Energy- and time-resolved detection of prompt gamma-rays for proton range verification

In this work, we present experimental results of a novel prompt gamma-ray detector for proton beam range verification. The detection system features an actively shielded cerium-doped lanthanum(III) bromide scintillator, coupled to a digital data acquisition system. The acquisition was synchronized to the cyclotron radio frequency to separate the prompt gamma-ray signals from the later-arriving neutron-induced background. We designed the detector to provide a high energy resolution and an effective reduction of background events, enabling discrete proton-induced prompt gamma lines to be resolved. Measuring discrete prompt gamma lines has several benefits for range verification. As the discrete energies correspond to specific nuclear transitions, the magnitudes of the different gamma lines have unique correlations with the proton energy and can be directly related to nuclear reaction cross sections. The quantification of discrete gamma lines also enables elemental analysis of tissue in the beam path, providing a better prediction of prompt gamma-ray yields. We present the results of experiments in which a water phantom was irradiated with proton pencil-beams in a clinical proton therapy gantry. A slit collimator was used to collimate the prompt gamma-rays, and measurements were performed at 27 positions along the path of proton beams with ranges of 9, 16 and 23 g cm(-2) in water. The magnitudes of discrete gamma lines at 4.44, 5.2 and 6.13 MeV were quantified. The prompt gamma lines were found to be clearly resolved in dimensions of energy and time, and had a reproducible correlation with the proton depth-dose curve. We conclude that the measurement of discrete prompt gamma-rays for in vivo range verification of clinical proton beams is feasible, and plan to further study methods and detector designs for clinical use.

[1]  D Robertson,et al.  Optimizing a three-stage Compton camera for measuring prompt gamma rays emitted during proton radiotherapy , 2010, Physics in medicine and biology.

[2]  Norman,et al.  Measurements of cross sections relevant to gamma -ray line astronomy. , 1988, Physical review. C, Nuclear physics.

[3]  I Buvat,et al.  Distributions of secondary particles in proton and carbon-ion therapy: a comparison between GATE/Geant4 and FLUKA Monte Carlo codes , 2013, Physics in medicine and biology.

[4]  S Beddar,et al.  Measurement and calculation of characteristic prompt gamma ray spectra emitted during proton irradiation , 2009, Physics in medicine and biology.

[5]  H. Paganetti Range uncertainties in proton therapy and the role of Monte Carlo simulations , 2012, Physics in medicine and biology.

[6]  P. Busca,et al.  Prompt gamma imaging with a slit camera for real-time range control in proton therapy , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[7]  J. Duprat,et al.  Measurements of nuclear $\gamma$-ray line emission in interactions of protons and $\alpha$ particles with N, O, Ne and Si , 2010, 1011.2561.

[8]  Lang,et al.  Cross sections for production of the 15.10-MeV and other astrophysically significant gamma-ray lines through excitation and spallation of 12C and 16O with protons. , 1986, Physical review. C, Nuclear physics.

[9]  Katia Parodi,et al.  Time-of-flight neutron rejection to improve prompt gamma imaging for proton range verification: a simulation study , 2012, Physics in medicine and biology.

[10]  Chan Hyeong Kim,et al.  Development of an array-type prompt gamma detection system for the online measurement of the range of the proton beam in a patient: A Monte Carlo feasibility study , 2008 .

[11]  M. Fukushima,et al.  Erratum to Levenberg-Marquardt methods with strong local convergence properties for solving nonlinear equations with convex constraints , 2005 .

[12]  H Paganetti,et al.  Systematic analysis of biological and physical limitations of proton beam range verification with offline PET/CT scans , 2009, Physics in medicine and biology.

[13]  Chan Hyeong Kim,et al.  Prompt gamma measurements for locating the dose falloff region in the proton therapy , 2006 .

[14]  Antti Saastamoinen,et al.  Measurement of characteristic prompt gamma rays emitted from oxygen and carbon in tissue-equivalent samples during proton beam irradiation , 2013, Physics in medicine and biology.

[15]  H. Kooy,et al.  Golden beam data for proton pencil-beam scanning , 2012, Physics in medicine and biology.

[16]  J. Duprat,et al.  Erratum: {gamma}-ray production by proton and {alpha}-particle induced reactions on {sup 12}C, {sup 16}O, {sup 24}Mg, and Fe [Phys. Rev. C 76, 034607 (2007)] , 2009 .

[17]  H Paganetti,et al.  Monte Carlo patient study on the comparison of prompt gamma and PET imaging for range verification in proton therapy , 2011, Physics in medicine and biology.

[18]  Denis Dauvergne,et al.  Real-time monitoring of the Bragg-peak position in ion therapy by means of single photon detection , 2010, Radiation and environmental biophysics.

[19]  Joao Seco,et al.  Simulation of prompt gamma-ray emission during proton radiotherapy , 2012, Physics in medicine and biology.

[20]  K. Parodi,et al.  Suppression of random coincidences during in-beam PET measurements at ion beam radiotherapy facilities , 2005, IEEE Transactions on Nuclear Science.

[21]  Eric B. Norman,et al.  Cross sections relevant to gamma-ray astronomy: Proton induced reactions , 1981 .

[22]  J D Harrison,et al.  Risks from ionising radiation: an HPA viewpoint paper for Safegrounds , 2011, Journal of radiological protection : official journal of the Society for Radiological Protection.

[23]  J. Kiener,et al.  gamma-ray production by inelastic proton scattering on O-16 and C-12 , 1998 .