Design of a pulsatile flow facility to evaluate thrombogenic potential of implantable cardiac devices.

Due to expensive nature of clinical trials, implantable cardiac devices should first be extensively characterized in vitro. Prosthetic heart valves (PHVs), an important class of these devices, have been shown to be associated with thromboembolic complications. Although various in vitro systems have been designed to quantify blood-cell damage and platelet activation caused by nonphysiological hemodynamic shear stresses in these PHVs, very few systems attempt to characterize both blood damage and fluid dynamics aspects of PHVs in the same test system. Various numerical modeling methodologies are also evolving to simulate the structural mechanics, fluid mechanics, and blood damage aspects of these devices. This article presents a completely hemocompatible small-volume test-platform that can be used for thrombogenicity studies and experimental fluid mechanics characterization. Using a programmable piston pump to drive freshly drawn human blood inside a cylindrical column, the presented system can simulate various physiological and pathophysiological conditions in testing PHVs. The system includes a modular device-mounting chamber, and in this presented case, a 23 mm St. Jude Medical (SJM) Regents® mechanical heart valve (MHV) in aortic position was used as the test device. The system was validated for its capability to quantify blood damage by measuring blood damage induced by the tester itself (using freshly drawn whole human blood). Blood damage levels were ascertained through clinically relevant assays on human blood while fluid dynamics were characterized using time-resolved particle image velocimetry (PIV) using a blood-mimicking fluid. Blood damage induced by the tester itself, assessed through Thrombin-anti-Thrombin (TAT), Prothrombin factor 1.2 (PF1.2), and hemolysis (Drabkins assay), was within clinically accepted levels. The hydrodynamic performance of the tester showed consistent, repeatable physiological pressure and flow conditions. In addition, the system contains proximity sensors to accurately capture leaflet motion during the entire cardiac cycle. The PIV results showed skewing of the leakage jet, caused by the asymmetric closing of the two leaflets. All these results are critical to characterizing the blood damage and fluid dynamics characteristics of the SJM Regents® MHV, proving the utility of this tester as a precise system for assessing the hemodynamics and thrombogenicity for various PHVs.

[1]  C. Ball,et al.  Effects of low temperature on shear-induced platelet aggregation and activation. , 2004, The Journal of trauma.

[2]  Danny Bluestein,et al.  Thrombogenic Potential of Innovia Polymer Valves versus Carpentier-Edwards Perimount Magna Aortic Bioprosthetic Valves , 2011, ASAIO journal.

[3]  S. Jackson,et al.  Intercellular calcium communication regulates platelet aggregation and thrombus growth , 2003, The Journal of cell biology.

[4]  Ajit P Yoganathan,et al.  Flow and thrombosis at orifices simulating mechanical heart valve leakage regions. , 2006, Journal of biomechanical engineering.

[5]  C. Akins,et al.  Results with mechanical cardiac valvular prostheses. , 1995, The Annals of thoracic surgery.

[6]  E. Bodnar The Medtronic Parallel valve and the lessons learned. , 1996, The Journal of heart valve disease.

[7]  Hee-Sun Kim,et al.  Structural simulations of prosthetic tri-leaflet aortic heart valves. , 2008, Journal of biomechanics.

[8]  Danny Bluestein,et al.  Flow-induced platelet activation and damage accumulation in a mechanical heart valve: numerical studies. , 2007, Artificial organs.

[9]  H Reul,et al.  In Vitro Thrombogenicity Testing of Artificial Organs , 1998, The International journal of artificial organs.

[10]  Ajit P Yoganathan,et al.  Fluid mechanics of heart valves. , 2004, Annual review of biomedical engineering.

[11]  M. Gharib,et al.  Vortex shedding as a mechanism for free emboli formation in mechanical heart valves. , 2000, Journal of biomechanical engineering.

[12]  Ulrich Steinseifer,et al.  In vitro evaluation of a novel hemodynamically optimized trileaflet polymeric prosthetic heart valve. , 2013, Journal of biomechanical engineering.

[13]  Wei Yin,et al.  Flow induced platelet activation in mechanical heart valves - in vitro studies , 2002, Proceedings of the Second Joint 24th Annual Conference and the Annual Fall Meeting of the Biomedical Engineering Society] [Engineering in Medicine and Biology.

[14]  Z. Ruggeri Mechanisms Initiating Platelet Thrombus Formation , 1997, Thrombosis and Haemostasis.

[15]  Neelakantan Saikrishnan,et al.  Micro Particle Image Velocimetry Measurements of Steady Diastolic Leakage Flow in the Hinge of a St. Jude Medical® Regent™ Mechanical Heart Valve , 2013, Annals of Biomedical Engineering.

[16]  D R Hose,et al.  The use of enzyme activated milk for in vitro simulation of prosthetic valve thrombosis. , 1996, The Journal of heart valve disease.

[17]  Fotis Sotiropoulos,et al.  Characterization of Hemodynamic Forces Induced by Mechanical Heart Valves: Reynolds vs. Viscous Stresses , 2008, Annals of Biomedical Engineering.

[18]  J. Svennevig,et al.  Solid Cerebral Microemboli and Cerebrovascular Symptoms in Patients With Prosthetic Heart Valves , 2008, Stroke.

[19]  Brian Savage,et al.  Initiation of Platelet Adhesion by Arrest onto Fibrinogen or Translocation on von Willebrand Factor , 1996, Cell.

[20]  Hélène A. Simon,et al.  FLUID MECHANICS OF ARTIFICIAL HEART VALVES , 2009, Clinical and experimental pharmacology & physiology.

[21]  Danny Bluestein,et al.  Flow-induced platelet activation in a St. Jude mechanical heart valve, a trileaflet polymeric heart valve, and a St. Jude tissue valve. , 2005, Artificial organs.

[22]  A. Yoganathan,et al.  Reduction of procoagulant potential of b-datum leakage jet flow in bileaflet mechanical heart valves via application of vortex generator arrays. , 2010, Journal of biomechanical engineering.

[23]  Steven Deutsch,et al.  Regurgitant flow field characteristics of the St. Jude bileaflet mechanical heart valve under physiologic pulsatile flow using particle image velocimetry. , 2003, Artificial organs.

[24]  Fotis Sotiropoulos,et al.  Flow in a mechanical bileaflet heart valve at laminar and near-peak systole flow rates: CFD simulations and experiments. , 2005, Journal of biomechanical engineering.

[25]  F. R. Rosendaal,et al.  Thromboembolic and Bleeding Complications in Patients With Mechanical Heart Valve Prostheses , 1994, Circulation.

[26]  Steven Deutsch,et al.  The Use of Fluid Mechanics to Predict Regions of Microscopic Thrombus Formation in Pulsatile VADs , 2014, Cardiovascular engineering and technology.

[27]  Lakshmi Prasad Dasi,et al.  Procoagulant Properties of Flow Fields in Stenotic and Expansive Orifices , 2007, Annals of Biomedical Engineering.

[28]  Mitchell W Krucoff,et al.  Clinical evaluation of cardiovascular devices: principles, problems, and proposals for European regulatory reform. Report of a policy conference of the European Society of Cardiology. , 2011, European heart journal.

[29]  H. Shankaran,et al.  Aspects of hydrodynamic shear regulating shear-induced platelet activation and self-association of von Willebrand factor in suspension. , 2003, Blood.

[30]  M. Roussel,et al.  Microemboli in cerebral circulation and alteration of cognitive abilities in patients with mechanical prosthetic heart valves. , 1998, Stroke.

[31]  E. Nauman,et al.  An in vitro model for preclinical testing of thrombogenicity of resorbable metallic stents. , 2015, Journal of biomedical materials research. Part A.

[32]  W. Dembitsky REMATCH and Beyond: The Cost of Treating Heart Failure Using an Implantable Left Ventricular Assist Device , 2006, Seminars in cardiothoracic and vascular anesthesia.

[33]  S. Felix,et al.  New stent surface materials: the impact of polymer-dependent interactions of human endothelial cells, smooth muscle cells, and platelets. , 2014, Acta biomaterialia.

[34]  R. Mason,et al.  Some species differences in fibrinolysis and blood coagulation. , 1971, Journal of biomedical materials research.

[35]  A P Yoganathan,et al.  The sensitivity of indicators of thrombosis initiation to a bileaflet prosthesis leakage stimulus. , 2001, The Journal of heart valve disease.

[36]  U. Steinseifer,et al.  In vitro comparison of bileaflet aortic heart valve prostheses. St. Jude Medical, CarboMedics, modified Edwards-Duromedics, and Sorin-Bicarbon valves. , 1993, The Journal of thoracic and cardiovascular surgery.

[37]  J. Hirsch,et al.  The role of whole blood in thrombin generation in contact with various titanium surfaces. , 2007, Biomaterials.

[38]  S. Deutsch,et al.  The Influence of Device Position on the Flow Within the Penn State 12 cc Pediatric Ventricular Assist Device , 2012, ASAIO journal.

[39]  Bartley P Griffith,et al.  Functional and biocompatibility performances of an integrated Maglev pump-oxygenator. , 2009, Artificial organs.

[40]  E. Edelman,et al.  Stent Thrombogenicity Early in High-Risk Interventional Settings Is Driven by Stent Design and Deployment and Protected by Polymer-Drug Coatings , 2011, Circulation.

[41]  J. Granada,et al.  Impact of Stent Surface on Thrombogenicity and Vascular Healing: A Comparative Analysis of Metallic and Polymeric Surfaces , 2013, Circulation. Cardiovascular interventions.

[42]  M. Ruel,et al.  Cerebral Microembolization After Bioprosthetic Aortic Valve Replacement: Comparison of Warfarin Plus Aspirin Versus Aspirin Only , 2012, Circulation.

[43]  Ulrich Steinseifer,et al.  A low-volume tester for the thrombogenic potential of mechanical heart valve prostheses. , 2011, The Journal of heart valve disease.

[44]  Samer A. Kseibi,et al.  Neuropsychological deficits after mechanical aortic valve replacement. , 2005, The Journal of heart valve disease.

[45]  Olga Pierrakos,et al.  Time-resolved DPIV analysis of vortex dynamics in a left ventricular model through bileaflet mechanical and porcine heart valve prostheses. , 2004, Journal of biomechanical engineering.

[46]  J. Christy,et al.  Evaluation of an in-vitro thrombosis assessment procedure by application to the Medtronic Parallel and St. Jude Medical valves. , 2004, The Journal of heart valve disease.

[47]  Matthias Loebe,et al.  What Price Support? Ventricular Assist Device Induced Systemic Response , 2003, ASAIO journal.

[48]  Shmuel Einav,et al.  Device Thrombogenicity Emulator (DTE)--design optimization methodology for cardiovascular devices: a study in two bileaflet MHV designs. , 2010, Journal of biomechanics.

[49]  C. Bode,et al.  Differences of platelet adhesion and thrombus activation on amorphous silicon carbide, magnesium alloy, stainless steel, and cobalt chromium stent surfaces , 2009, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[50]  G Rosenberg,et al.  Relative blood damage in the three phases of a prosthetic heart valve flow cycle. , 1993, ASAIO journal.

[51]  S. Goto,et al.  Distinct mechanisms of platelet aggregation as a consequence of different shearing flow conditions. , 1998, The Journal of clinical investigation.

[52]  R. Sodian,et al.  Long-term self-management of anticoagulation therapy after mechanical heart valve replacement in outside trial conditions. , 2012, Interactive cardiovascular and thoracic surgery.

[53]  B. Meuris Research on biological and mechanical heart valves: experimental studies in chronic animal models. , 2002, Verhandelingen - Koninklijke Academie voor Geneeskunde van Belgie.

[54]  K Watanabe,et al.  The role of von Willebrand factor and fibrinogen in platelet aggregation under varying shear stress. , 1991, The Journal of clinical investigation.

[55]  Gaetano Burriesci,et al.  Polymeric heart valves: new materials, emerging hopes. , 2009, Trends in biotechnology.

[56]  L. Hillis,et al.  Intra-aortic balloon counterpulsation. , 2006, The American journal of cardiology.

[57]  D. Ku,et al.  Fluid mechanics of vascular systems, diseases, and thrombosis. , 1999, Annual review of biomedical engineering.

[58]  A. Federici,et al.  Activation-independent platelet adhesion and aggregation under elevated shear stress. , 2005, Blood.

[59]  Steven Deutsch,et al.  Flow Field Study Comparing Design Iterations of a 50 cc Left Ventricular Assist Device , 2011, ASAIO journal.

[60]  Steven Deutsch,et al.  Flow Visualization of a Pediatric Ventricular Assist Device During Stroke Volume Reductions Related to Weaning , 2011, Annals of Biomedical Engineering.

[61]  M. Rosenfeld,et al.  Can vortices in the flow across mechanical heart valves contribute to cavitation? , 2006, Medical and Biological Engineering and Computing.

[62]  M. Borggrefe,et al.  Thrombogenicity of sirolimus-eluting stents and bare metal stents: evaluation in the early phase after stent implantation. , 2010, In vivo.

[63]  Hélène A. Simon,et al.  A Numerical Investigation of Blood Damage in the Hinge Area of Aortic Bileaflet Mechanical Heart Valves During the Leakage Phase , 2012, Annals of Biomedical Engineering.

[64]  Shmuel Einav,et al.  The Syncardia(™) total artificial heart: in vivo, in vitro, and computational modeling studies. , 2013, Journal of biomechanics.

[65]  Craig R Forest,et al.  Microfluidic system for simultaneous optical measurement of platelet aggregation at multiple shear rates in whole blood. , 2012, Lab on a chip.

[66]  A. Wolberg,et al.  A systematic evaluation of the effect of temperature on coagulation enzyme activity and platelet function. , 2004, The Journal of trauma.

[67]  Arnan Mitchell,et al.  A shear gradient–dependent platelet aggregation mechanism drives thrombus formation , 2009, Nature Medicine.

[68]  Fotis Sotiropoulos,et al.  Simulation of the Three-Dimensional Hinge Flow Fields of a Bileaflet Mechanical Heart Valve Under Aortic Conditions , 2009, Annals of Biomedical Engineering.

[69]  M. Scharfschwerdt,et al.  In-Vitro Localization of Initial Flow-Induced Thrombus Formation in Bileaflet Mechanical Heart Valves , 2009, ASAIO journal.

[70]  Ulrich Steinseifer,et al.  Transient, Three-Dimensional Flow Field Simulation through a Mechanical, Trileaflet Heart Valve Prosthesis , 2011, ASAIO journal.

[71]  A. Yoganathan,et al.  Thrombin formation in vitro in response to shear-induced activation of platelets. , 2007, Thrombosis research.

[72]  M. Xenos,et al.  Erratum to: Evaluation of Shear-Induced Platelet Activation Models Under Constant and Dynamic Shear Stress Loading Conditions Relevant to Devices , 2013, Annals of Biomedical Engineering.

[73]  T. Schmitz-Rode,et al.  Thrombogenic evaluation of two mechanical heart valve prostheses using a new in-vitro test system. , 2009, The Journal of heart valve disease.

[74]  A. Yoganathan,et al.  Passive flow control of bileaflet mechanical heart valve leakage flow. , 2008, Journal of biomechanics.

[75]  Danny Bluestein,et al.  Flow-Induced Platelet Activation in Bileaflet and Monoleaflet Mechanical Heart Valves , 2004, Annals of Biomedical Engineering.

[76]  J. Christy,et al.  An in-vitro technique for assessment of thrombogenicity in mechanical prosthetic cardiac valves: evaluation with a range of valve types. , 2004, The Journal of heart valve disease.