Repositioning accuracy of a commercially available thermoplastic mask system.

BACKGROUND AND PURPOSE To evaluate the repositioning accuracy of a commercially available thermoplastic mask system for single dose radiosurgery treatments and fractionated treatment courses. PATIENTS AND METHODS The repositioning accuracy of the Raycast-HP mask system (Orfit Industries, Wijnegem, Belgium) was analyzed. Twenty-two patients that were treated by intensity-modulated radiation therapy (IMRT) or intensity modulated radiosurgery (IMRS) for 43 intracranial lesions, underwent repeated CT imaging during their course of treatment, or as a positional control immediately before radiosurgery. We evaluated multiple anatomical landmark coordinates and their respective shifts in consecutive repeated CT-controls. An iterative optimization algorithm allowed for the calculation of the x, y and z-components of translation of the target isocenter(s) for each repeated CT, as well as rotation in the respective CT data sets. In addition to absolute target isocenter translation, the total magnitude vector (i.e. sum-vector) of isocenter motion was calculated along with patient rotations about the three principle axes. RESULTS Fifty-five control CT datasets were analyzed for the target isocenter's respective position relative to the original treatment planning CT simulation. Mean target isocenter translation was 0.74+/-0.53, 0.75+/-0.60 and 0.93+/-0.78 mm in x, y and z-directions, respectively. Mean rotation about the x, y and z-axes was 0.67+/-0.66, 0.61+/-0.63 and 0.67+/-0.61 degrees, respectively. The respective median and mean magnitude vectors of isocenter translation were 1.28 and 1.59+/-0.84 mm. Analysis of the accuracy of the first setup control, representative of setup accuracy for radiosurgery treatments, compared with setup accuracy throughout a fractionated radiation treatment course were statistically equivalent (P= 0.15) thus indicating no measurable deterioration of setup accuracy throughout the treatment course. CONCLUSIONS The analyzed Orfit thermoplastic mask system performed favorably compared with other mask immobilization systems for which peer-reviewed repositioning data exist. While the performance of the system for fractionated treatment courses was considered to be excellent, use of this mask system for radiosurgery immobilization in our clinic is subject to additional quality assurance measures to prohibit the delivery of treatments with target dislocations larger than 2 mm. The measured data in the present study should enable the users of this system to assign appropriate margins for the generation of planning target volumes.

[1]  B. Salter,et al.  The TALON removable head frame system for stereotactic radiosurgery/radiotherapy: measurement of the repositioning accuracy. , 2001, International journal of radiation oncology, biology, physics.

[2]  V Grégoire,et al.  Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors. , 2001, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[3]  K. Bratengeier,et al.  CT simulation in stereotactic brain radiotherapy--analysis of isocenter reproducibility with mask fixation. , 1997, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[4]  S L Meeks,et al.  Initial clinical experience with frameless stereotactic radiosurgery: analysis of accuracy and feasibility. , 2001, International journal of radiation oncology, biology, physics.

[5]  F. Lohr,et al.  Radiosurgery alone or in combination with whole-brain radiotherapy for brain metastases. , 1998, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[6]  O Jäkel,et al.  Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system. , 2001, International journal of radiation oncology, biology, physics.

[7]  R Bale,et al.  Repositioning accuracy: comparison of a noninvasive head holder with thermoplastic mask for fractionated radiotherapy and a case report. , 1998, International journal of radiation oncology, biology, physics.

[8]  C A Pelizzari,et al.  Repositioning accuracy of a noninvasive head fixation system for stereotactic radiotherapy. , 1996, Medical physics.

[9]  W Schlegel,et al.  Photogrammetric accuracy measurements of head holder systems used for fractionated radiotherapy. , 1994, International journal of radiation oncology, biology, physics.

[10]  J. Tsai,et al.  A non-invasive immobilization system and related quality assurance for dynamic intensity modulated radiation therapy of intracranial and head and neck disease. , 1999, International journal of radiation oncology, biology, physics.