional 0.5% leakage occurs between the leaves. Alignment of the leaves to form a straight edge results in a penumbra profile which compares favorably with the standard technique of using alloy blocks. When the edge is stepped, the isodose lines follow the leaf pattern and the boundary is poorly defined compared to divergent blocks.

[1]  P. K. Kijewski,et al.  Computer-controlled radiation therapy. , 1977, Radiology.

[2]  J J Weinkam,et al.  Automatic variation of field size and dose rate in rotation therapy. , 1977, International journal of radiation oncology, biology, physics.

[3]  P K Kijewski,et al.  Dose optimization with computer-controlled gantry rotation, collimator motion and dose-rate variation. , 1983, International journal of radiation oncology, biology, physics.

[4]  Penumbral measurements in water for high-energy x rays. , 1986, Medical physics.

[5]  M Goitein,et al.  The influence of the size of the grid used for dose calculation on the accuracy of dose estimation. , 1989, Medical physics.

[6]  J. Sayeg,et al.  A new method for characterizing beta-ray ophthalmic applicator sources. , 1991, Medical physics.

[7]  D. J. Dawson,et al.  Analysis of physical parameters associated with the measurement of high-energy x-ray penumbra. , 1984, Medical physics.

[8]  R. Nath,et al.  Photon energy dependence of the sensitivity of radiochromic film and comparison with silver halide film and LiF TLDs used for brachytherapy dosimetry. , 1991, Medical physics.

[9]  G K Svensson,et al.  A computer-controlled radiation therapy machine for pelvic and para-aortic nodal areas. , 1981, International journal of radiation oncology, biology, physics.

[10]  A Brahme,et al.  Shaping of arbitrary dose distributions by dynamic multileaf collimation. , 1988, Physics in medicine and biology.

[11]  A R Smith,et al.  Evaluation of multileaf collimator design for a photon beam. , 1992, International journal of radiation oncology, biology, physics.

[12]  C. Sibata,et al.  Influence of detector size in photon beam profile measurements. , 1991, Physics in medicine and biology.

[13]  G E Hanks,et al.  Conformal static field therapy for low volume low grade prostate cancer with rigid immobilization. , 1991, International journal of radiation oncology, biology, physics.

[14]  R Mohan,et al.  Improved dose distributions for 3D conformal boost treatments in carcinoma of the nasopharynx. , 1991, International journal of radiation oncology, biology, physics.

[15]  A. Boyer,et al.  Potential and limitations of invariant kernel conformal therapy. , 1991, Medical physics.