An expandable catheter loop coil for intravascular MRI in larger blood vessels

The present study proposes a catheter system with an expandable coil etched on a polyimide foil. The catheter system combines the advantages of a small insertion diameter when the coil is rolled up in a protective carrier sheath with an increased signal‐to‐noise ratio (SNR) and penetration depth when the coil is pushed out. After imaging, the coil can be retracted into the sheath and folded back into the initial rolled‐up configuration due to the tapered geometry of the carrier foil. The catheter system was tested on two healthy anesthetized pigs, including tracking and high‐resolution intravascular imaging. To reduce artifacts in high‐resolution images induced by catheter motion in the pulsatile blood flow, a motion‐gating method was implemented that combines a flow‐compensated two‐dimensional fast low angle shot (FLASH) imaging sequence with the acquisition of projection data for retrospective gating. Using the projection data for motion detection, image SNR was increased by up to 500% over uncorrected images, and anatomic structures of 150 μm size could be differentiated in the aorta. Magn Reson Med, 2010. © 2009 Wiley‐Liss, Inc.

[1]  S. Souza,et al.  Real‐time position monitoring of invasive devices using magnetic resonance , 1993, Magnetic resonance in medicine.

[2]  V. Fuster,et al.  Characterization of Atherosclerotic Plaques by Magnetic Resonance Imaging , 2000, Annals of the New York Academy of Sciences.

[3]  Wolfhard Semmler,et al.  MR‐guided intravascular procedures: Real‐time parameter control and automated slice positioning with active tracking coils , 2004, Journal of magnetic resonance imaging : JMRI.

[4]  R Frayne,et al.  A rapid 2D time‐resolved variable‐rate k‐space sampling MR technique for passive catheter tracking during endovascular procedures , 1998, Magnetic resonance in medicine.

[5]  Jeffrey L Duerk,et al.  Toward rapid high resolution in vivo intravascular MRI: Evaluation of vessel wall conspicuity in a porcine model using multiple imaging protocols , 2006, Journal of magnetic resonance imaging : JMRI.

[6]  B K Rutt,et al.  An expandable intravenous RF coil for arterial wall imaging , 1998, Journal of magnetic resonance imaging : JMRI.

[7]  D B Plewes,et al.  MR imaging of blood vessels with an intravascular coil , 1992, Journal of magnetic resonance imaging : JMRI.

[8]  J L Duerk,et al.  Intravascular (catheter) NMR receiver probe: Preliminary design analysis and application to canine iliofemoral imaging , 1992, Magnetic resonance in medicine.

[9]  Volkmar Schulz,et al.  Transmission line for improved RF safety of interventional devices , 2005, Magnetic resonance in medicine.

[10]  C J Bakker,et al.  Heating Around Intravascular Guidewires by Resonating RF Waves , 2000, Journal of magnetic resonance imaging : JMRI.

[11]  Wolfhard Semmler,et al.  B1 field-insensitive transformers for RF-safe transmission lines , 2006, Magnetic Resonance Materials in Physics, Biology and Medicine.

[12]  René M. Botnar,et al.  Motion artifact reduction and vessel enhancement for free‐breathing navigator‐gated coronary MRA using 3D k‐space reordering , 2001, Magnetic resonance in medicine.

[13]  Arne Reykowski,et al.  Active device tracking and high‐resolution intravascular MRI using a novel catheter‐based, opposed‐solenoid phased array coil , 2004, Magnetic resonance in medicine.

[14]  J. Debatin,et al.  Single‐loop coil concepts for intravascular magnetic resonance imaging , 1999, Magnetic resonance in medicine.

[15]  W. Nitz,et al.  On the heating of linear conductive structures as guide wires and catheters in interventional MRI , 2001, Journal of magnetic resonance imaging : JMRI.

[16]  D. I. Hoult,et al.  The NMR receiver: A description and analysis of design , 1978 .

[17]  Peter Kellman,et al.  Image reconstruction in SNR units: A general method for SNR measurement † , 2005, Magnetic resonance in medicine.