Observation on retrieved Hylamer glenoids in shoulder arthroplasty: problems associated with sterilization by gamma irradiation in air.

Hylamer is an ultra-high molecular weight polyethylene that was used in the glenoid prosthesis of the Global Shoulder. It was developed from a joint venture with DePuy Orthopaedics, Inc, and E.I. DuPont Company. In presterilization evaluations it was noted to have improved mechanical strength with superior creep and wear resistance. However, after sterilization was performed with gamma irradiation in an air environment from 1990 to 1993, the Hylamer glenoids were affected by oxidation, which altered the mechanical properties, and they became brittle and eroded. From 1993 to 1995 Hylamer glenoids were sterilized by gamma irradiation in a nitrogen environment. From 1995 to 1998 sterilization was by gas plasma processes, which do not create the oxidation problem. Since 1998 the glenoid prosthesis of the Global Shoulder has been made exclusively of Enduron and is gas sterilized. The primary purpose of this article is to provide orthopaedic surgeons the best information available about Hylamer glenoids. We make suggestions concerning the diagnosis and treatment of the patients who had a Hylamer glenoid inserted between 1990 and 1993.

[1]  J. Charnley,et al.  Rate of wear in total hip replacement. , 1975, Clinical orthopaedics and related research.

[2]  J. Currier,et al.  An analysis of hylamer and polyethylene bearings from retrieved acetabular components. , 1998, Orthopedics.

[3]  T. Wright,et al.  Post-irradiation aging of ultra-high molecular weight polyethylene. , 1994, The Journal of bone and joint surgery. American volume.

[4]  W. Harris,et al.  The problem is osteolysis. , 1995, Clinical orthopaedics and related research.

[5]  R. Zipser,et al.  Prostaglandins in hyporeninaemic hypoaldosteronism. , 1978, Lancet.

[6]  J. Graeter,et al.  Early osteolysis with Hylamer acetabular liners. , 1998, The Journal of arthroplasty.

[7]  A H Burstein,et al.  Ultra-high molecular weight polyethylene. The material and its use in total joint implants. , 1994, The Journal of bone and joint surgery. American volume.

[8]  R. Poss,et al.  Complications of Total Hip Arthroplasty Associated with the Use of an Acetabular Component with a Hylamer Liner* , 1997, The Journal of bone and joint surgery. American volume.

[9]  C. M. Agrawal,et al.  Isolation and characterization of polyethylene wear debris associated with osteolysis following total shoulder arthroplasty. , 1999, The Journal of bone and joint surgery. American volume.

[10]  M. Spector,et al.  Early delamination of a Hylamer-M tibial insert. , 1996, The Journal of arthroplasty.

[11]  W. Harris,et al.  Periprosthetic bone loss in total hip arthroplasty. Polyethylene wear debris and the concept of the effective joint space. , 1992, The Journal of bone and joint surgery. American volume.

[12]  Donald L. Wise,et al.  Encyclopedic Handbook of Biomaterials and Bioengineering , 1995 .

[13]  John M. Kelly ULTRA-HIGH MOLECULAR WEIGHT POLYETHYLENE* , 2002 .

[14]  T. Wright,et al.  The effect of carbon fiber reinforcement on contact area, contact pressure, and time-dependent deformation in polyethylene tibial components. , 1981, Journal of biomedical materials research.

[15]  B. K. Vaughn,et al.  Early osteolysis following total hip arthroplasty with use of a Hylamer liner in combination with a modular ceramic femoral head. A case report. , 1999, The Journal of bone and joint surgery. American volume.

[16]  K. Saum,et al.  Impact of gamma sterilization on clinical performance of polyethylene in the knee. , 1996, The Journal of arthroplasty.

[17]  C. Engh,et al.  Examination of wear in Duraloc acetabular components: two- to five-year evaluation of Hylamer and Enduron liners. , 1998, The Journal of arthroplasty.

[18]  A. Bantjes,et al.  Anticoagulant activity of a synthetic heparinoid in relation to molecular weight and N-sulfate content. , 1981, Journal of biomedical materials research.