Metallurgical Applications of X‐Ray Fluorescent Analysis
暂无分享,去创建一个
Techniques are described for fluorescent analysis using an x‐ray Geiger counter goniometer. Consideration is given to the influence of potential variations, specimen preparation, and thickness of the diffracting crystal.The observed intensity of fluorescent radiation from a number of common commercially pure metallic elements was compared under a constant primary radiation that can be closely maintained in practice.Intensities of fluorescent spectra were measured on constituents of Fe‐Ni, Fe‐Cr, and Fe‐Ni‐Cr alloys. Calibration curves correlating line intensity with weight percentage are included for binary Fe‐Ni and Fe‐Cr alloys.The thickness beyond which the fluorescent intensity becomes constant has been determined for pure iron, nickel, and chromium. Up to the above thickness for each element an empirical plot of fluorescent intensity against thickness of layer has been established. These empirical plots can be applied to measure the thickness of very thin coatings.Illustrations are given indicating t...