A high sensitivity fiber optic macro-bend based gas flow rate transducer for low flow rates: theory, working principle, and static calibration.

A novel fiber optic macro-bend based gas flowmeter for low flow rates is presented. Theoretical analysis of the sensor working principle, design, and static calibration were performed. The measuring system consists of: an optical fiber, a light emitting diode (LED), a Quadrant position sensitive Detector (QD), and an analog electronic circuit for signal processing. The fiber tip undergoes a deflection in the flow, acting like a cantilever. The consequent displacement of light spot center is monitored by the QD generating four unbalanced photocurrents which are function of fiber tip position. The analog electronic circuit processes the photocurrents providing voltage signal proportional to light spot position. A circular target was placed on the fiber in order to increase the sensing surface. Sensor, tested in the measurement range up to 10 l min(-1), shows a discrimination threshold of 2 l min(-1), extremely low fluid dynamic resistance (0.17 Pa min l(-1)), and high sensitivity, also at low flow rates (i.e., 33 mV min l(-1) up to 4 l min(-1) and 98 mV min l(-1) from 4 l min(-1) up to 10 l min(-1)). Experimental results agree with the theoretical predictions. The high sensitivity, along with the reduced dimension and negligible pressure drop, makes the proposed transducer suitable for medical applications in neonatal ventilation.

[1]  Eric Udd,et al.  Fiber Optic Sensors , 1992 .

[2]  Sergio Silvestri,et al.  An optical fiber based flow transducer for infant ventilation: Measurement principle and calibration , 2011, 2011 IEEE International Symposium on Medical Measurements and Applications.

[3]  A Cantilever Force Sensor Combined With a Spherical Reflecting Mirror for Sensitivity Enhancement of an Optical Detection System , 2009, IEEE Journal of Selected Topics in Quantum Electronics.

[4]  Qingping Yang,et al.  DP flow sensor using optical fibre Bragg grating , 2001 .

[5]  Ping Lu,et al.  Fiber Bragg grating sensor for simultaneous measurement of flow rate and direction , 2008 .

[6]  Andrea Scorza,et al.  Preliminary evaluation of a fiber-optic sensor for flow measurements in pulmonary ventilators , 2011, 2011 IEEE International Symposium on Medical Measurements and Applications.

[7]  Yong Zhao,et al.  Novel target type flowmeter based on a differential fiber Bragg grating sensor , 2005 .

[8]  Mark Johnson Photodetection and Measurement: Maximizing Performance in Optical Systems , 2003 .

[9]  Sergio Silvestri,et al.  A novel target-type low pressure drop bidirectional optoelectronic air flow sensor for infant artificial ventilation: measurement principle and static calibration. , 2011, The Review of scientific instruments.

[10]  Xu Guang Huang,et al.  A Simple Fiber-Optic Flowmeter Based on Bending Loss , 2009, IEEE Sensors Journal.

[11]  Sergio Silvestri,et al.  Optical-Fiber Measurement Systems for Medical Applications , 2011 .

[12]  R. N. Karekar,et al.  Optical fiber‐based macrobend free air flow sensor using a hinge joint: A preliminary report , 2008 .