End-expiration respiratory gating for a high-resolution stationary cardiac SPECT system

Respiratory and cardiac motions can degrade myocardial perfusion SPECT (MPS) image quality and reduce defect detection and quantitative accuracy. In this study, we developed a dual respiratory and cardiac gating system for a high-resolution fully stationary cardiac SPECT scanner in order to improve the image quality and defect detection. Respiratory motion was monitored using a compressive sensor pillow connected to a dual respiratory-cardiac gating box, which sends cardiac triggers only during end-expiration phases to the single cardiac trigger input on the SPECT scanners. The listmode data were rebinned retrospectively into end-expiration frames for respiratory motion reduction or eight cardiac gates only during end-expiration phases to compensate for both respiratory and cardiac motions. The proposed method was first validated on a motion phantom in the presence and absence of multiple perfusion defects, and then applied on 11 patient studies with and without perfusion defects. In the normal phantom studies, the end-expiration gated SPECT (EXG-SPECT) reduced respiratory motion blur and increased myocardium to blood pool contrast by 51.2% as compared to the ungated images. The proposed method also yielded an average of 11.2% increase in myocardium to defect contrast as compared to the ungated images in the phantom studies with perfusion defects. In the patient studies, EXG-SPECT significantly improved the myocardium to blood pool contrast (p < 0.005) by 24% on average as compared to the ungated images, and led to improved perfusion uniformity across segments on polar maps for normal patients. For a patient with defect, EXG-SPECT improved the defect contrast and definition. The dual respiratory-cardiac gating further reduced the blurring effect, increased the myocardium to blood pool contrast significantly by 36% (p < 0.05) compared to EXG-SPECT, and further improved defect characteristics and visualization of fine structures at the expense of increased noise on the patient with defect. The results showed that the proposed methods can effectively reduce motion blur in the images caused by both respiratory and cardiac motions, which may lead to more accurate defect detection and quantifications. This approach can be easily adapted in routine clinical practice on currently available commercial systems.

[1]  T. Ruddy,et al.  Single-Phase CT Aligned to Gated PET for Respiratory Motion Correction in Cardiac PET/CT , 2010, The Journal of Nuclear Medicine.

[2]  Paul E Kinahan,et al.  The impact of respiratory motion on tumor quantification and delineation in static PET/CT imaging , 2009, Physics in medicine and biology.

[3]  David Atkinson,et al.  A study of the motion and deformation of the heart due to respiration , 2002, IEEE Transactions on Medical Imaging.

[4]  Guido Germano,et al.  Advances in technical aspects of myocardial perfusion SPECT imaging , 2009, Journal of Nuclear Cardiology.

[5]  Respiratory gating for a stationary dedicated cardiac SPECT system , 2011, 2011 IEEE Nuclear Science Symposium Conference Record.

[6]  Xiao Jin,et al.  Event-by-event respiratory motion correction for PET with 3D internal-1D external motion correlation. , 2013, Medical physics.

[7]  Arik Wolak,et al.  Motion-Frozen Myocardial Perfusion SPECT Improves Detection of Coronary Artery Disease in Obese Patients , 2008, Journal of Nuclear Medicine.

[8]  O Caselles,et al.  The retrospective binning method improves the consistency of phase binning in respiratory-gated PET/CT , 2012, Physics in medicine and biology.

[9]  A. Sinusas,et al.  Potential impact of hybrid CZT SPECT/CT imaging on estimation accuracy of left ventricular volumes and ejection fraction: A phantom study , 2013, 2013 IEEE Nuclear Science Symposium and Medical Imaging Conference (2013 NSS/MIC).

[10]  B. Tsui,et al.  Evaluation of Respiratory Motion Effect on Defect Detection in Myocardial Perfusion SPECT: A Simulation Study , 2009, IEEE Transactions on Nuclear Science.

[11]  G. Lucignani Respiratory and cardiac motion correction with 4D PET imaging: shooting at moving targets , 2009, European Journal of Nuclear Medicine and Molecular Imaging.

[12]  Xiaoyi Jiang,et al.  Motion Correction in Dual Gated Cardiac PET Using Mass-Preserving Image Registration , 2012, IEEE Transactions on Medical Imaging.

[13]  O. Schober,et al.  List Mode–Driven Cardiac and Respiratory Gating in PET , 2009, Journal of Nuclear Medicine.

[14]  P. Green Bayesian reconstructions from emission tomography data using a modified EM algorithm. , 1990, IEEE transactions on medical imaging.

[15]  Joyoni Dey,et al.  Estimation and correction of cardiac respiratory motion in SPECT in the presence of limited-angle effects due to irregular respiration. , 2010, Medical physics.

[16]  A J Sinusas,et al.  Quantification of SPECT myocardial perfusion images: Methodology and validation of the Yale-CQ method , 1999, Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology.

[17]  I. Burger,et al.  Real-time breath-hold triggering of myocardial perfusion imaging with a novel cadmium-zinc-telluride detector gamma camera , 2010, European Journal of Nuclear Medicine and Molecular Imaging.

[18]  Haim Azhari,et al.  Correction of Heart Motion Due to Respiration in Clinical Myocardial Perfusion SPECT Scans Using Respiratory Gating , 2007, Journal of Nuclear Medicine.

[19]  H. Azhari,et al.  Correction for respiration artefacts in myocardial perfusion SPECT is more effective when reconstructions supporting collimator detector response compensation are applied , 2009, Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology.

[20]  Kris Thielemans,et al.  Quiescent period respiratory gating for PET/CT. , 2010, Medical physics.

[21]  L. Barnden,et al.  Contributions of subdiaphragmatic activity, attenuation, and diaphragmatic motion to inferior wall artifact in attenuation-corrected Tc-99m myocardial perfusion SPECT , 2005, Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology.

[22]  R. deKemp,et al.  Respiratory phase alignment improves blood-flow quantification in Rb82 PET myocardial perfusion imaging. , 2013, Medical physics.

[23]  Michael A. King,et al.  Comparison of methods of acquiring attenuation maps for cardiac SPECT in the presence of respiratory motion , 2013, Journal of Nuclear Cardiology.

[24]  Valerie Treyer,et al.  Nuclear Myocardial Perfusion Imaging with a Cadmium-Zinc-Telluride Detector Technique: Optimized Protocol for Scan Time Reduction , 2010, Journal of Nuclear Medicine.

[25]  Haim Azhari,et al.  Dual “motion-frozen heart” combining respiration and contraction compensation in clinical myocardial perfusion SPECT imaging , 2009, Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology.

[26]  Xiaoyi Jiang,et al.  Lung motion correction on respiratory gated 3-D PET/CT images , 2006, IEEE Transactions on Medical Imaging.

[27]  Xiaoyi Jiang,et al.  Respiratory Motion Correction in 3-D PET Data With Advanced Optical Flow Algorithms , 2008, IEEE Transactions on Medical Imaging.

[28]  Tommi Noponen,et al.  Dual-gated cardiac PET–Clinical feasibility study , 2010, European Journal of Nuclear Medicine and Molecular Imaging.

[29]  Joyoni Dey,et al.  Improved image reconstruction for dedicated cardiac SPECT with truncated projections , 2012 .

[30]  Vladimir Tolmachev,et al.  Imaging of Human Epidermal Growth Factor Receptor Type 2 Expression with 18F-Labeled Affibody Molecule ZHER2:2395 in a Mouse Model for Ovarian Cancer , 2012, The Journal of Nuclear Medicine.

[31]  Leonid Tsukerman,et al.  A fast cardiac gamma camera with dynamic SPECT capabilities: design, system validation and future potential , 2010, European Journal of Nuclear Medicine and Molecular Imaging.

[32]  Nassir Navab,et al.  Dual cardiac–respiratory gated PET: implementation and results from a feasibility study , 2007, European Journal of Nuclear Medicine and Molecular Imaging.

[33]  Guido Germano,et al.  "Motion-frozen" display and quantification of myocardial perfusion. , 2004, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[34]  Sue Francis,et al.  Physiological measurements using ultra-high field fMRI: a review , 2014, Physiological measurement.

[35]  Jason G Parker,et al.  Respiratory motion correction in gated cardiac SPECT using quaternion-based, rigid-body registration. , 2009, Medical physics.