Estimating the Magnitude and Phase Response of a 50 GHz Sampling Oscilloscope Using the "Nose-to-Nose" Method

We describe estimation of the magnitude and phase response of a sampling oscilloscope with 50 GHz bandwidth using the nose-to-nose method. The measurements are corrected for the non-ideal properties of the oscilloscope and calibration apparatus, including mismatch and time-base distortion, drift, and jitter. The mean and standard deviation of repeated measurements of an ensemble of three oscilloscope samplers are reported, along with attempts to verify the magnitude calibration using a swept sine-wave method.

[1]  Paul D. Hale,et al.  Least-squares estimation of time-base distortion of sampling oscilloscopes , 1999, IEEE Trans. Instrum. Meas..

[2]  W. Gans The Measurement and Deconvolution of Time Jitter in Equivalent-Time Waveform Samplers , 1983, IEEE Transactions on Instrumentation and Measurement.

[3]  Jan Verspecht,et al.  Analysis of Interconnection Networks and Mismatch in the Nose-to-Nose Calibration , 2000, 55th ARFTG Conference Digest.

[4]  J. Verspecht Broadband sampling oscilloscope characterization with the "nose-to-nose" calibration procedure: a theoretical analysis , 1994, Conference Proceedings. 10th Anniversary. IMTC/94. Advanced Technologies in I & M. 1994 IEEE Instrumentation and Measurement Technolgy Conference (Cat. No.94CH3424-9).

[5]  John R. Juroshek,et al.  A Direct Calibration Method for Measuring Equivalent Source Mismatch | NIST , 1997 .

[6]  Kate A. Remley,et al.  Nose-to-Nose Response of a 20-GHz Sampling Circuit , 2000, 54th ARFTG Conference Digest.

[7]  Kate A. Remley,et al.  Realistic sampling-circuit model for a nose-to-nose simulation , 2000, 2000 IEEE MTT-S International Microwave Symposium Digest (Cat. No.00CH37017).

[8]  Yves Rolain,et al.  Signal reconstruction for nonequidistant finite length sample sets: a "KIS" approach , 1998, IMTC/98 Conference Proceedings. IEEE Instrumentation and Measurement Technology Conference. Where Instrumentation is Going (Cat. No.98CH36222).

[9]  J. Verspecht,et al.  Individual characterization of broadband sampling oscilloscopes with a 'nose-to-nose' calibration procedure , 1993, 1993 IEEE Instrumentation and Measurement Technology Conference.