Identification of microemboli during haemodialysis using Doppler ultrasound.

BACKGROUND Doppler ultrasound methods were used during haemodialysis sessions for the detection of microemboli and determination of their origin. METHODS A 2-MHz ultrasound probe (Multidop X(4) DWL((TM))) was used to assess the number of microembolic signals (MES) in the subclavian vein downstream from the arteriovenous fistula before the dialysis session and over two periods of 15 min at the beginning and end of haemodialysis sessions in 25 patients without previous cardiovascular disease. A similar probe was used during in vitro studies to detect MES at different sites in the dialysis machine (before and downstream from the blood pump, and before and downstream from the air trap). RESULTS No MES were detected during in vivo studies before haemodialysis sessions. MES were registered in all patients (100%) at the beginning and end of the haemodialysis procedure at an average of 12.7+/-9 and 16. 7+/-11.5 signals/min respectively. The average intensity of MES was 19.2+/-5.0 dB and 19.4+/-3.9 dB respectively. No MES were detected on the arterial line during in vitro studies. In contrast, 19+/-6 MES/min were detected after the blood pump, 13+/-4.2 before the air trap, and 16.5+/-5.5 thereafter. CONCLUSIONS In all patients, MES were recorded during haemodialysis sessions in the drainage vein from arteriovenous fistulae. The results of in vitro studies indicate that MES are formed by the blood pump of the haemodialysis machine. The intensity of the MES suggests that they correspond to synthetic particles or microbubbles, which are not detected by the air trap. The final destination of these microbubbles will be assessed in further studies.

[1]  K. Albertine,et al.  Quantification of damage by air emboli to lung microvessels in anesthetized sheep. , 1984, Journal of applied physiology: respiratory, environmental and exercise physiology.

[2]  E. Ritz,et al.  Reduction of silicone particle release during haemodialysis. , 1985, Proceedings of the European Dialysis and Transplant Association - European Renal Association. European Dialysis and Transplant Association - European Renal Association. Congress.

[3]  W. Hörl,et al.  Studies of biocompatibility of different dialyzer membranes: role of complement system, intracellular calcium and inositol-triphosphate. , 1994, Clinical nephrology.

[4]  M.A. Moehring,et al.  Pulse Doppler ultrasound detection, characterization and size estimation of emboli in flowing blood , 1994, IEEE Transactions on Biomedical Engineering.

[5]  J. Gold,et al.  Comparison of transcranial Doppler ultrasonography and transesophageal echocardiography to monitor emboli during coronary artery bypass surgery. , 1996, Stroke.

[6]  H S Markus,et al.  Asymptomatic embolization predicts stroke and TIA risk in patients with carotid artery stenosis. , 1999, Stroke.

[7]  M Kaps,et al.  Variability of Doppler microembolic signal counts in patients with prosthetic cardiac valves. , 1995, Stroke.

[8]  S. Harbottle,et al.  Particle spallation induced by blood pumps in hemodialysis tubing sets. , 1986, Artificial organs.

[9]  P. Dimakakos,et al.  Asymptomatic embolization predicts stroke and TIA risk in patients with carotid artery stenosis. , 1999, Stroke.

[10]  Manfred Kaps,et al.  Consensus on Microembolus Detection by TCD , 1998 .

[11]  G. Donelli,et al.  Particle migration from haemodialysis circuit: electron microscopy and microprobe analysis. , 1988, Biomaterials, artificial cells, and artificial organs.

[12]  H. Guénard,et al.  Membrane diffusion of the lungs in patients with chronic renal failure. , 1993, The European respiratory journal.

[13]  K. Lees,et al.  Doppler Emboli Signals Vary According to Stroke Subtype , 1994, Stroke.

[14]  A. Tsuchida,et al.  Evaluation of blood coagulation-fibrinolysis system in patients receiving chronic hemodialysis. , 1996, Nephron.

[15]  J. Svennevig,et al.  Cerebral microembolic signals during cardiopulmonary bypass surgery. Frequency, time of occurrence, and association with patient and surgical characteristics. , 1997, Stroke.

[16]  L R Sauvage,et al.  Detection of middle cerebral artery emboli during carotid endarterectomy using transcranial Doppler ultrasonography. , 1990, Stroke.

[17]  M. Roussel,et al.  Microemboli in cerebral circulation and alteration of cognitive abilities in patients with mechanical prosthetic heart valves. , 1998, Stroke.

[18]  B. Meyrick,et al.  Continuous air embolization into sheep causes sustained pulmonary hypertension and increased pulmonary vasoreactivity. , 1988, The American journal of pathology.

[19]  K. Ang,et al.  Enhanced platelet aggregation and membrane biocompatibility: possible influence on thrombosis and embolism in hemodialysis patients. , 1987, Nephron.

[20]  M. Ziemen,et al.  Enhanced platelet aggregation as a risk factor for vascular disease and its thromboembolic complications. , 1975, Acta Clinica Belgica.

[21]  M. Brown,et al.  Computerized Detection of Cerebral Emboli and Discrimination From Artifact Using Doppler Ultrasound , 1993, Stroke.

[22]  A. Muir,et al.  Altered lung vascular permeability during intermittent haemodialysis. , 1988, Nephrology, dialysis, transplantation : official publication of the European Dialysis and Transplant Association - European Renal Association.

[23]  T. M. Runge,et al.  Tubing Spallation in Extracorporeal Circuits. An in Vitro Study Using an Electronic Particle Counter , 1992, The International journal of artificial organs.