Time-of-flight neutron rejection to improve prompt gamma imaging for proton range verification: a simulation study

Therapeutic proton and heavier ion beams generate prompt gamma photons that may escape from the patient. In principle, this allows for real-time, in situ monitoring of the treatment delivery, in particular, the hadron range within the patient, by imaging the emitted prompt gamma rays. Unfortunately, the neutrons simultaneously created with the prompt photons create a background that may obscure the prompt gamma signal. To enhance the accuracy of proton dose verification by prompt gamma imaging, we therefore propose a time-of-flight (TOF) technique to reject this neutron background, involving a shifting time window to account for the propagation of the protons through the patient. Time-resolved Monte Carlo simulations of the generation and transport of prompt gamma photons and neutrons upon irradiation of a PMMA phantom with 100, 150 and 200 MeV protons were performed using Geant4 (version 9.2.p02) and MCNPX (version 2.7.D). The influence of angular collimation and TOF selection on the prompt gamma and neutron longitudinal profiles is studied. Furthermore, the implications of the proton beam microstructure (characterized by the proton bunch width and repetition period) are investigated. The application of a shifting TOF window having a width of ΔTOF(z) = 1.0 ns appears to reduce the neutron background by more than 99%. Subsequent application of an energy threshold does not appear to sharpen the distal falloff of the prompt gamma profile but reduces the tail that is observed beyond the proton range. Investigations of the influence of the beam time structure show that TOF rejection of the neutron background is expected to be effective for typical therapeutic proton cyclotrons.

[1]  D Robertson,et al.  Optimizing a three-stage Compton camera for measuring prompt gamma rays emitted during proton radiotherapy , 2010, Physics in medicine and biology.

[2]  O Jäkel,et al.  The Heidelberg Ion Therapy Center. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[3]  M. Daube-Witherspoon,et al.  The imaging performance of a LaBr3-based PET scanner , 2010, Physics in medicine and biology.

[4]  Denis Dauvergne,et al.  Design of a Compton camera for 3D prompt-γ imaging during ion beam therapy , 2011 .

[5]  S. Dolinsky,et al.  Multi-Pixel Photon Counters for TOF PET Detector and Its Challenges , 2009, IEEE Transactions on Nuclear Science.

[6]  P Chauvel,et al.  Monte Carlo simulation of a protontherapy platform devoted to ocular melanoma. , 2005, Medical physics.

[7]  M. Foresta,et al.  Silicon Photomultipliers (SiPM) as novel photodetectors for PET , 2011 .

[8]  Wolfgang Enghardt,et al.  A Compton imager for in-vivo dosimetry of proton beams—A design study , 2011 .

[9]  M. Bajard,et al.  Monitoring the Bragg peak location of 73 MeV/u carbon ions by means of prompt γ-ray measurements , 2008, 0809.0185.

[10]  Shinichi Minohara,et al.  Heavy ion synchrotron for medical use —HIMAC project at NIRS-Japan— , 1992 .

[11]  F. Stichelbaut,et al.  Application of the HICAM camera for imaging of prompt gamma rays in measurements of proton beam range , 2011, 2011 IEEE Nuclear Science Symposium Conference Record.

[12]  Sam Beddar,et al.  Material efficiency studies for a Compton camera designed to measure characteristic prompt gamma rays emitted during proton beam radiotherapy , 2011, Physics in medicine and biology.

[13]  Giacomo Cuttone,et al.  Hadrontherapy: An open source, Geant4-based application for proton-ion therapy studies , 2009, 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC).

[14]  T. Bortfeld IMRT: a review and preview , 2006, Physics in medicine and biology.

[15]  Freek Beekman,et al.  Real-time prompt gamma monitoring in spot-scanning proton therapy using imaging through a knife-edge-shaped slit , 2012, Physics in medicine and biology.

[16]  R. K. Bull,et al.  Stopping powers for electrons and positrons: ICRU Report 37; 271 pp.; 24 figures; U.S. $24.00. , 1986 .

[17]  J. Schippers Beam Delivery Systems for Particle Radiation Therapy: Current Status and Recent Developments , 2009 .

[18]  F Verhaegen,et al.  Monte Carlo calculations of positron emitter yields in proton radiotherapy , 2012, Physics in medicine and biology.

[19]  C. Lacasta,et al.  First tests in the application of silicon photomultiplier arrays to dose monitoring in hadron therapy , 2011 .

[20]  Chan Hyeong Kim,et al.  Prompt gamma measurements for locating the dose falloff region in the proton therapy , 2006 .

[21]  C. Piemonte,et al.  The DLED Algorithm for Timing Measurements on Large Area SiPMs Coupled to Scintillators , 2012, IEEE Transactions on Nuclear Science.

[22]  D. Ondreka,et al.  The Heidelberg Ion Therapy (HIT) Accelerator Coming into Operation , 2009 .

[23]  S Beddar,et al.  Measurement and calculation of characteristic prompt gamma ray spectra emitted during proton irradiation , 2009, Physics in medicine and biology.

[24]  A. Lomax,et al.  Intensity modulated proton therapy and its sensitivity to treatment uncertainties 1: the potential effects of calculational uncertainties , 2008, Physics in medicine and biology.

[25]  P. Busca,et al.  Prompt gamma imaging with a slit camera for real-time range control in proton therapy , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[26]  Christian Hilbes,et al.  The PSI Gantry 2: a second generation proton scanning gantry. , 2004, Zeitschrift fur medizinische Physik.

[27]  D. Schaart,et al.  The lower bound on the timing resolution of scintillation detectors , 2012, Physics in medicine and biology.

[28]  D. Schaart,et al.  Monolithic LaBr3:Ce crystals on silicon photomultiplier arrays for time-of-flight positron emission tomography , 2012, Physics in medicine and biology.

[29]  Harald Paganetti,et al.  Relative biological effectiveness (RBE) values for proton beam therapy. , 2002, International journal of radiation oncology, biology, physics.

[30]  D. Schardt,et al.  Magnetic scanning system for heavy ion therapy , 1993 .

[31]  P. Lecoq,et al.  Time based readout of a silicon photomultiplier (SiPM) for time of flight positron emission tomography (TOF-PET) , 2009, 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC).

[32]  David Sarrut,et al.  Optimization of GEANT4 settings for Proton Pencil Beam Scanning simulations using GATE , 2010 .

[33]  Masahiro Endo,et al.  Broad-beam three-dimensional irradiation system for heavy-ion radiotherapy at HIMAC , 1999 .

[34]  R. Mohan,et al.  An MCNPX Monte Carlo model of a discrete spot scanning proton beam therapy nozzle. , 2010, Medical physics.

[35]  Ky Cheung Biomedical Imaging and Intervention Journal Intensity Modulated Radiotherapy: Advantages, Limitations and Future Developments , 2022 .

[36]  Maurizio Conti,et al.  Improving time resolution in time-of-flight PET , 2011 .

[37]  H. G. Hughes,et al.  Cross-Section Evaluations to 150 MeV for Accelerator-Driven Systems and Implementation in MCNPX , 1999 .

[38]  P. Busca,et al.  89 PROMPT GAMMA IMAGING WITH A SLIT CAMERA FOR REAL TIME RANGE CONTROL IN PROTON THERAPY , 2012 .

[39]  A J Lomax,et al.  Intensity modulated proton therapy and its sensitivity to treatment uncertainties 2: the potential effects of inter-fraction and inter-field motions , 2008, Physics in medicine and biology.

[40]  G Ciangaru,et al.  Prompt gamma-ray emission from biological tissues during proton irradiation: a preliminary study , 2009, Physics in medicine and biology.

[41]  H Paganetti,et al.  Monte Carlo patient study on the comparison of prompt gamma and PET imaging for range verification in proton therapy , 2011, Physics in medicine and biology.

[42]  Michael Scholz,et al.  Effectiveness of carbon ion radiotherapy in the treatment of skull-base chordomas. , 2007, International journal of radiation oncology, biology, physics.

[43]  Magdalena Rafecas,et al.  Detector characterization and first coincidence tests of a Compton telescope based on LaBr3 crystals and SiPMs , 2012 .

[44]  Denis Dauvergne,et al.  Real-time monitoring of the Bragg-peak position in ion therapy by means of single photon detection , 2010, Radiation and environmental biophysics.

[45]  B. Zwaans,et al.  Arrays of digital Silicon Photomultipliers — Intrinsic performance and application to scintillator readout , 2010, IEEE Nuclear Science Symposuim & Medical Imaging Conference.