Long-term reliability of Al2O3 and Parylene C bilayer encapsulated Utah electrode array based neural interfaces for chronic implantation

OBJECTIVE We focus on improving the long-term stability and functionality of neural interfaces for chronic implantation by using bilayer encapsulation. APPROACH We evaluated the long-term reliability of Utah electrode array (UEA) based neural interfaces encapsulated by 52 nm of atomic layer deposited Al2O3 and 6 µm of Parylene C bilayer, and compared these to devices with the baseline Parylene-only encapsulation. Three variants of arrays including wired, wireless, and active UEAs were used to evaluate this bilayer encapsulation scheme, and were immersed in phosphate buffered saline (PBS) at 57 °C for accelerated lifetime testing. MAIN RESULTS The median tip impedance of the bilayer encapsulated wired UEAs increased from 60 to 160 kΩ during the 960 days of equivalent soak testing at 37 °C, the opposite trend to that typically observed for Parylene encapsulated devices. The loss of the iridium oxide tip metallization and etching of the silicon tip in PBS solution contributed to the increase of impedance. The lifetime of fully integrated wireless UEAs was also tested using accelerated lifetime measurement techniques. The bilayer coated devices had stable power-up frequencies at ∼910 MHz and constant radio-frequency signal strength of -50 dBm during up to 1044 days (still under testing) of equivalent soaking time at 37 °C. This is a significant improvement over the lifetime of ∼100 days achieved with Parylene-only encapsulation at 37 °C. The preliminary samples of bilayer coated active UEAs with a flip-chip bonded ASIC chip had a steady current draw of ∼3 mA during 228 days of soak testing at 37 °C. An increase in the current draw has been consistently correlated to device failures, so is a sensitive metric for their lifetime. SIGNIFICANCE The trends of increasing electrode impedance of wired devices and performance stability of wireless and active devices support the significantly greater encapsulation performance of this bilayer encapsulation compared with Parylene-only encapsulation. The bilayer encapsulation should significantly improve the in vivo lifetime of neural interfaces for chronic implantation.

[1]  F. Solzbacher,et al.  Bi-layer encapsulation of utah array based nerual interfaces by atomic layer deposited Al2O3 and parylene C , 2013, 2013 Transducers & Eurosensors XXVII: The 17th International Conference on Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS & EUROSENSORS XXVII).

[2]  Sheryl R. Kane,et al.  Electrical Performance of Penetrating Microelectrodes Chronically Implanted in Cat Cortex , 2013, IEEE Transactions on Biomedical Engineering.

[3]  E Erik Langereis,et al.  Plasma-assisted atomic layer deposition of Al2O3 moisture permeation barriers on polymers , 2006 .

[4]  Jon A. Mukand,et al.  Neuronal ensemble control of prosthetic devices by a human with tetraplegia , 2006, Nature.

[5]  R.R. Harrison,et al.  HermesC: Low-Power Wireless Neural Recording System for Freely Moving Primates , 2009, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[6]  K. Horch,et al.  A silicon-based, three-dimensional neural interface: manufacturing processes for an intracortical electrode array , 1991, IEEE Transactions on Biomedical Engineering.

[7]  Florian Solzbacher,et al.  Lifetime assessment of atomic-layer-deposited Al2O3-Parylene C bilayer coating for neural interfaces using accelerated age testing and electrochemical characterization. , 2014, Acta biomaterialia.

[8]  N. Lago,et al.  Long term assessment of axonal regeneration through polyimide regenerative electrodes to interface the peripheral nerve. , 2005, Biomaterials.

[9]  Christian Ethier,et al.  Intrafascicular stimulation of monkey arm nerves evokes coordinated grasp and sensory responses. , 2013, Journal of neurophysiology.

[10]  David J. Warren,et al.  An automated system for measuring tip impedance and among-electrode shunting in high-electrode count microelectrode arrays , 2009, Journal of Neuroscience Methods.

[11]  Byron M. Yu,et al.  A high-performance brain–computer interface , 2006, Nature.

[12]  A Vanhoestenberghe,et al.  Corrosion of silicon integrated circuits and lifetime predictions in implantable electronic devices , 2013, Journal of neural engineering.

[13]  R. Normann,et al.  Chronic recording capability of the Utah Intracortical Electrode Array in cat sensory cortex , 1998, Journal of Neuroscience Methods.

[14]  Se Stephen Potts,et al.  Ultra-Thin Aluminium Oxide Films Deposited by Plasma-Enhanced Atomic Layer Deposition for Corrosion Protection , 2011 .

[15]  C. Dawes Laser Welding: A Practical Guide , 1993 .

[16]  Florian Solzbacher,et al.  Long-term in-vivo Investigation of Parylene-C as Encapsulation Material for Neural Interfaces , 2011 .

[17]  F. Solzbacher,et al.  Plasma-assisted atomic layer deposition of Al(2)O(3) and parylene C bi-layer encapsulation for chronic implantable electronics. , 2012, Applied physics letters.

[18]  Florian Solzbacher,et al.  Long-Term Bilayer Encapsulation Performance of Atomic Layer Deposited Al $_{\bf 2}$O$_{\bf 3}$ and Parylene C for Biomedical Implantable Devices , 2013, IEEE Transactions on Biomedical Engineering.

[19]  Florian Solzbacher,et al.  Long term in vitro functional stability and recording longevity of fully integrated wireless neural interfaces based on the Utah Slant Electrode Array , 2011, Journal of neural engineering.

[20]  R.R. Harrison,et al.  A wireless neural interface for chronic recording , 2008, 2008 IEEE Biomedical Circuits and Systems Conference.

[21]  M. Ghovanloo,et al.  A 15-Channel Wireless Neural Recording System Based on Time Division Multiplexing of Pulse Width Modulated Signals , 2006, 2006 International Conference on Microtechnologies in Medicine and Biology.

[22]  Reid R. Harrison,et al.  High channel-count neural interfaces for multiple degree- of-freedom neuroprosthetics , 2011 .

[23]  Marco Sampietro,et al.  Atomic layer deposited Al2O3 as a capping layer for polymer based transistors , 2007 .

[24]  T. Stieglitz,et al.  Characterization of parylene C as an encapsulation material for implanted neural prostheses. , 2010, Journal of biomedical materials research. Part B, Applied biomaterials.

[25]  Andrew S. Whitford,et al.  Cortical control of a prosthetic arm for self-feeding , 2008, Nature.

[26]  D W L Hukins,et al.  Accelerated aging for testing polymeric biomaterials and medical devices. , 2008, Medical engineering & physics.

[27]  Daryl R Kipke,et al.  The insulation performance of reactive parylene films in implantable electronic devices. , 2009, Biomaterials.

[28]  Toh-Ming Lu,et al.  Chemical Vapor Deposition Polymerization , 2004 .

[29]  Corrosion Behavior of Parylene-Metal-Parylene Thin Films in Saline , 2007 .

[30]  Y. P. Wu,et al.  Impact properties of PBGA assemblies reflowed in nitrogen ambient and compressed air , 2000 .

[31]  R. K. Roy,et al.  Biomedical applications of diamond-like carbon coatings: a review. , 2007, Journal of biomedical materials research. Part B, Applied biomaterials.

[32]  Steven M. George,et al.  Ca test of Al2O3 gas diffusion barriers grown by atomic layer deposition on polymers , 2006 .

[33]  R.R. Harrison,et al.  Wireless Neural Recording With Single Low-Power Integrated Circuit , 2009, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[34]  S. Scott Neuroscience: Converting thoughts into action , 2006, Nature.

[35]  Stuart F Cogan,et al.  Electrical performance of penetrating microelectrodes chronically implanted in cat cortex , 2011, EMBC.

[36]  M. H. Tanielian,et al.  Evaluation and characterization of reliable non-hermetic conformal coatings for microelectromechanical system (MEMS) device encapsulation , 2000 .

[37]  F Solzbacher,et al.  A Wafer-Scale Etching Technique for High Aspect Ratio Implantable MEMS Structures. , 2010, Sensors and actuators. A, Physical.

[38]  M. Ghovanloo,et al.  Using Pulse Width Modulation for Wireless Transmission of Neural Signals in Multichannel Neural Recording Systems , 2009, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[39]  Daryl R. Kipke,et al.  Wireless implantable microsystems: high-density electronic interfaces to the nervous system , 2004, Proceedings of the IEEE.

[40]  Richard A. Normann,et al.  Selective stimulation and recording using a slanted multielectrode array , 1999, Proceedings of the First Joint BMES/EMBS Conference. 1999 IEEE Engineering in Medicine and Biology 21st Annual Conference and the 1999 Annual Fall Meeting of the Biomedical Engineering Society (Cat. N.

[41]  F. Solzbacher,et al.  Integrated wireless neural interface based on the Utah electrode array , 2009, Biomedical microdevices.

[42]  Florian Solzbacher,et al.  Influence of system integration and packaging for a wireless neural interface on its wireless powering performance , 2008, 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[43]  F. Solzbacher,et al.  Characterization of a-SiC(x):H thin films as an encapsulation material for integrated silicon based neural interface devices. , 2007, Thin Solid Films.

[44]  Richard A Normann,et al.  Technology Insight: future neuroprosthetic therapies for disorders of the nervous system , 2007, Nature Clinical Practice Neurology.

[45]  Stuart F Cogan,et al.  Plasma-enhanced chemical vapor deposited silicon carbide as an implantable dielectric coating. , 2003, Journal of biomedical materials research. Part A.

[46]  F. Solzbacher,et al.  Self-aligned tip deinsulation of atomic layer deposited Al2O3 and parylene C coated Utah electrode array based neural interfaces , 2014, Journal of micromechanics and microengineering : structures, devices, and systems.

[47]  G. Loeb,et al.  Parylene as a Chronically Stable, Reproducible Microelectrode Insulator , 1977, IEEE Transactions on Biomedical Engineering.

[48]  Patrick A Tresco,et al.  The brain tissue response to implanted silicon microelectrode arrays is increased when the device is tethered to the skull. , 2007, Journal of biomedical materials research. Part A.

[49]  J. J. Licari,et al.  Coating materials for electronic applications : polymers, processes, reliability, testing , 2003 .

[50]  John P. Donoghue,et al.  Connecting cortex to machines: recent advances in brain interfaces , 2002, Nature Neuroscience.

[51]  G. Gerlach,et al.  Smart Hydrogel-Based Biochemical Microsensor Array for Medical Diagnostics , 2012 .

[52]  Jerald D. Kralik,et al.  Real-time prediction of hand trajectory by ensembles of cortical neurons in primates , 2000, Nature.

[53]  Atomic layer deposited Al2O3 and parylene C dual-layer encapsulation for biomedical implantable devices , 2013, 2013 Transducers & Eurosensors XXVII: The 17th International Conference on Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS & EUROSENSORS XXVII).

[54]  Sharon Gerecht,et al.  Acellular implantable and injectable hydrogels for vascular regeneration , 2015, Biomedical materials.

[55]  Jerzy Kanicki,et al.  Thin-Film Encapsulation of Organic Light-Emitting Devices , 2007 .

[56]  Florian Solzbacher,et al.  Encapsulation of an Integrated Neural Interface Device With Parylene C , 2009, IEEE Transactions on Biomedical Engineering.

[57]  Michael Szwarc,et al.  POLY-PARA-XYLELENE: ITS CHEMISTRY AND APPLICATION IN COATING TECHNOLOGY. , 1976 .

[58]  Toh-Ming Lu,et al.  Chemical Vapor Deposition Polymerization: The Growth and Properties of Parylene Thin Films , 2003 .