Detection of electrode interchange in right precordial and posterior ECG leads

This study presents a method for automated detection of misplaced supplementary precordial leads, including the right-sided V3R, V4R and the posterior V8, V9 leads. Considering their uncommon use in clinical routine, a lead reversal is quite probable and could result in erroneous diagnosis and treatment. The method allows real-time implementation by scoring inter-lead cross-correlations over continuous 4s episodes, scanning the normal progression of PQRST waveforms within leads [V4R, V3R, V3, V4] and [V4, V5, V6, V8, V9]. A large 16-lead ECG database with 1333 chest pain patients is used to test the performance of the method for all possible 23 swaps between the supplementary leads V4R, V3R, V8, V9, assuming correct positions of the standard V1-V6. The sensitivity (Se) for lead reversals is Se=94.1±4.6%, ranged between 78.5% and 97.8%, with the most difficult detection of V3R/V4R swap (Se=78.5%), V4R/V9 swap (Se=83.7%), V8/V9 swap (Se=91.8%). The achieved specificity for the correct lead positions is Sp=83.4%.

[1]  Ferenc Follath,et al.  Electrocardiographic artifacts due to electrode misplacement and their frequency in different clinical settings. , 2007, The American journal of emergency medicine.

[2]  Xiaopeng Zhao,et al.  Automatic detection of ECG electrode misplacement: a tale of two algorithms , 2012, Physiological measurement.

[3]  I. Hoffman,et al.  A flatline electrocardiogram in lead II is a marker for right arm/right leg electrode switch. , 2007, Journal of electrocardiology.

[4]  K. C. Ho,et al.  Use of the sinus P wave in diagnosing electrocardiographic limb lead misplacement not involving the right leg (ground) lead. , 2001, Journal of electrocardiology.

[5]  Saeed Babaeizadeh,et al.  Automatic detection of ECG cable interchange by analyzing both morphology and interlead relations. , 2014, Journal of electrocardiology.

[6]  Irena Jekova,et al.  Recognition of diagnostically useful ECG recordings: Alert for corrupted or interchanged leads , 2011, 2011 Computing in Cardiology.

[7]  O. Pahlm,et al.  Detection of frequently overlooked electrocardiographic lead reversals using artificial neural networks. , 1996, The American journal of cardiology.

[8]  J A Kors,et al.  Accurate automatic detection of electrode interchange in the electrocardiogram. , 2001, The American journal of cardiology.

[9]  J A Milliken,et al.  Recognition of electrocardiographic left arm/left leg lead reversal. , 1997, The American journal of cardiology.

[10]  Irena Jekova,et al.  Detection of electrode interchange in precordial and orthogonal ECG leads , 2013, Computing in Cardiology 2013.

[11]  Reginald T Ho,et al.  Simple Diagnosis of Limb‐Lead Reversals by Predictable Changes in QRS Axis , 2006, Pacing and clinical electrophysiology : PACE.