Report of a Multilateral Accurate Measurement Comparison on a High-Precision Multimeter to Evaluate the Traceability Transfer from INRIM in the Field of Low-Frequency Electrical Quantities

As the calibration activity of modern electrical secondary laboratories operating in the five low-frequency quantities is carried out by means of high-precision digital multimeters (DMMs), it is of fundamental importance for their correct technical surveillance through suitable inter-laboratory comparisons (ILCs). The laboratory for calibration of multifunction electrical instruments of the National Institute of Metrology Research (INRIM-Lab) provides bilateral and multilateral ILCs involving high-precision DMMs to check the measurement capabilities of electrical laboratories. In particular, this paper deals with the last multilateral ILC that INRIM-Lab provided and carried out as reference measurements’ provider to which took part twelve Italian-accredited laboratories divided into two loops. The ILC was involved in a high-precision 8.5 digit high-precision DMM to be verified in a grid of about forty measurement points. The 2σ relative uncertainties of the INRIM-Lab calibrations ranged from 1.6 × 10−6 to 1.5 × 10−4, while those of the laboratories ranged from 2.0 × 10−6 to 2.2 × 10−3. The ILC had satisfactory results confirming the correctness of both the dissemination from INRIM and the accreditation process by the Italian accreditation body for calibration laboratories.

[1]  E. Gasparotto,et al.  Analysis of a national comparison in the field of electrical low dc resistance , 2014 .

[2]  Flavio Galliana,et al.  Inter-laboratories comparison at 100 GΩ and 1 TΩ level to evaluate the traceability transfer from INRIM in the field of high dc resistance , 2012 .

[3]  F. Galliana,et al.  Accurate comparison between INRIM and a secondary calibration laboratory using a top-class multifunction electrical calibrator , 2017 .

[4]  Vinicius Alves Pessanha,et al.  Proposta para acreditação da divisão de produção de painéis sorológicos de Bio-Manguinhos / Fiocruz na norma ISO / IEC 17043 - Conformity Assessment – General Requirements for Proficiency Testing , 2011 .

[5]  Marco Del Moro,et al.  The Strategic Role of Inter-Laboratory Comparison Among International Institutions to Assure Confidence: Report of the INRIM and the European Spatial Agency Comparison on Electrical Quantities , 2018 .

[6]  Umberto Pogliano,et al.  Stability evaluation of high-precision multifunction instruments for traceability transfer , 2000, IEEE Trans. Instrum. Meas..

[7]  Flavio Galliana,et al.  Report of the Italian inter-laboratories comparison of high dc resistance on the calibration of a 10MΩ and a 1GΩ resistors , 2009 .

[8]  Leonard Steinborn,et al.  International Organization for Standardization ISO/IEC 17025 General Requirements for the Competence of Testing and Calibration Laboratories , 2004 .

[9]  F. Francone,et al.  Portable DC Voltage and Resistance Reference with Switching Unit for Calibration of Electrical Precision Multifunction Instruments , 2018, Conference on Precision Electromagnetic Measurements.

[10]  F. Galliana,et al.  1 {\Omega}-10 k{\Omega} high precision transportable setup to calibrate multifunction electrical instruments , 2015, 1505.04398.