Limitations of transcutaneous carbon dioxide measurements for assessing long-term mechanical ventilation.

STUDY OBJECTIVES Transcutaneous CO(2) pressure (Ptcco(2)) and transcutaneous O(2) pressure (Ptco(2)) measurements are routinely used in pediatric ICUs in order to avoid serial arterial punctures. The aim of this study was to determine the value of Ptcco(2) assessment during the evaluation of home ventilation in 12 adult patients with COPD or restrictive respiratory failure in the stable state (mean [+/- SD] basal Paco(2), 48.8 +/- 8.3 mm Hg) who were treated by mask or tracheotomy-mediated ventilation. METHODS After radial catheter insertion, patients were instructed to breathe spontaneously for 40 min and then to receive ventilation for 40 min according to their individual home ventilation modalities. An in vivo calibration was performed in the initial stage of the study in order to optimize the arterial Pco(2) and Ptcco(2) values. Every 5 min, transcutaneous measurements were performed and simultaneously compared with arterial values. MEASUREMENTS AND RESULTS Ptcco(2) and Ptco(2) were correlated with arterial values (p < 0.0001) except for Paco(2) values of > 56 mm Hg and Pao(2) values of > 115 mm Hg. During ventilation, Paco(2) decreased >or= 4 mm Hg in seven patients. Ptcco(2) variations recorded during consecutive 5-min periods while the patient received mechanical ventilation were well correlated with the arterial variations (p = 0.0033), with a delay of < 5 min. CONCLUSION Ptcco(2) values and variations accurately reflected Paco(2) values and variations during mechanical ventilation. However, the accuracy of these data seems to be restricted to patients with Paco(2) values of < 56 mm Hg.

[1]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[2]  B Hannhart,et al.  Validity of transcutaneous oxygen/carbon dioxide pressure measurement in the monitoring of mechanical ventilation in stable chronic respiratory failure. , 1999, The European respiratory journal.

[3]  C. Planès,et al.  Arterial blood gases during exercise: validity of transcutaneous measurements. , 2001, Archives of physical medicine and rehabilitation.

[4]  A. Beckett,et al.  AKUFO AND IBARAPA. , 1965, Lancet.

[5]  B. Make,et al.  Mechanical ventilation beyond the intensive care unit. Report of a consensus conference of the American College of Chest Physicians. , 1998, Chest.

[6]  J. Dziodzio,et al.  Trancutaneous PCO2 and end-tidal PCO2 in ventilated adults , 1992 .

[7]  Martin J. Tobin,et al.  Principles and Practice of Intensive Care Monitoring. , 1997 .

[8]  J. Chevrolet,et al.  Transcutaneous PCO2 to monitor noninvasive mechanical ventilation in adults: assessment of a new transcutaneous PCO2 device. , 1998, Chest.

[9]  J. Tobias,et al.  Comparison of end-tidal and transcutaneous measures of carbon dioxide during general anaesthesia in severely obese adults. , 2003, British journal of anaesthesia.

[10]  A. Rebuck,et al.  Response characteristics of a dual transcutaneous oxygen/carbon dioxide monitoring system. , 1991, Chest.

[11]  D. Holmgren,et al.  Repeated blood gas monitoring in healthy children and adolescents by the transcutaneous route , 2003, Pediatric pulmonology.

[12]  K. Sumikawa,et al.  A Comparative Evaluation of Transcutaneous and End-Tidal Measurements of CO2 in Thoracic Anesthesia , 2003, Anesthesia and analgesia.

[13]  N. Cherniack,et al.  Oxygen and carbon dioxide gas stores of the body. , 1970, Physiological reviews.

[14]  C. Lang,et al.  Apnea testing guided by continuous transcutaneous monitoring of partial pressure of carbon dioxide. , 1998, Critical care medicine.

[15]  R. Carter,et al.  Use of a combined oxygen and carbon dioxide transcutaneous electrode in the estimation of gas exchange during exercise. , 1993, Thorax.