Temperature and SAR measurement errors in the evaluation of metallic linear structures heating during MRI using fluoroptic® probes

The purpose of this work is to evaluate the error associated with temperature and SAR measurements using fluoroptic temperature probes on pacemaker (PM) leads during magnetic resonance imaging (MRI). We performed temperature measurements on pacemaker leads, excited with a 25, 64, and 128 MHz current. The PM lead tip heating was measured with a fluoroptic thermometer (Luxtron, Model 3100, USA). Different contact configurations between the pigmented portion of the temperature probe and the PM lead tip were investigated to find the contact position minimizing the temperature and SAR underestimation. A computer model was used to estimate the error made by fluoroptic probes in temperature and SAR measurement. The transversal contact of the pigmented portion of the temperature probe and the PM lead tip minimizes the underestimation for temperature and SAR. This contact position also has the lowest temperature and SAR error. For other contact positions, the maximum temperature error can be as high as -45%, whereas the maximum SAR error can be as high as -54%. MRI heating evaluations with temperature probes should use a contact position minimizing the maximum error, need to be accompanied by a thorough uncertainty budget and the temperature and SAR errors should be specified.

[1]  M. Niehaus,et al.  Electromagnetic interference in patients with implanted pacemakers or cardioverter-defibrillators , 2001, Heart.

[2]  Frank G Shellock,et al.  MR Safety and the American College of Radiology White Paper. , 2002, AJR. American journal of roentgenology.

[3]  Isaac Chang,et al.  Finite Element Analysis of Hepatic Radiofrequency Ablation Probes using Temperature-Dependent Electrical Conductivity , 2003, Biomedical engineering online.

[4]  S. Souza,et al.  Real‐time position monitoring of invasive devices using magnetic resonance , 1993, Magnetic resonance in medicine.

[5]  LEONARD T. BLOUIN,et al.  Assessment of Effects of a Radiofrequency Energy Field and Thermistor Location in an Electrode Catheter on the Accuracy of Temperature Measurement , 1991, Pacing and clinical electrophysiology : PACE.

[6]  J. Felmlee,et al.  American College of Radiology White Paper on MR Safety: 2004 update and revisions. , 2004, AJR. American journal of roentgenology.

[7]  W. Nitz,et al.  On the heating of linear conductive structures as guide wires and catheters in interventional MRI , 2001, Journal of magnetic resonance imaging : JMRI.

[8]  Robert E Lenkinski,et al.  Radiofrequency ablation: importance of background tissue electrical conductivity--an agar phantom and computer modeling study. , 2005, Radiology.

[9]  F G Shellock,et al.  Radiofrequency Energy-Induced Heating of Bovine Articular Cartilage Using a Bipolar Radiofrequency Electrode , 2000, American Journal of Sports Medicine.

[10]  Henry R. Halperin,et al.  Modern Pacemaker and Implantable Cardioverter/Defibrillator Systems Can Be Magnetic Resonance Imaging Safe: In Vitro and In Vivo Assessment of Safety and Function at 1.5 T , 2004, Circulation.

[11]  J Gieseke,et al.  MR imaging and cardiac pacemakers: in-vitro evaluation and in-vivo studies in 51 patients at 0.5 T. , 2000, Radiology.

[12]  F. Shellock,et al.  Physiologic responses to an MR imaging procedure performed at a specific absorption rate of 6.0 W/kg. , 1994, Radiology.

[13]  Richard G Trohman,et al.  Interference in Implanted Cardiac Devices, Part II , 2002, Pacing and clinical electrophysiology : PACE.

[14]  C L Dumoulin,et al.  Intravascular MR tracking catheter: preliminary experimental evaluation. , 1995, AJR. American journal of roentgenology.

[15]  F. Shellock Thermal Responses in Human Subjects Exposed to Magnetic Resonance Imaging a , 1992, Annals of the New York Academy of Sciences.

[16]  H. K. Charles,et al.  Catheter‐tracking FOV MR fluoroscopy , 1998, Magnetic resonance in medicine.

[17]  P S Ruggera,et al.  In vitro assessment of tissue heating near metallic medical implants by exposure to pulsed radio frequency diathermy. , 2003, Physics in medicine and biology.

[18]  R W Günther,et al.  Catheter visualization using locally induced, actively controlled field inhomogeneities , 1997, Magnetic resonance in medicine.

[19]  H Calkins,et al.  Temperature Monitoring During Radiofrequency Ablation , 1996, Journal of cardiovascular electrophysiology.

[20]  Jean A. Tkach,et al.  Evaluation of specific absorption rate as a dosimeter of MRI‐related implant heating , 2004, Journal of magnetic resonance imaging : JMRI.

[21]  K Bachmann,et al.  Effects of magnetic resonance imaging on cardiac pacemakers and electrodes. , 1997, American heart journal.