Magnetization transfer effects in multislice MR imaging.

A theoretical model is presented which describes the effects of magnetization transfer in multislice MR imaging of a tissue-mimicking phantom composed of cross-linked agar gel. The model is successful in explaining differences between single and multislice image signal intensities observed for the agar gel but not seen in a simple aqueous solution. Magnetization transfer leads to a reduction in the image signal intensity of a slice of interest due to off-resonance RF irradiation arising from 90 degrees and 180 degrees pulses intended for neighboring slices. The contribution of magnetization transfer to multislice MR imaging depends on the amount of off-resonance RF irradiation during the imaging sequence repetition interval. For the tissue-mimicking agar gel, conventional spin-echo multislice imaging gave rise to a negligible image signal intensity reduction (< or = 2%); however, fast spin-echo (FSE) imaging, which employs up to 16 times as many RF pulses per slice, exhibited as much as a 13% reduction in image signal intensity (13 slices). The reduction in multislice image signal intensity due to magnetization transfer is sample specific and is shown to be more dramatic for in vivo human leg muscle (10% for conventional spin echo, 40% for FSE) where magnetization transfer rates are greater than in the cross-linked agar gel.

[1]  T. Child,et al.  Steady‐state and pulsed NMR studies of gelation in aqueous agarose , 1972 .

[2]  S. Ablett,et al.  Nuclear magnetic resonance investigations of polysaccharide films, sols, and gels: I. Agarose , 1978 .

[3]  J. Gore,et al.  Quantitative studies of hydrodynamic effects and cross‐relaxation in protein solutions and tissues with proton and deuteron longitudinal relaxation times , 1990, Magnetic resonance in medicine.

[4]  S. Kennedy,et al.  Comparison of agarose and cross-linked protein gels as magnetic resonance imaging phantoms. , 1991, Magnetic resonance imaging.

[5]  I. Young,et al.  Variations in slice shape and absorption as artifacts in the determination of tissue parameters in NMR imaging , 1985, Magnetic resonance in medicine.

[6]  J. B. Kneeland,et al.  Effect of intersection spacing on MR image contrast and study time. , 1986, Radiology.

[7]  F. Hyder,et al.  Applications of nuclear magnetic cross‐relaxation spectroscopy to tissues , 1991, Magnetic resonance in medicine.

[8]  A. Crawley,et al.  Errors in T2 estimation using multislice multiple‐echo imaging , 1987, Magnetic resonance in medicine.

[9]  R S Balaban,et al.  Quantitative 1H magnetization transfer imaging in vivo , 1991, Magnetic resonance in medicine.

[10]  J C Gore,et al.  Errors in the measurements of T2 using multiple‐echo MRI techniques. I. Effects of radiofrequency pulse imperfections , 1986, Magnetic resonance in medicine.

[11]  R. Balaban,et al.  Magnetization transfer contrast (MTC) and tissue water proton relaxation in vivo , 1989, Magnetic resonance in medicine.

[12]  E. Samulski,et al.  The measurement of cross-relaxation effects in the proton NMR spin-lattice relaxation of water in biological systems: Hydrated collagen and muscle☆ , 1978 .

[13]  G D Fullerton,et al.  Characterization of proton NMR relaxation times in normal and pathological tissues by correlation with other tissue parameters. , 1984, Magnetic resonance imaging.

[14]  R M Henkelman,et al.  Effect of multislice interference on image contrast in T2- and T1-weighted MR images. , 1988, AJNR. American journal of neuroradiology.

[15]  Bob S. Hu,et al.  Pulsed saturation transfer contrast , 1992, Magnetic resonance in medicine.

[16]  H Engels,et al.  Incidental magnetization transfer contrast in standard multislice imaging. , 1990, Magnetic resonance imaging.

[17]  J Hennig,et al.  RARE imaging: A fast imaging method for clinical MR , 1986, Magnetic resonance in medicine.

[18]  E. Madsen,et al.  Tissue-mimicking gelatin-agar gels for use in magnetic resonance imaging phantoms. , 1988, Medical physics.

[19]  J. Pope,et al.  A pulsed NMR study of nuclear spin-lattice relaxation in the off resonance rotating frame , 1974 .

[20]  S. Forsén,et al.  Study of Moderately Rapid Chemical Exchange Reactions by Means of Nuclear Magnetic Double Resonance , 1963 .

[21]  P. Pfannenstiel,et al.  Errors in T1-determination using multislice technique and Gaussian slice profiles. , 1988, Magnetic resonance imaging.

[22]  S. H. Koenig,et al.  Magnetic cross-relaxation among protons in protein solutions. , 1978, Biochemistry.

[23]  H. C. Torrey Transient Nutations in Nuclear Magnetic Resonance , 1949 .

[24]  R V Mulkern,et al.  Magnetization transfer effects in multislice RARE sequences , 1992, Magnetic resonance in medicine.

[25]  A. Crawley,et al.  A stimulated echo artifact from slice interference in magnetic resonance imaging. , 1987, Medical physics.

[26]  T. Schleich,et al.  Incorporation of magnetization transfer into the formalism for rotating-frame spin-lattice proton NMR relaxation in the presence of an off-resonance-irradiation field , 1991 .

[27]  M. Bronskill,et al.  Variation in measured transverse relaxation in tissue resulting from spin locking with the CPMG sequence , 1988 .

[28]  L P Panych,et al.  Comparing the FAISE method with conventional dual‐echo sequences , 1991, Journal of magnetic resonance imaging : JMRI.