A fast and accurate simulator for the design of birdcage coils in MRI

The birdcage coils are extensively used in MRI systems since they introduce a high signal to noise ratio and a high radiofrequency magnetic field homogeneity that guarantee a large field of view. The present article describes the implementation of a birdcage coil simulator, operating in high-pass and low-pass modes, using magnetostatic analysis of the coil. Respect to other simulators described in literature, our simulator allows to obtain in short time not only the dominant frequency mode, but also the complete resonant frequency spectrum and the relevant magnetic field pattern with high accuracy. Our simulator accounts for all the inductances including the mutual inductances between conductors. Moreover, the inductance calculation includes an accurately birdcage geometry description and the effect of a radiofrequency shield. The knowledge of all the resonance modes introduced by a birdcage coil is twofold useful during birdcage coil design: —higher order modes should be pushed far from the fundamental one, —for particular applications, it is necessary to localize other resonant modes (as the Helmholtz mode) jointly to the dominant mode. The knowledge of the magnetic field pattern allows to a priori verify the field homogeneity created inside the coil, when varying the coil dimension and mainly the number of the coil legs. The coil is analyzed using equivalent circuit method. Finally, the simulator is validated by implementing a low-pass birdcage coil and comparing our data with the literature.

[1]  T. M. Fitzgerald,et al.  Generalized electrical analysis of low-pass and high-pass birdcage resonators. , 1991, Magnetic resonance imaging.

[2]  R.L. Magin,et al.  A simple method to incorporate the effects of an RF shield into RF resonator analysis for MRI applications , 1995, IEEE Transactions on Biomedical Engineering.

[3]  M. Morich,et al.  Novel quadrature birdcage coil for a vertical B0 field open MRI system , 2000, Magnetic resonance in medicine.

[4]  Mark C. Leifer,et al.  Resonant Modes of the Birdcage Coil , 1997 .

[5]  Raghbir Singh Khandpur,et al.  Handbook of Biomedical Instrumentation , 1987 .

[6]  Frederick Warren Grover,et al.  Inductance Calculations: Working Formulas and Tables , 1981 .

[7]  Luc Darrasse,et al.  Quick measurement of NMR‐coil sensitivity with a dual‐loop probe , 1993 .

[8]  Jianming Jin Electromagnetic Analysis and Design in Magnetic Resonance Imaging , 1998 .

[9]  D. Hoult,et al.  Quadrature detection coils—A further √2 improvement in sensitivity , 1983 .

[10]  P M Joseph,et al.  A technique for double resonant operation of birdcage imaging coils. , 1989, IEEE transactions on medical imaging.

[11]  J. Schenck,et al.  An efficient, highly homogeneous radiofrequency coil for whole-body NMR imaging at 1.5 T , 1985 .

[12]  P. Cahill,et al.  Explicit treatment of mutual inductance in eight-column birdcage resonators. , 1992, Magnetic resonance imaging.

[13]  Tropp Mutual Inductance in the Bird-Cage Resonator , 1997, Journal of magnetic resonance.

[14]  J. Tropp The theory of the bird-cage resonator , 1989 .

[15]  K. Thulborn,et al.  Experimentally verified, theoretical design of dual‐tuned, low‐pass birdcage radiofrequency resonators for magnetic resonance imaging and magnetic resonance spectroscopy of human brain at 3.0 Tesla , 1999, Magnetic resonance in medicine.

[16]  Stuart Crozier,et al.  Optimized Small-Bore, High-Pass Resonator Designs , 1995 .

[17]  Michael B. Smith,et al.  A method to create an optimum current distribution and homogeneous B1 field for elliptical birdcage coils , 1997, Magnetic resonance in medicine.

[18]  Christopher M. Collins,et al.  A Method for Accurate Calculation ofB1Fields in Three Dimensions. Effects of Shield Geometry on Field Strength and Homogeneity in the Birdcage Coil , 1997 .