Axially Extended-Volume C-Arm CT Using a Reverse Helical Trajectory in the Interventional Room

C-arm computed tomography (CT) is an innovative technique that enables a C-arm system to generate 3-D images from a set of 2-D X-ray projections. This technique can reduce treatment-related complications and may improve interventional efficacy and safety. However, state-of-the-art C-arm systems rely on a circular short scan for data acquisition, which limits coverage in the axial direction. This limitation was reported as a problem in hepatic vascular interventions. To solve this problem, as well as to further extend the value of C-arm CT, axially extended-volume C-arm CT is needed. For example, such an extension would enable imaging the full aorta, the peripheral arteries or the spine in the interventional room, which is currently not feasible. In this paper, we demonstrate that performing long object imaging using a reverse helix is feasible in the interventional room. This demonstration involved developing a novel calibration method, assessing geometric repeatability, implementing a reconstruction method that applies to real reverse helical data, and quantitatively evaluating image quality. Our results show that: 1) the reverse helical trajectory can be implemented and reliably repeated on a multiaxis C-arm system; and 2) a long volume can be reconstructed with satisfactory image quality using reverse helical data.

[1]  Xiaochuan Pan,et al.  Exact reconstruction of volumetric images in reverse helical cone-beam CT. , 2008, Medical physics.

[2]  F. Noo,et al.  FDK-type reconstruction algorithms for the reverse helical trajectory , 2011, 2011 IEEE Nuclear Science Symposium Conference Record.

[3]  Thomas Albrecht,et al.  Contrast-Enhanced Abdominal Angiographic CT for Intra-abdominal Tumor Embolization: A New Tool for Vessel and Soft Tissue Visualization , 2007, CardioVascular and Interventional Radiology.

[4]  Xiaochuan Pan,et al.  Exact reconstruction of volumetric images in reverse helical cone-beam CT. , 2008, Medical physics.

[5]  N. Navab,et al.  Enhanced 3-D-reconstruction algorithm for C-arm systems suitable for interventional procedures , 2000, IEEE Transactions on Medical Imaging.

[6]  Xiaochuan Pan,et al.  A BPF-FBP tandem algorithm for image reconstruction in reverse helical cone-beam CT. , 2010, Medical physics.

[7]  Frédéric Noo,et al.  Cone-beam artifact evaluation of the factorization method. , 2011, Medical physics.

[8]  Herlen Alencar,et al.  Translumbar type II endoleak repair using angiographic CT. , 2006, Journal of vascular and interventional radiology : JVIR.

[9]  Ken Ueda,et al.  Cone-Beam CT with Flat-Panel-Detector Digital Angiography System: Early Experience in Abdominal Interventional Procedures , 2006, CardioVascular and Interventional Radiology.

[10]  J. Remy,et al.  Pulmonary nodules: detection with thick-section spiral CT versus conventional CT. , 1993, Radiology.

[11]  Joachim Hornegger,et al.  Ellipse-line-ellipse source trajectory and its R-line coverage for long-object cone-beam imaging with a C-arm system , 2012, Medical Imaging.

[12]  Anne Rougee,et al.  Geometrical calibration for 3D x-ray imaging , 1993, Medical Imaging.

[13]  Gilles Soulez,et al.  Three-dimensional C-arm cone-beam CT: applications in the interventional suite. , 2008, Journal of vascular and interventional radiology : JVIR.

[14]  Günter Lauritsch,et al.  Line plus arc source trajectories and their R-line coverage for long-object cone-beam imaging with a C-arm system. , 2011, Physics in medicine and biology.

[15]  O. Faugeras Three-dimensional computer vision: a geometric viewpoint , 1993 .

[16]  R Fahrig,et al.  Three-dimensional computed tomographic reconstruction using a C-arm mounted XRII: image-based correction of gantry motion nonidealities. , 2000, Medical physics.

[17]  Nassir Navab,et al.  Enhanced 3D-reconstruction algorithms for C-Arm based interventional procedures , 2000, IEEE Trans. Medical Imaging.

[18]  Xiaochuan Pan,et al.  Non-circular cone beam CT trajectories: A preliminary investigation on a clinical scanner , 2010, IEEE Nuclear Science Symposuim & Medical Imaging Conference.

[19]  Ravi Murthy,et al.  Impact of C-arm CT on hepatic arterial interventions for hepatic malignancies. , 2007, Journal of vascular and interventional radiology : JVIR.

[20]  H. Tuy AN INVERSION FORMULA FOR CONE-BEAM RECONSTRUCTION* , 1983 .

[21]  David V. Finch CONE BEAM RECONSTRUCTION WITH SOURCES ON A CURVE , 1985 .

[22]  Timothy A. Brunner,et al.  Low contrast 3D reconstruction from C-arm data , 2005, SPIE Medical Imaging.

[23]  G. Wang,et al.  A general cone-beam reconstruction algorithm , 1993, IEEE Trans. Medical Imaging.

[24]  Rui Liao,et al.  System to Guide Transcatheter Aortic Valve Implantations Based on Interventional C-Arm CT Imaging , 2010, MICCAI.

[25]  Andreas Keil,et al.  Analysis of vertical and horizontal circular C-arm trajectories , 2011, Medical Imaging.

[26]  F. Noo,et al.  Direct determination of geometric alignment parameters for cone-beam scanners. , 2009, Physics in medicine and biology.

[27]  Nassir Navab,et al.  Recovering the X-ray projection geometry for three-dimensional tomographic reconstruction with additional sensors: Attached camera versus external navigation system , 2003, Medical Image Anal..

[28]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[29]  Drazenko Babic,et al.  Percutaneous Vertebroplasty: Preliminary Experiences with Rotational Acquisitions and 3D Reconstructions for Therapy Control , 2006, CardioVascular and Interventional Radiology.

[30]  Günter Lauritsch,et al.  First Experimental Results on Long-object Imaging using a Reverse Helical Trajectory with a Carm System , 2012 .

[31]  Deshan Yang,et al.  Physical phantom studies of helical cone-beam CT with exact reconstruction. , 2012, Medical physics.

[32]  Michael D. Silver,et al.  Determination and correction of the wobble of a C-arm gantry , 2000, Medical Imaging: Image Processing.

[33]  Benno Heigl,et al.  Improving 3D image quality of x-ray C-arm imaging systems by using properly designed pose determination systems for calibrating the projection geometry , 2003, SPIE Medical Imaging.

[34]  Günter Lauritsch,et al.  Geometric calibration of the circle-plus-arc trajectory. , 2007, Physics in medicine and biology.