Large area pixel detector WIDEPIX with full area sensitivity composed of 100 Timepix assemblies with edgeless sensors

The superior properties of the single particle counting semiconductor pixel detectors in radiation imaging are well known. They are namely: very high dynamic range due to digital counting, absence of integration and read-out noise, high spatial resolution and energy sensitivity. The major disadvantage of current pixel devices preventing their broad exploitation has been their relatively small sensitive area of few cm2. This disadvantage is often solved using tiling method placing many detector units side by side forming a large matrix. The current tiling techniques require rather large gaps of few millimeters between tiles. These gaps stand as areas insensitive to radiation which is acceptable only in some applications such as diffraction imaging. However standard transmission radiography requires fully continuous area sensitivity. In this article we present the new large area device WIDEPIX composed of a matrix of 10 × 10 tiles of silicon pixel detectors Timepix (each of 256 × 256 pixels with pitch of 55 μm) having fully sensitive area of 14.3 × 14.3 cm2 without any gaps between the tiles. The device contains a total of 6.5 mega pixels. This achievement was reached thanks to new technology of edgeless semiconductor sensors together with precise alignment technique and multilevel architecture of readout electronics. The mechanical construction of the device is fully modular and scalable. This concept allows replacing any single detector tile which significantly improves production yield. The first results in the field of X-ray radiography and material sensitive X-ray radiography are presented in this article.

[1]  Energy selective X-ray imaging with Medipix , 2010, IEEE Nuclear Science Symposuim & Medical Imaging Conference.

[3]  K. Smith,et al.  Semiconductor pixel detectors , 2000 .

[4]  Juha Kalliopuska,et al.  Alternative fabrication process for edgeless detectors on 6 in. wafers , 2011 .

[5]  J. Jakubek,et al.  Equalization of Medipix2 imaging detector energy thresholds using measurement of polychromatic X-ray beam attenuation , 2011 .

[6]  Jan Jakubek,et al.  Data processing and image reconstruction methods for pixel detectors , 2007 .

[7]  B. Schmitt,et al.  EIGER: Next generation single photon counting detector for X-ray applications , 2011 .

[8]  J. Jakubek,et al.  X-ray beam hardening based material recognition in micro-imaging , 2011 .

[9]  Michael Campbell,et al.  Medipix3 CT for material sciences , 2013 .

[10]  Rafael Ballabriga,et al.  Timepix, a 65k programmable pixel readout chip for arrival time, energy and/or photon counting measurements , 2007 .

[11]  Vaclav Kraus,et al.  FITPix ? fast interface for Timepix pixel detectors , 2011 .

[12]  J. Kalliopuska,et al.  Processing and characterization of edgeless radiation detectors for large area detection , 2013 .

[13]  Daniel Turecek,et al.  Pixelman: a multi-platform data acquisition and processing software package for Medipix2, Timepix and Medipix3 detectors , 2011 .

[14]  B. Schmitt,et al.  Performance of single-photon-counting PILATUS detector modules , 2009, Journal of synchrotron radiation.

[15]  P. Pangaud,et al.  Performance and Applications of the CdTe- and Si-XPAD3 photon counting 2D detector , 2011 .

[16]  J. Jakubek,et al.  Semiconductor Pixel detectors and their applications in life sciences , 2009 .