Accuracy of 4 digital scanning systems on prepared teeth digitally isolated from a complete dental arch

Statement of problem. The accuracy of digital scanners is acceptable for scanning a complete dental arch. However, whether that accuracy is sufficient for only 1 tooth within the dental scan of a complete dental arch is unclear. Purpose. The purpose of this in vitro study was to evaluate and compare the accuracy of 4 intraoral scanners on a complete dental arch and on prepared teeth digitally isolated from the digital scan in terms of trueness and precision. Material and methods. A model of a complete dental arch with tooth preparations was scanned 40 times with each of the 4 digital scanners. Their accuracy was evaluated by using 3‐dimensional (3D) software to compare the test models with a highly accurate reference model. The data were digitally processed to isolate the prepared teeth and evaluate them in the same way. The data were statistically analyzed using the Levene test and the Tamhane's T2 test (&agr;=.05). Results. In scans of a complete dental arch, the True Definition scanner had the best accuracy values, followed by TRIOS, iTero, and Omnicam. For prepared teeth isolated from the dental arch, both True Definition and TRIOS had the best values, followed by iTero and Omnicam. Conclusions. In both long‐span scans of the complete dental arch and isolated prepared teeth, the True Definition scanner had the greatest accuracy, closely followed by TRIOS.

[1]  Mahmoud Serag,et al.  A Comparative Study of the Accuracy of Dies Made from Digital Intraoral Scanning vs. Elastic Impressions: An In Vitro Study , 2018, Journal of prosthodontics : official journal of the American College of Prosthodontists.

[2]  R. Nedelcu,et al.  Scanning accuracy and precision in 4 intraoral scanners: an in vitro comparison based on 3-dimensional analysis. , 2014, The Journal of prosthetic dentistry.

[3]  In-Sung Yeo,et al.  In vitro marginal fit of three all-ceramic crown systems. , 2003, The Journal of prosthetic dentistry.

[4]  Albert Mehl,et al.  Precision of guided scanning procedures for full-arch digital impressions in vivo , 2017, Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie.

[5]  Accuracy of digital and conventional impression techniques and workflow , 2013, BDJ.

[6]  Stefan Wolfart,et al.  The clinical accuracy of single crowns exclusively fabricated by digital workflow—the comparison of two systems , 2013, Clinical Oral Investigations.

[7]  Henrik Skjerven,et al.  Marginal and Internal Fit of Cobalt-Chromium Fixed Dental Prostheses Generated from Digital and Conventional Impressions , 2014, International journal of dentistry.

[8]  A. Manabe,et al.  Comparison of digital intraoral scanners by single-image capture system and full-color movie system. , 2017, Bio-medical materials and engineering.

[9]  Lapo Governi,et al.  A Comparative Analysis Of Intraoral 3d Digital Scanners For Restorative Dentistry , 2008 .

[10]  Takashi Miyazaki,et al.  Effect of sintering on the marginal and internal fit of CAD/CAM-fabricated zirconia frameworks. , 2007, Dental materials journal.

[11]  Agustín Pascual-Moscardó,et al.  Relationship between resolution and accuracy of four intraoral scanners in complete-arch impressions , 2018, Journal of clinical and experimental dentistry.

[12]  G. Heydecke,et al.  Evaluation of the marginal fit of single-unit, complete-coverage ceramic restorations fabricated after digital and conventional impressions: A systematic review and meta-analysis. , 2016, The Journal of prosthetic dentistry.

[13]  J. Strub,et al.  Evaluation of the marginal accuracy of different all-ceramic crown systems after simulation in the artificial mouth. , 1999, Journal of oral rehabilitation.

[14]  Jack L Ferracane,et al.  Evaluation of different methods of optical impression making on the marginal gap of onlays created with CEREC 3D. , 2010, Operative dentistry.

[15]  L Tapie,et al.  Understanding dental CAD/CAM for restorations--the digital workflow from a mechanical engineering viewpoint. , 2015, International journal of computerized dentistry.

[16]  Agneta Oden,et al.  A three-dimensional evaluation of a laser scanner and a touch-probe scanner. , 2006, The Journal of prosthetic dentistry.

[17]  Djordje Vukelic,et al.  Comparative analysis on measuring performances of dental intraoral and extraoral optical 3D digitization systems , 2014 .

[18]  Sergio Rubinstein Accurate impression technique: a key to successful prosthetics. , 2010, Dentistry today.

[19]  Sebastian B. M. Patzelt,et al.  Accuracy of full-arch scans using intraoral scanners , 2014, Clinical Oral Investigations.

[20]  Mark Ludlow,et al.  Evaluation of the accuracy of 7 digital scanners: An in vitro analysis based on 3‐dimensional comparisons , 2017, The Journal of prosthetic dentistry.

[21]  Y. Chun,et al.  A comparison of the precision of three-dimensional images acquired by 2 digital intraoral scanners: effects of tooth irregularity and scanning direction , 2016, Korean journal of orthodontics.

[22]  T. Attin,et al.  In vivo precision of conventional and digital methods of obtaining complete-arch dental impressions. , 2016, The Journal of prosthetic dentistry.

[23]  A Ender,et al.  Influence of scanning strategies on the accuracy of digital intraoral scanning systems. , 2013, International journal of computerized dentistry.

[24]  Jin-Hun Jeon,et al.  Accuracy of complete-arch model using an intraoral video scanner: An in vitro study. , 2016, The Journal of prosthetic dentistry.

[25]  Ariel J Raigrodski,et al.  Contemporary materials and technologies for all-ceramic fixed partial dentures: a review of the literature. , 2004, The Journal of prosthetic dentistry.

[26]  M. Dehurtevent,et al.  Influence of dentist experience with scan spray systems used in direct CAD/CAM impressions. , 2015, The Journal of prosthetic dentistry.

[27]  Benjamin M. Wu,et al.  Effects of Computer-Aided Manufacturing Technology on Precision of Clinical Metal-Free Restorations , 2015, BioMed research international.

[28]  T. Jemt,et al.  Accuracy of implant-supported prostheses in the edentulous jaw: analysis of precision of fit between cast gold-alloy frameworks and master casts by means of a three-dimensional photogrammetric technique. , 1995, Clinical oral implants research.

[29]  Florian Beuer,et al.  Precision of fit: zirconia three-unit fixed dental prostheses , 2009, Clinical Oral Investigations.

[30]  Andreas Syrek,et al.  Clinical evaluation of all-ceramic crowns fabricated from intraoral digital impressions based on the principle of active wavefront sampling. , 2010, Journal of dentistry.

[31]  J. Katsoulis,et al.  Impact of digital intraoral scan strategies on the impression accuracy using the TRIOS Pod scanner. , 2016, Quintessence international.

[32]  Hiroshi Hirayama,et al.  Marginal and internal adaptation of ceramic crown restorations fabricated with CAD/CAM technology and the heat-press technique. , 2014, The Journal of prosthetic dentistry.

[33]  P Pera,et al.  In vitro marginal adaptation of alumina porcelain ceramic crowns. , 1994, The Journal of prosthetic dentistry.

[34]  J. Mclean,et al.  The estimation of cement film thickness by an in vivo technique , 1971, British Dental Journal.

[35]  Sebastian Quaas,et al.  Direct mechanical data acquisition of dental impressions for the manufacturing of CAD/CAM restorations. , 2007, Journal of dentistry.

[36]  P. Vallittu,et al.  Digital Versus Conventional Impressions in Fixed Prosthodontics: A Review , 2018, Journal of prosthodontics : official journal of the American College of Prosthodontists.

[37]  J R Strub,et al.  Marginal fit of restorations before and after cementation in vivo. , 1993, The International journal of prosthodontics.

[38]  Göran Sjögren,et al.  Accuracy of ceramic restorations made using an in-office optical scanning technique: an in vitro study. , 2014, Operative dentistry.

[39]  Federico Mandelli,et al.  Evaluation of the accuracy of extraoral laboratory scanners with a single-tooth abutment model: A 3D analysis. , 2017, Journal of prosthodontic research.

[40]  Josef Schweiger,et al.  A new method for the evaluation of the accuracy of full-arch digital impressions in vitro , 2015, Clinical Oral Investigations.

[41]  B. Duane,et al.  Evidence for use of intraoral scanners under clinical conditions for obtaining full-arch digital impressions is insufficient , 2017, Evidence-Based Dentistry.

[42]  Matts Andersson,et al.  Digitization of simulated clinical dental impressions: virtual three-dimensional analysis of exactness. , 2009, Dental materials : official publication of the Academy of Dental Materials.

[43]  A. Mehl,et al.  In-vitro evaluation of the accuracy of conventional and digital methods of obtaining full-arch dental impressions. , 2015, Quintessence international.