The bionic eye: A review of multielectrode arrays

The notion of creating artificial vision using visual prostheses has been well represented though science fiction literature and films. When we think of retinal prostheses, we immediately think of fictional characters like The Terminator scanning across a bar to assess patrons for appropriately fitting clothing, or Star Trek’s Geordi La Forge with his VISOR, a visual instrument and sensory organ replacement placed across his eyes and attached into his temples to provide him with vision. Such devices are no longer farfetched. In the past 20 years, significant research has been undertaken across the globe in the race for a “Bionic Eye”. Advances in Bionic Eye research have come from improvements in the design and fabrication of multielectrode arrays (MEAs) for medical applications. MEAs are already commonplace in medicine with use in applications such as the cochlear device, cardiac pacemakers, and deep brain stimulators where interfacing with neuronal cell populations is required. The use of MEAs for vision prostheses is currently of significant interest. For the most part, retinal prostheses have dominated the research landscape owing to the ease of access and direct contact to the retinal ganglion nerve cells. However, MEAs are also in use for direct stimulation into the optic nerve [1]. Retinal prostheses bypass the damaged photoreceptor cells within the retina and instead replace the degenerate retina with electrical stimulation to the nerve cells. Using electrical stimulation, stimulated retinal ganglion cells have been shown to elicit a percept in the form of a phosphene in blind patients [2–6]. Accordingly, the two diseases commonly linked to the justification for Bionic Eye research are age-related macular degeneration (AMD) and retinitis pigmentosa (RP), diseases which lead to progressive loss of photoreceptor cells and diseases where the patient has had previous vision and thus exhibits prior visual-brain pathways. At present, there has been no reliable cure for any of the retinal diseases that target the photoreceptor cells, and thus the development of prosthetic devices is a viable clinical treatment option [7–9]. The best physical location for a retinal implant remains undecided. Although this chapter concentrates on retinal prostheses, other artificial vision devices have been investigated at sites external to the eye such as the optic nerve [10] and the visual cortex [11]. Referring back to retinal prostheses, at present there have been three sites best identified for potential implant positioning: (i) epiretinal; (ii) subretinal; and (iii) suprachoroidal. Figure 24.1 shows a drawing of an eye, highlighting the three device positions relative to the retinal ganglion nerve cell layer. The implant sites have been selected for various reasons such as distance from the nerve cells, surgical ease and mechanical stability. The majority of clinical groups have opted for the epiretinal implantation of their Bionic Eye devices [3, 12–16]. The epiretinal position offers the advantage that the surgical procedure is relatively well understood and the device is closely aligned with the site it is actively stimulating. In particular, the epiretinal position aligns the device against the intact retinal ganglion cells

[1]  R. Hornig,et al.  The IMI Retinal Implant System , 2007 .

[2]  R. Hornig,et al.  Chronic Epiretinal Chip Implant in Blind Patients With Retinitis Pigmentosa: Long-Term Clinical Results , 2007 .

[3]  T.L. Rose,et al.  Electrical stimulation with Pt electrodes. VIII. Electrochemically safe charge injection limits with 0.2 ms pulses (neuronal application) , 1990, IEEE Transactions on Biomedical Engineering.

[4]  Spencer C. Chen,et al.  Simulating prosthetic vision: I. Visual models of phosphenes , 2009, Vision Research.

[5]  Joseph F. Rizzo,et al.  Communication and control system for a 15-channel hermetic retinal prosthesis , 2011, Biomed. Signal Process. Control..

[6]  Thomas Laube,et al.  Acute electrical stimulation of the human retina with an epiretinal electrode array , 2012, Acta ophthalmologica.

[7]  Joel Villalobos,et al.  A wide-field suprachoroidal retinal prosthesis is stable and well tolerated following chronic implantation. , 2013, Investigative ophthalmology & visual science.

[8]  U. Schnakenberg,et al.  Sputtered platinum–iridium layers as electrode material for functional electrostimulation , 2011 .

[9]  J M Marston,et al.  Electrical stimulation with pt electrodes. IV. Factors influencing Pt dissolution in inorganic saline. , 1980, Biomaterials.

[10]  T. Tokuda,et al.  In vivo Stimulation on Rabbit Retina using CMOS LSI-based Multi-Chip Flexible Stimulator for Retinal Prosthesis , 2007, 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[11]  T Fujikado,et al.  Laboratory investigation of microelectronics-based stimulators for large-scale suprachoroidal transretinal stimulation (STS) , 2007, Journal of neural engineering.

[12]  Mark S Humayun,et al.  Frequency and amplitude modulation have different effects on the percepts elicited by retinal stimulation. , 2012, Investigative ophthalmology & visual science.

[13]  Kate Fox,et al.  An all-diamond, hermetic electrical feedthrough array for a retinal prosthesis. , 2014, Biomaterials.

[14]  Jessy D. Dorn,et al.  The Detection of Motion by Blind Subjects With the Epiretinal 60-Electrode (Argus II) Retinal Prosthesis. , 2013, JAMA ophthalmology.

[15]  J. Weiland,et al.  Perceptual thresholds and electrode impedance in three retinal prosthesis subjects , 2005, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[16]  S. Cogan,et al.  Sputtered iridium oxide films for neural stimulation electrodes. , 2009, Journal of biomedical materials research. Part B, Applied biomaterials.

[17]  S. B. Brummer,et al.  Electrochemical Considerations for Safe Electrical Stimulation of the Nervous System with Platinum Electrodes , 1977, IEEE Transactions on Biomedical Engineering.

[18]  Alfred Stett,et al.  Electronic Implants Provide Continuous Stable Percepts in Blind Volunteers Only if the Image Receiver is Directly Linked to Eye Movement , 2010 .

[19]  John P. Donoghue,et al.  Connecting cortex to machines: recent advances in brain interfaces , 2002, Nature Neuroscience.

[20]  Chris E. Williams,et al.  Visual cortex responses to suprachoroidal electrical stimulation of the retina: effects of electrode return configuration. , 2012, Journal of neural engineering.

[21]  Eduardo Fernandez,et al.  Toward the development of a cortically based visual neuroprosthesis , 2009, Journal of neural engineering.

[22]  E. Zrenner,et al.  Visualisation of active subretinal implants with external connections by high-resolution CT , 2010, British Journal of Ophthalmology.

[23]  J. Menzel-Severing Emerging techniques to treat corneal neovascularisation , 2012, Eye.

[24]  Ronald T. Leung,et al.  Electrical stimulation of retinal ganglion cells with diamond and the development of an all diamond retinal prosthesis. , 2012, Biomaterials.

[25]  J. Weiland,et al.  Visual performance using a retinal prosthesis in three subjects with retinitis pigmentosa. , 2007, American journal of ophthalmology.

[26]  Bernd Eggers Artificial Sight Basic Research Biomedical Engineering And Clinical Advances , 2016 .

[27]  James D. Weiland,et al.  In vitro electrical properties for iridium oxide versus titanium nitride stimulating electrodes , 2002, IEEE Transactions on Biomedical Engineering.

[28]  J. Weiland,et al.  Pattern electrical stimulation of the human retina , 1999, Vision Research.

[29]  S. J. Kim,et al.  A Suprachoroidal Electrical Retinal Stimulator Design for Long-Term Animal Experiments and In Vivo Assessment of Its Feasibility and Biocompatibility in Rabbits , 2008, Journal of biomedicine & biotechnology.

[30]  Albrecht Rothermel,et al.  A CMOS Chip With Active Pixel Array and Specific Test Features for Subretinal Implantation , 2009, IEEE Journal of Solid-State Circuits.

[31]  B. Wilhelm,et al.  Spatial resolution and perception of patterns mediated by a subretinal 16-electrode array in patients blinded by hereditary retinal dystrophies. , 2011, Investigative ophthalmology & visual science.

[32]  R. Hornig,et al.  Long Term Tolerability of the First Wireless Implant for Electrical Epiretinal Stimulation , 2009 .

[33]  Alfred Stett,et al.  Subretinal electronic chips allow blind patients to read letters and combine them to words , 2010, Proceedings of the Royal Society B: Biological Sciences.

[34]  Benoît Gérard,et al.  Pattern recognition with the optic nerve visual prosthesis. , 2003, Artificial organs.

[35]  Eberhart Zrenner,et al.  Factors Affecting Perceptual Thersholds of Subretinal Electric Stimulation in Blind Volunteers , 2010 .

[36]  R. Hornig,et al.  Visual Perception After Long-Term Implantation of a Retinal Implant , 2008 .

[37]  Lois S. Robblee,et al.  Charge Injection Properties of Thermally-Prepared Iridium Oxide Films , 1985 .

[38]  T. Stieglitz,et al.  Development of an inductively coupled epiretinal vision prosthesis , 2004, The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[39]  R. Hornig,et al.  [Development of an epiretinal prosthesis for stimulation of the human retina]. , 2005, Der Ophthalmologe : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft.

[40]  Alice K. Cho,et al.  Retinal prostheses: current clinical results and future needs. , 2011, Ophthalmology.

[41]  R. Hornig,et al.  Long-Term Stability of Stimulation Thresholds Obtained From a Human Patient With a Prototype of an Epiretinal Retina Prosthesis , 2009 .

[42]  Arup Roy,et al.  Factors affecting perceptual thresholds in epiretinal prostheses. , 2008, Investigative ophthalmology & visual science.

[43]  Bedrich J. Hosticka,et al.  Single chip CMOS imagers and flexible microelectronic stimulators for a retina implant system , 2000 .

[44]  H. Meffin,et al.  Diamond penetrating electrode array for Epi-Retinal Prosthesis , 2010, 2010 Annual International Conference of the IEEE Engineering in Medicine and Biology.

[45]  S. J. Kim,et al.  Biocompatibility of polyimide microelectrode array for retinal stimulation , 2004 .

[46]  X. Beebe,et al.  Charge injection limits of activated iridium oxide electrodes with 0.2 ms pulses in bicarbonate buffered saline (neurological stimulation application) , 1988, IEEE Transactions on Biomedical Engineering.

[47]  M. Mladejovsky,et al.  Artificial Vision for the Blind: Electrical Stimulation of Visual Cortex Offers Hope for a Functional Prosthesis , 1974, Science.

[48]  Ryad Benosman,et al.  Neural stimulation for visual rehabilitation: Advances and challenges , 2013, Journal of Physiology-Paris.

[49]  G J Suaning,et al.  Fabrication of implantable microelectrode arrays by laser cutting of silicone rubber and platinum foil , 2005, Journal of neural engineering.

[50]  Seung Woo Lee,et al.  Feasibility of microelectrode array (MEA) based on silicone-polyimide hybrid for retina prosthesis. , 2009, Investigative ophthalmology & visual science.

[51]  Thomas Stieglitz,et al.  Development of a micromachined epiretinal vision prosthesis , 2009, Journal of neural engineering.

[52]  R. Ali,et al.  Prospects for retinal gene replacement therapy. , 2009, Trends in genetics : TIG.

[53]  István Bókkon,et al.  Phosphene phenomenon: A new concept , 2008, Biosyst..

[54]  Subjects Blind From Outer Retinal Dystrophies Are Able To Consistently Read Short Sentences Using The ArgusTM Ii Retinal Prosthesis System , 2011 .

[55]  R. H. Propst,et al.  Visual perception elicited by electrical stimulation of retina in blind humans. , 1996, Archives of ophthalmology.

[56]  Ákos Zarándy,et al.  Vision Restoration and Vision Chip Technologies , 2011, FET.

[57]  E. Zrenner,et al.  Compound subretinal prostheses with extra-ocular parts designed for human trials: successful long-term implantation in pigs , 2007, Graefe's Archive for Clinical and Experimental Ophthalmology.

[58]  F. Hambrecht,et al.  CRITERIA FOR SELECTING ELECTRODES FOR ELECTRICAL STIMULATION: THEORETICAL AND PRACTICAL CONSIDERATIONS , 1983, Annals of the New York Academy of Sciences.

[59]  Eberhart Zrenner,et al.  Fading Of Perception In Retinal Implants Is A Function Of Time And Space Between Sites Of Stimulation , 2011 .

[60]  James Weiland,et al.  Artificial vision: needs, functioning, and testing of a retinal electronic prosthesis. , 2009, Progress in brain research.

[61]  J. Rizzo Update on Retinal Prosthetic Research: The Boston Retinal Implant Project , 2011, Journal of neuro-ophthalmology : the official journal of the North American Neuro-Ophthalmology Society.

[62]  Philip R. Troyk,et al.  In vitro comparison of the charge-injection limits of activated iridium oxide (AIROF) and platinum-iridium microelectrodes , 2005, IEEE Transactions on Biomedical Engineering.

[63]  Nigel H Lovell,et al.  Substrate dependent stability of conducting polymer coatings on medical electrodes. , 2012, Biomaterials.

[64]  A. Y. Chow,et al.  The artificial silicon retina microchip for the treatment of vision loss from retinitis pigmentosa. , 2004, Archives of ophthalmology.

[65]  S. Kelly,et al.  Perceptual efficacy of electrical stimulation of human retina with a microelectrode array during short-term surgical trials. , 2003, Investigative ophthalmology & visual science.

[66]  Chris E. Williams,et al.  Visual cortex responses to single- and simultaneous multiple-electrode stimulation of the retina: implications for retinal prostheses. , 2012, Investigative ophthalmology & visual science.

[67]  Thomas Laube,et al.  A method and technical equipment for an acute human trial to evaluate retinal implant technology , 2005, Journal of neural engineering.

[68]  Takashi Fujikado,et al.  Electrophysiological studies of the feasibility of suprachoroidal-transretinal stimulation for artificial vision in normal and RCS rats. , 2004, Investigative ophthalmology & visual science.

[69]  Yvonne Hsu-Lin Luo,et al.  The Argus II Retinal Prosthesis System , 2019, Prosthesis.

[70]  H. Kishima,et al.  Testing of semichronically implanted retinal prosthesis by suprachoroidal-transretinal stimulation in patients with retinitis pigmentosa. , 2011, Investigative ophthalmology & visual science.

[71]  S. Kelly,et al.  Methods and perceptual thresholds for short-term electrical stimulation of human retina with microelectrode arrays. , 2003, Investigative ophthalmology & visual science.

[72]  Joseph F. Rizzo,et al.  A Hermetic Wireless Subretinal Neurostimulator for Vision Prostheses , 2011, IEEE Transactions on Biomedical Engineering.

[73]  Farhad Hafezi,et al.  Patients Blinded By Outer Retinal Dystrophies Are Able To Perceive Color Using The ArgusTm II Retinal Prosthesis System , 2011 .

[74]  T. Yuen,et al.  NEUROPATHOLOGICAL EFFECTS OF INTRACEREBRAL PLATINUM SALT INJECTIONS , 1977, Journal of neuropathology and experimental neurology.

[75]  W. A. Drohan,et al.  Realization of a 15-channel, hermetically-encased wireless subretinal prosthesis for the blind , 2009, 2009 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[76]  B. Sellhaus,et al.  Implantation and explantation of a wireless epiretinal retina implant device: observations during the EPIRET3 prospective clinical trial. , 2009, Investigative ophthalmology & visual science.

[77]  Seung Woo Lee,et al.  Development of microelectrode arrays for artificial retinal implants using liquid crystal polymers. , 2009, Investigative ophthalmology & visual science.

[78]  P. Andrews,et al.  Embryonic stem cells and retinal repair , 2007, Mechanisms of Development.

[79]  Devyani Nanduri,et al.  Percept Properties of Single Electrode Stimulation in Retinal Prosthesis Subjects , 2011 .

[80]  J. Weiland,et al.  Retinal Prosthesis , 2014, IEEE Transactions on Biomedical Engineering.

[81]  Y. Tano,et al.  Evaluation of phosphenes elicited by extraocular stimulation in normals and by suprachoroidal-transretinal stimulation in patients with retinitis pigmentosa , 2007, Graefe's Archive for Clinical and Experimental Ophthalmology.

[82]  M. Humayun,et al.  Artificial vision through neuronal stimulation , 2012, Neuroscience Letters.

[83]  Farhad Hafezi,et al.  Patients blinded by outer retinal dystrophies are able to perceive simultaneous colors using the Argus® II Retinal Prosthesis System , 2012 .

[84]  S. Cogan Neural stimulation and recording electrodes. , 2008, Annual review of biomedical engineering.

[85]  Avi Caspi,et al.  The ArgusTM II Retinal Prosthesis Provides Complex Form Vision for a Subject Blinded by Retinitis Pigmentosa , 2010 .

[86]  Farhad Hafezi,et al.  Temporal properties of visual perception on electrical stimulation of the retina. , 2012, Investigative ophthalmology & visual science.

[87]  Joseph F. Rizzo,et al.  Development and Implantation of a Minimally Invasive Wireless Subretinal Neurostimulator , 2009, IEEE Transactions on Biomedical Engineering.

[88]  T. Wachtler,et al.  Stimulation with a wireless intraocular epiretinal implant elicits visual percepts in blind humans. , 2011, Investigative ophthalmology & visual science.

[89]  Philip R. Troyk,et al.  Potential-biased, asymmetric waveforms for charge-injection with activated iridium oxide (AIROF) neural stimulation electrodes , 2006, IEEE Transactions on Biomedical Engineering.

[90]  J. Sahel,et al.  Subjects Blinded By Outer Retinal Dystrophies Are Able To Recognize Outlined Shapes Using The Argus® Ii Retinal Prosthesis System: A Comparison With The Full Shapes Recognition Task , 2012 .

[91]  Chris E. Williams,et al.  Evaluation of stimulus parameters and electrode geometry for an effective suprachoroidal retinal prosthesis , 2010, Journal of neural engineering.

[92]  Kate Fox,et al.  Ultra-nanocrystalline diamond electrodes: optimization towards neural stimulation applications , 2012, Journal of neural engineering.

[93]  W. Mokwa,et al.  Intraocular epiretinal prosthesis to restore vision in blind humans , 2008, 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[94]  J. Mortimer,et al.  Platinum for neural stimulation: voltammetry considerations , 2010, Journal of neural engineering.

[95]  L. Poole-Warren,et al.  Conducting polymer electrodes for visual prostheses , 2010, 2010 Annual International Conference of the IEEE Engineering in Medicine and Biology.

[96]  S.F. Cogan,et al.  Sputtered iridium oxide films (SIROFs) for low-impedance neural stimulation and recording electrodes , 2004, The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[97]  Avi Caspi,et al.  Feasibility study of a retinal prosthesis: spatial vision with a 16-electrode implant. , 2009, Archives of ophthalmology.

[98]  Jessy D. Dorn,et al.  Interim results from the international trial of Second Sight's visual prosthesis. , 2012, Ophthalmology.

[99]  Penelope J. Allen,et al.  Development of a surgical approach for a wide-view suprachoroidal retinal prosthesis: evaluation of implantation trauma , 2012, Graefe's Archive for Clinical and Experimental Ophthalmology.

[100]  Stuart F Cogan,et al.  Over-pulsing degrades activated iridium oxide films used for intracortical neural stimulation , 2004, Journal of Neuroscience Methods.

[101]  Intracortical capacitor electrodes: preliminary evaluation , 1982, Journal of Neuroscience Methods.