Handheld scanning probes for optical coherence tomography: developments, applications, and perspectives

We present the handheld scanning probes that we have recently developed in our current project for biomedical imaging in general and for Optical Coherence Tomography (OCT) in particular. OCT is an established, but dynamic imagistic technique based on laser interferometry, which offers micrometer resolutions and millimeters penetration depths. With regard to existing devices, the newly developed handheld probes are simple, light and relatively low cost. Their design is described in detail to allow for the reproduction in any lab, including for educational purposes. Two probes are constructed almost entirely from off-the-shelf components, while a third, final variant is constructed with dedicated components, in an ergonomic design. The handheld probes have uni-dimensional (1D) galvanometer scanners therefore they achieve transversal sections through the biological sample investigated - in contrast to handheld probes equipped with bi-dimensional (2D) scanners that can also achieve volumetric (3D) reconstructions of the samples. These latter handheld probes are therefore also discussed, as well as the possibility to equip them with galvanometer 2D scanners or with Risley prisms. For galvanometer scanners the optimal scanning functions studied in a series of previous works are pointed out; these functions offer a higher temporal efficiency/duty cycle of the scanning process, as well as artifact-free OCT images. The testing of the handheld scanning probes in dental applications is presented, for metal ceramic prosthesis and for teeth.

[1]  Jan Awrejcewicz Modeling, Simulation and Control of Nonlinear Engineering Dynamical Systems , 2009 .

[2]  Jeehyun Kim,et al.  Handheld Optical Coherence Tomography Scanner for Primary Care Diagnostics , 2011, IEEE Transactions on Biomedical Engineering.

[3]  Freddy T. Nguyen,et al.  Optical coherence tomography: a review of clinical development from bench to bedside. , 2007, Journal of biomedical optics.

[4]  N. S. Kapany,et al.  Laser Retinal Photocoagulator , 1965 .

[5]  Jean Montagu Scanners: Galvanometric and Resonant , 2015 .

[6]  Redmond P. Aylward,et al.  Advanced galvanometer‐based optical scanner design , 2003 .

[7]  Adrian Bradu,et al.  Surface imaging of metallic material fractures using optical coherence tomography. , 2014, Applied optics.

[8]  G. Gelikonov,et al.  In vivo OCT imaging of hard and soft tissue of the oral cavity. , 1998, Optics express.

[9]  J P Rolland,et al.  Invariant resolution dynamic focus OCM based on liquid crystal lens. , 2007, Optics express.

[10]  Kye-Sung Lee,et al.  Three-dimensional imaging of normal skin and nonmelanoma skin cancer with cellular resolution using Gabor domain optical coherence microscopy , 2012, Journal of biomedical optics.

[11]  Virgil-Florin Duma,et al.  Optimal scanning function of a galvanometer scanner for an increased duty cycle , 2010 .

[12]  Susana Marcos,et al.  Full OCT anterior segment biometry: an application in cataract surgery , 2013, Biomedical optics express.

[13]  Virgil-Florin Duma,et al.  Optimization of scanning and command functions of galvanometer-based scanners , 2011, Optical Metrology.

[14]  Adrian Bradu,et al.  Noninvasive Quantitative Evaluation of the Dentin Layer during Dental Procedures Using Optical Coherence Tomography , 2015, Comput. Math. Methods Medicine.

[15]  Martin F. Kraus,et al.  Handheld ultrahigh speed swept source optical coherence tomography instrument using a MEMS scanning mirror. , 2013, Biomedical optics express.

[16]  Adrian Gh. Podoleanu,et al.  Perspectives of optical scanning in OCT , 2010, BiOS.

[17]  J. S. Gadhok,et al.  Achieving high-duty cycle sawtooth scanning with galvanometric scanners , 1999, Optics & Photonics.

[18]  B E Bouma,et al.  Forward-imaging instruments for optical coherence tomography. , 1997, Optics letters.

[19]  Cosmin Sinescu,et al.  Optical Imaging of Oral Squamous Cell Carcinoma Using Optical Coherence Tomography and Micro CT , 2013 .

[20]  Adrian Gh. Podoleanu,et al.  Research Lead Student Projects on Multi-Disciplinary Optomechatronics with Applications in Biomedical Imaging , 2014 .

[21]  Virgil-Florin Duma,et al.  Experimental investigations of the scanning functions of galvanometer-based scanners with applications in OCT. , 2011, Applied optics.

[22]  Gerald F. Marshall,et al.  Handbook of Optical and Laser Scanning , 2004 .

[23]  Adrian Bradu,et al.  Assessment of the sealant/tooth interface using optical coherence tomography , 2015 .

[24]  Virgil-Florin Duma,et al.  Building an optomechatronics group in a young university in Western Romania , 2014, Other Conferences.

[25]  L. Yannuzzi,et al.  Combined multiplanar optical coherence tomography and confocal scanning ophthalmoscopy. , 2004, Journal of biomedical optics.

[26]  Johannes F de Boer,et al.  Angiography of the retina and the choroid with phase-resolved OCT using interval-optimized backstitched B-scans. , 2012, Optics express.

[27]  Gerald F. Marshall,et al.  Risley prism scan patterns , 1999, Optics & Photonics.

[28]  Adrian Bradu,et al.  Dual instrument for in vivo and ex vivo OCT imaging in an ENT department , 2012, Biomedical optics express.

[29]  J. Rolland,et al.  Optimization of galvanometer scanning for optical coherence tomography. , 2015, Applied Optics.

[30]  Changhuei Yang,et al.  Sensitivity advantage of swept source and Fourier domain optical coherence tomography. , 2003, Optics express.

[31]  Virgil-Florin Duma Scanning in biomedical imaging: from classical devices to handheld heads and micro-systems , 2014, International Conference on Lasers in Medicine.

[32]  Virgil-Florin Duma,et al.  Mathematical Functions of a 2-D Scanner with Oscillating Elements , 2009 .

[33]  G. Ripandelli,et al.  Optical coherence tomography. , 1998, Seminars in ophthalmology.

[34]  Wolfgang Drexler,et al.  State-of-the-art retinal optical coherence tomography , 2008, Progress in Retinal and Eye Research.

[35]  Alberto J. Palma,et al.  Recent developments in handheld and portable optosensing-a review. , 2011, Analytica chimica acta.

[36]  Changhuei Yang,et al.  Handheld forward-imaging needle endoscope for ophthalmic optical coherence tomography inspection. , 2008, Journal of biomedical optics.

[37]  A. Godavarty,et al.  Hand-held based near-infrared optical imaging devices: a review. , 2009, Medical engineering & physics.

[38]  Liren Liu,et al.  Laser coarse-fine coupling scanning method by steering double prisms. , 2012, Applied optics.

[39]  Yajun Li Single-mirror beam steering system: analysis and synthesis of high-order conic-section scan patterns. , 2008, Applied optics.

[40]  J. Fujimoto,et al.  Optical Coherence Tomography , 1991 .

[41]  Brett E Bouma,et al.  Two-axis magnetically-driven MEMS scanning catheter for endoscopic high-speed optical coherence tomography. , 2007, Optics express.

[42]  Virgil-Florin Duma,et al.  Modeling of Risley prisms devices for exact scan patterns , 2013, Optical Metrology.

[43]  Panomsak Meemon,et al.  Gabor-based fusion technique for Optical Coherence Microscopy. , 2010, Optics express.

[44]  Adrian Bradu,et al.  Design and testing of prototype handheld scanning probes for optical coherence tomography , 2014, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.