The design of a finger wear simulator and preliminary results

Abstract A dual-cycle finger wear simulator has been designed, manufactured and commissioned. The simulator interspersed dynamic flexion-extension motion under light load with a heavier static ‘pinch’ load to a test prosthesis immersed in a lubricant heated to 37°C. A validation test was undertaken on a size 2 Swanson prosthesis, leading to prosthesis failure in less than 1 million cycles. A second test was carried out on a Durham metacarpophalangeal prosthesis. After 4.8 million cycles a total wear factor for the joint of 0.60 × 10−6 mm3/N m was calculated, with no cracks or damage visible. Both test results compare well with earlier tests undertaken on the Stokoe finger wear simulator.

[1]  K. Doi,et al.  Alumina ceramic finger implants: a preliminary biomaterial and clinical evaluation. , 1984, The Journal of hand surgery.

[2]  Walker Ps,et al.  Laboratory evaluation of a metal-plastic type of metacarpophalangeal joint prosthesis. , 1975 .

[3]  J. J. Niebauer,et al.  Silicone-dacron hinge prosthesis. Design, evaluation, and application. , 1969, Annals of the rheumatic diseases.

[4]  A. Unsworth,et al.  Wear of cross-linked polyethylene against itself: a material suitable for surface replacement of the finger joint. , 1991, Journal of biomedical engineering.

[5]  K. An,et al.  Three-dimensional geometric analysis of the metacarpophalangeal joint. , 1988, The Journal of hand surgery.

[6]  A Unsworth,et al.  A microcomputer controlled hand assessment system used for clinical measurement. , 1985, Engineering in medicine.

[7]  S. Simon,et al.  Environmental fatigue testing of silastic finger joint prostheses. , 1972, Journal of biomedical materials research.

[8]  A. Swanson,et al.  Flexible implant arthroplasty for arthritic finger joints: rationale, technique, and results of treatment. , 1972, The Journal of bone and joint surgery. American volume.

[10]  A Unsworth,et al.  A Finger Function Simulator and the Laboratory Testing of Joint Replacements , 1990, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[11]  A A Amis,et al.  Finger joint force predictions related to design of joint replacements. , 1982, Journal of biomedical engineering.

[12]  A new metacaprophalangeal joint prosthesis , 1975 .

[13]  A. Swanson Flexible implant arthroplasty for arthritic finger joints , 1973 .

[14]  Anthony Unsworth,et al.  Dimensions of the Metacarpo-Phalangeal Joint with Particular Reference to Joint Prostheses , 1979 .

[15]  P S Walker,et al.  An apparatus to assess function of the hand. , 1978, The Journal of hand surgery.

[16]  B. Weightman,et al.  The Laboratory Development of a new Metacarpophalangeal Prosthesis , 1983, The Hand.

[17]  V. Mathiowetz,et al.  Grip and pinch strength: normative data for adults. , 1985, Archives of physical medicine and rehabilitation.

[18]  H. Harasaki,et al.  A new finger joint prosthesis. , 1993, ASAIO journal.