Fast T2 Mapping With Improved Accuracy Using Undersampled Spin-Echo MRI and Model-Based Reconstructions With a Generating Function

A model-based reconstruction technique for accelerated T2 mapping with improved accuracy is proposed using undersampled Cartesian spin-echo magnetic resonance imaging (MRI) data. The technique employs an advanced signal model for T2 relaxation that accounts for contributions from indirect echoes in a train of multiple spin echoes. An iterative solution of the nonlinear inverse reconstruction problem directly estimates spin-density and T2 maps from undersampled raw data. The algorithm is validated for simulated data as well as phantom and human brain MRI at 3T. The performance of the advanced model is compared to conventional pixel-based fitting of echo-time images from fully sampled data. The proposed method yields more accurate T2 values than the mono-exponential model and allows for retrospective undersampling factors of at least 6. Although limitations are observed for very long T2 relaxation times, respective reconstruction problems may be overcome by a gradient dampening approach. The analytical gradient of the utilized cost function is included as Appendix . The source code is made available to the community.

[1]  Mariya Doneva,et al.  Compressed sensing reconstruction for magnetic resonance parameter mapping , 2010, Magnetic resonance in medicine.

[2]  Rudolf Stollberger,et al.  Closed‐form solution for T2 mapping with nonideal refocusing of slice selective CPMG sequences , 2015, Magnetic resonance in medicine.

[3]  J. Gore,et al.  Errors in the measurements of T2 using multiple‐echo MRI techniques. II. Effects of static field inhomogeneity , 1986, Magnetic resonance in medicine.

[4]  J. Hennig Echoes—how to generate, recognize, use or avoid them in MR‐imaging sequences. Part I: Fundamental and not so fundamental properties of spin echoes , 1991 .

[5]  Nikola Stikov,et al.  Practical medical applications of quantitative MR relaxometry , 2012, Journal of magnetic resonance imaging : JMRI.

[6]  Avinash C. Kak,et al.  Principles of computerized tomographic imaging , 2001, Classics in applied mathematics.

[7]  V. Kiselev,et al.  Calculation of diffusion effect for arbitrary pulse sequences. , 2003, Journal of magnetic resonance.

[8]  Y. Zur An algorithm to calculate the NMR signal of a multi spin-echo sequence with relaxation and spin-diffusion. , 2004, Journal of magnetic resonance.

[9]  M. Glas,et al.  Principles of Computerized Tomographic Imaging , 2000 .

[10]  Jens Frahm,et al.  Model‐based nonlinear inverse reconstruction for T2 mapping using highly undersampled spin‐echo MRI , 2011, Journal of magnetic resonance imaging : JMRI.

[11]  Jens Frahm,et al.  Model-Based Iterative Reconstruction for Radial Fast Spin-Echo MRI , 2009, IEEE Transactions on Medical Imaging.

[12]  L. Comtet,et al.  Advanced Combinatorics: The Art of Finite and Infinite Expansions , 1974 .

[13]  N. Lukzen,et al.  Analytical derivation of multiple spin echo amplitudes with arbitrary refocusing angle. , 2007, Journal of magnetic resonance.

[14]  J C Gore,et al.  Errors in the measurements of T2 using multiple‐echo MRI techniques. I. Effects of radiofrequency pulse imperfections , 1986, Magnetic resonance in medicine.

[15]  Ali Bilgin,et al.  T2 relaxometry with indirect echo compensation from highly undersampled data , 2013, Magnetic resonance in medicine.

[16]  R. Lebel,et al.  Transverse relaxometry with stimulated echo compensation , 2010, Magnetic resonance in medicine.

[17]  J. Hennig Multiecho imaging sequences with low refocusing flip angles , 1988 .

[18]  Rik Van de Walle,et al.  Reconstruction of MR images from data acquired on a general nonregular grid by pseudoinverse calculation , 2000, IEEE Transactions on Medical Imaging.

[19]  C. S. Poon,et al.  Practical T2 quantitation for clinical applications , 1992, Journal of magnetic resonance imaging : JMRI.

[20]  Ali Bilgin,et al.  T2 mapping from highly undersampled data by reconstruction of principal component coefficient maps using compressed sensing , 2012, Magnetic resonance in medicine.

[21]  Herbert S. Wilf,et al.  Generating functionology , 1990 .

[22]  F. Wiesinger,et al.  B1 mapping by Bloch‐Siegert shift , 2010, Magnetic resonance in medicine.

[23]  V J Schmithorst,et al.  Spatial variation in cartilage T2 of the knee , 2001, Journal of magnetic resonance imaging : JMRI.

[24]  T. Hohage,et al.  Image reconstruction by regularized nonlinear inversion—Joint estimation of coil sensitivities and image content , 2008, Magnetic resonance in medicine.

[25]  A. Crawley,et al.  Errors in T2 estimation using multislice multiple‐echo imaging , 1987, Magnetic resonance in medicine.

[26]  D. Woessner Effects of Diffusion in Nuclear Magnetic Resonance Spin-Echo Experiments , 1961 .

[27]  William W. Hager,et al.  A New Conjugate Gradient Method with Guaranteed Descent and an Efficient Line Search , 2005, SIAM J. Optim..

[28]  T J Mosher,et al.  Functional cartilage MRI T2 mapping: evaluating the effect of age and training on knee cartilage response to running. , 2010, Osteoarthritis and cartilage.