AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy.

The introduction of advanced techniques and technology in radiotherapy has greatly improved our ability to deliver highly conformal tumor doses while minimizing the dose to adjacent organs at risk. Despite these tremendous improvements, there remains a general concern about doses to normal tissues that are not the target of the radiation treatment; any "nontarget" radiation should be minimized as it offers no therapeutic benefit. As patients live longer after treatment, there is increased opportunity for late effects including second cancers and cardiac toxicity to manifest. Complicating the management of these issues, there are unique challenges with measuring, calculating, reducing, and reporting nontarget doses that many medical physicists may have limited experience with. Treatment planning systems become dramatically inaccurate outside the treatment field, necessitating a measurement or some other means of assessing the dose. However, measurements are challenging because outside the treatment field, the radiation energy spectrum, dose rate, and general shape of the dose distribution (particularly the percent depth dose) are very different and often require special consideration. Neutron dosimetry is also particularly challenging, and common errors in methodology can easily manifest as errors of several orders of magnitude. Task Group 158 was, therefore, formed to provide guidance for physicists in terms of assessing and managing nontarget doses. In particular, the report: (a) highlights major concerns with nontarget radiation; (b) provides a rough estimate of doses associated with different treatment approaches in clinical practice; (c) discusses the uses of dosimeters for measuring photon, electron, and neutron doses; (d) discusses the use of calculation techniques for dosimetric evaluations; (e) highlights techniques that may be considered for reducing nontarget doses; (f) discusses dose reporting; and (g) makes recommendations for both clinical and research practice.

[1]  H. Vega-Carrillo,et al.  Monte Carlo simulation estimates of neutron doses to critical organs of a patient undergoing 18MV x-ray LINAC-based radiotherapy. , 2005, Medical physics.

[2]  A. Bozkurt,et al.  VIP-MAN: AN IMAGE-BASED WHOLE-BODY ADULT MALE MODEL CONSTRUCTED FROM COLOR PHOTOGRAPHS OF THE VISIBLE HUMAN PROJECT FOR MULTI-PARTICLE MONTE CARLO CALCULATIONS , 2000, Health physics.

[3]  K. Kase,et al.  Neutron sources in the Varian Clinac 2100C/2300C medical accelerator calculated by the EGS4 code. , 1997, Health physics.

[4]  L. Marks,et al.  Prospective assessment of radiotherapy‐associated cardiac toxicity in breast cancer patients: Analysis of data 3 to 6 years after treatment , 2007, Cancer.

[5]  C. R. Edwards,et al.  Near surface photon energy spectra outside a 6 MV field edge. , 2004, Physics in medicine and biology.

[6]  Radhe Mohan,et al.  Out-of-field photon dose following removal of the flattening filter from a medical accelerator , 2010, Physics in medicine and biology.

[7]  U. Schneider,et al.  Systematic measurements of whole-body dose distributions for various treatment machines and delivery techniques in radiation therapy. , 2012, Medical physics.

[8]  Timothy C Zhu,et al.  Energy dependence of commercially available diode detectors for in-vivo dosimetry. , 2007, Medical physics.

[9]  B. Gottschalk Neutron dose in scattered and scanned proton beams: in regard to Eric J. Hall (Int J Radiat Oncol Biol Phys 2006;65:1-7). , 2006, International journal of radiation oncology, biology, physics.

[10]  S. L. Morgan-Fletcher Prescribing, Recording and Reporting Photon Beam Therapy (Supplement to ICRU Report 50), ICRU Report 62. ICRU, pp. ix+52, 1999 (ICRU Bethesda, MD) $65.00 ISBN 0-913394-61-0 , 2001 .

[11]  Kenneth E Rosenzweig,et al.  Acute skin toxicity following stereotactic body radiation therapy for stage I non-small-cell lung cancer: who's at risk? , 2008, International journal of radiation oncology, biology, physics.

[12]  David S Followill,et al.  Out-of-field photon and neutron dose equivalents from step-and-shoot intensity-modulated radiation therapy. , 2005, International journal of radiation oncology, biology, physics.

[13]  X. Xu,et al.  Better radiation weighting factors for neutrons generated from proton treatment are needed. , 2010, Radiation protection dosimetry.

[14]  A. Mesbahi A Monte Carlo study on neutron and electron contamination of an unflattened 18-MV photon beam. , 2009, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[15]  V Saenko,et al.  The Chernobyl accident and its consequences. , 2011, Clinical oncology (Royal College of Radiologists (Great Britain)).

[16]  S. Kry,et al.  Accuracy of out-of-field dose calculations by a commercial treatment planning system , 2010, Physics in medicine and biology.

[17]  Steve B. Jiang,et al.  The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75. , 2007, Medical physics.

[18]  U. Schneider,et al.  Comparative simulations of neutron dose in soft tissue and phantom materials for proton and carbon ion therapy with actively scanned beams. , 2011, Medical physics.

[19]  U. Schneider,et al.  Measurements of the neutron dose equivalent for various radiation qualities, treatment machines and delivery techniques in radiation therapy , 2014, Physics in medicine and biology.

[20]  W. Newhauser,et al.  Equivalent dose and effective dose from stray radiation during passively scattered proton radiotherapy for prostate cancer , 2008, Physics in medicine and biology.

[21]  J. Seco,et al.  Comparison of out-of-field photon doses in 6 MV IMRT and neutron doses in proton therapy for adult and pediatric patients , 2010, Physics in medicine and biology.

[22]  G. Starkschall,et al.  Surface dose for megavoltage photon beams outside the treatment field. , 1983, Medical physics.

[23]  C. Berg,et al.  Second solid cancers after radiotherapy for breast cancer in SEER cancer registries , 2009, British Journal of Cancer.

[24]  D. Followill,et al.  A Monte Carlo model for out-of-field dose calculation from high-energy photon therapy. , 2007, Medical physics.

[25]  E. Hall,et al.  Radiation-induced second cancers: the impact of 3D-CRT and IMRT. , 2003, International journal of radiation oncology, biology, physics.

[26]  D. Hoel Second Primary Cancers and Cardiovascular Disease after Radiation Therapy , 2012 .

[27]  H. Schuhmacher,et al.  Quality factors, ambient and personal dose equivalent for neutrons, based on the new ICRU stopping power data for protons and alpha particles , 1995 .

[28]  D. L. Preston,et al.  Solid Cancer Incidence in Atomic Bomb Survivors: 1958–1998 , 2007, Radiation research.

[29]  D W O Rogers,et al.  Monte Carlo calculation of nine megavoltage photon beam spectra using the BEAM code. , 2002, Medical physics.

[30]  David J. Brenner,et al.  Estimating radiation-induced cancer risks at very low doses: rationale for using a linear no-threshold approach , 2006, Radiation and environmental biophysics.

[31]  G. G. Eigenwillig Comment on “Mortality from cardiovascular diseases in the German uranium miners cohort study, 1946–1998” by Kreuzer M, Kreisheimer M, Kandel M, Schnelzer M, Tschense A, Grosche B (2006) Radiat Environ Biophys 45:159–166 , 2007, Radiation and environmental biophysics.

[32]  J. W. Vieira,et al.  Comparison between effective doses for voxel-based and stylized exposure models from photon and electron irradiation , 2005, Physics in medicine and biology.

[33]  R. Nath,et al.  Depth dose-equivalent and effective energies of photoneutrons generated by 6-18 MV X-ray beams for radiotherapy. , 2001, Health physics.

[34]  T. Bortfeld,et al.  Neutron dose in proton radiation therapy: in regard to Eric J. Hall (Int J Radiat Oncol Biol Phys 2006;65:1-7). , 2006, International journal of radiation oncology, biology, physics.

[35]  Mark J Rivard,et al.  Comparison of organ doses for patients undergoing balloon brachytherapy of the breast with HDR 192Ir or electronic sources using monte carlo simulations in a heterogeneous human phantom. , 2010, Medical physics.

[36]  Harald Paganetti,et al.  A review of dosimetry studies on external-beam radiation treatment with respect to second cancer induction , 2008, Physics in medicine and biology.

[37]  M. Butson,et al.  Energy dependence corrections to MOSFET dosimetric sensitivity , 2009, Australasian Physics & Engineering Sciences in Medicine.

[38]  A. Edwards,et al.  Neutron RBE values and their relationship to judgements in radiological protection. , 1999, Journal of radiological protection : official journal of the Society for Radiological Protection.

[39]  I Kawrakow,et al.  Monte Carlo simulation and patient dosimetry for a kilovoltage cone-beam CT unit. , 2009, Medical physics.

[40]  L. Dewerd,et al.  Response of an implantable MOSFET dosimeter to Ir192 HDR radiation. , 2008, Medical physics.

[41]  D. Lefkopoulos,et al.  Experimental assessment of out‐of‐field dose components in high energy electron beams used in external beam radiotherapy , 2015, Journal of applied clinical medical physics.

[42]  W. Newhauser,et al.  Calculations of neutron dose equivalent exposures from range-modulated proton therapy beams , 2005, Physics in medicine and biology.

[43]  F. Gómez,et al.  A new active method for the measurement of slow-neutron fluence in modern radiotherapy treatment rooms , 2010, Physics in medicine and biology.

[44]  K. Langen,et al.  Measurement of the tissue to A-150 tissue equivalent plastic kerma ratio at two p(66)Be neutron therapy facilities. , 2003, Physics in medicine and biology.

[45]  Eric E Klein,et al.  Peripheral doses from pediatric IMRT. , 2006, Medical physics.

[46]  B. Bednarz,et al.  A STUDY OF THE SHIELDING USED TO REDUCE LEAKAGE AND SCATTERED RADIATION TO THE FETUS IN A PREGNANT PATIENT TREATED WITH A 6-MV EXTERNAL X-RAY BEAM , 2009, Health physics.

[47]  Wesley E Bolch,et al.  Age-dependent organ and effective dose coefficients for external photons: a comparison of stylized and voxel-based paediatric phantoms , 2006, Physics in medicine and biology.

[48]  R. Pollock,et al.  Factors Affecting the Response of the Bubble Detector BD-100 and a Comparison of its Response to CR-39 , 1988 .

[49]  S. Kry,et al.  Measurement of High-Energy Neutron Spectra with a Bonner Sphere Extension System , 2009 .

[50]  Shuichi Ozawa,et al.  A dose comparison study between XVI and OBI CBCT systems. , 2008, Medical physics.

[51]  L. Dewerd,et al.  Microionization chamber air-kerma calibration coefficients as a function of photon energy for x-ray spectra in the range of 20-250 kVp relative to 60Co. , 2013, Medical physics.

[52]  Michael L. Taylor,et al.  Assessment of out-of-field doses in radiotherapy of brain lesions in children. , 2011, International journal of radiation oncology, biology, physics.

[53]  S Agosteo,et al.  Secondary neutron and photon dose in proton therapy. , 1998, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[54]  B. Poppe,et al.  Experimental study on photon-beam peripheral doses, their components and some possibilities for their reduction , 2010, Physics in medicine and biology.

[55]  Niko Papanikolaou,et al.  Flattening filter‐free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group , 2015, Journal of applied clinical medical physics.

[56]  S. Chiu‐Tsao,et al.  Use of new radiochromic devices for peripheral dose measurement: potential in-vivo dosimetry application , 2009, Biomedical imaging and intervention journal.

[57]  J. DeMarco,et al.  Shielding implications for secondary neutrons and photons produced within the patient during IMPT. , 2013, Medical physics.

[58]  A. B. Brill,et al.  NCRP Report No. 116, Limitation of Exposure to Ionizing Radiation National Council on Radiation Protection, Bethesda, MD, 88 pages, $22.50 , 1994 .

[59]  D. Followill,et al.  Variations in photon energy spectra of a 6 MV beam and their impact on TLD response. , 2011, Medical physics.

[60]  K. Kase,et al.  Neutron fluence and energy spectra around the Varian Clinac 2100C/2300C medical accelerator. , 1998, Health physics.

[61]  Ronald I. Ewing,et al.  A new type of neutron spectrometer , 1960 .

[62]  Jack Valentin,et al.  Biological effects after prenatal irradiation (embryo and fetus) , 2003 .

[63]  Brian Wang,et al.  Monte Carlo modeling of a high-sensitivity MOSFET dosimeter for low- and medium-energy photon sources. , 2004, Medical physics.

[64]  W. Rühm,et al.  Iterative unfolding for Bonner sphere spectrometers using the MSANDB code – Sensitivity analysis and dose calculation , 2010 .

[65]  J. F. Briesmeister MCNP-A General Monte Carlo N-Particle Transport Code , 1993 .

[66]  D. Followill,et al.  Effects of tertiary MLC configuration on secondary neutron spectra from 18 MV x-ray beams for the Varian 21EX linear accelerator. , 2009, Medical physics.

[67]  T. Kanai,et al.  Measurement of absorbed dose, quality factor, and dose equivalent in water phantom outside of the irradiation field in passive carbon-ion and proton radiotherapies. , 2010, Medical physics.

[68]  R. Macklis,et al.  A mobile shield to reduce scatter radiation to the contralateral breast during radiotherapy for breast cancer: preclinical results. , 1999, International journal of radiation oncology, biology, physics.

[69]  Timothy C. Zhu,et al.  Diode in Vivo Dosimetry for Patients Receiving External Beam Radiation Therapy , 2005 .

[70]  M. Hudson,et al.  Chronic health conditions in adult survivors of childhood cancer. , 2006, The New England journal of medicine.

[71]  A. Nahum,et al.  Effect of size and composition of the central electrode on the response of cylindrical ionization chambers in high-energy photon and electron beams , 1993 .

[72]  Susanna Guatelli,et al.  Tissue equivalency of phantom materials for neutron dosimetry in proton therapy. , 2009, Medical physics.

[73]  K. Burn,et al.  Neutron spectra in a tissue equivalent phantom during photon radiotherapy treatment by LINACS. , 2004, Radiation protection dosimetry.

[74]  H. Paganetti,et al.  Comparison of whole-body phantom designs to estimate organ equivalent neutron doses for secondary cancer risk assessment in proton therapy , 2012, Physics in medicine and biology.

[75]  Jacques Ferlay,et al.  Estimates of the cancer burden in Europe from radioactive fallout from the Chernobyl accident , 2006, International journal of cancer.

[76]  An Liu,et al.  Targeted total marrow irradiation using three-dimensional image-guided tomographic intensity-modulated radiation therapy: an alternative to standard total body irradiation. , 2006, Biology of Blood and Marrow Transplantation.

[77]  G T Chen,et al.  Online repositioning during treatment of the prostate: a study of potential limits and gains. , 1993, International journal of radiation oncology, biology, physics.

[78]  Radhe Mohan,et al.  Dosimetric properties of photon beams from a flattening filter free clinical accelerator , 2005, Physics in medicine and biology.

[79]  S. Darby,et al.  Cardiac risks of breast-cancer radiotherapy: a contemporary view. , 2006, Clinical oncology (Royal College of Radiologists (Great Britain)).

[80]  D. Followill,et al.  Energy response of optically stimulated luminescent dosimeters for non-reference measurement locations in a 6 MV photon beam , 2012, Physics in medicine and biology.

[81]  Harald Paganetti,et al.  Assessment of organ-specific neutron equivalent doses in proton therapy using computational whole-body age-dependent voxel phantoms , 2008, Physics in medicine and biology.

[82]  Barbara Kaser-Hotz,et al.  Estimation of radiation-induced cancer from three-dimensional dose distributions: Concept of organ equivalent dose. , 2005, International journal of radiation oncology, biology, physics.

[83]  I. M. Jones,et al.  DNA damage among thyroid cancer and multiple cancer cases, controls, and long-lived individuals. , 2005, Mutation research.

[84]  C. Land,et al.  Coronary heart disease after radiotherapy for peptic ulcer disease. , 2005, International journal of radiation oncology, biology, physics.

[85]  J. Bourhis,et al.  Role of cancer treatment in long-term overall and cardiovascular mortality after childhood cancer. , 2010, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[86]  L. Pierce,et al.  Contralateral breast dose after whole-breast irradiation: an analysis by treatment technique. , 2012, International journal of radiation oncology, biology, physics.

[87]  K. F. Eckerman,et al.  Specific absorbed fractions of energy at various ages from internal photon sources: 6, Newborn , 1987 .

[88]  J. Zoetelief,et al.  Assessment of induction of secondary tumours due to various radiotherapy modalities. , 2006, Radiation protection dosimetry.

[89]  Ping Xia,et al.  Investigation of the use of MOSFET for clinical IMRT dosimetric verification. , 2002, Medical physics.

[90]  Asghar Mesbahi,et al.  Dosimetric characteristics of unflattened 6 MV photon beams of a clinical linear accelerator: a Monte Carlo study. , 2007, Applied Radiation and Isotopes.

[91]  D. Rogers,et al.  Monte Carlo calculations of kQ, the beam quality conversion factor. , 2010, Medical physics.

[92]  S. Roy,et al.  Shielding for neutron scattered dose to the fetus in patients treated with 18 MV x-ray beams. , 2000, Medical physics.

[93]  W. S. Snyder,et al.  Estimates of absorbed fractions for monoenergetic photon sources uniformly distributed in various organs of a heterogeneous phantom. , 1974, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[94]  P. V. D. Giessen Peridose, a software program to calculate the dose outside the primary beam in radiation therapy , 2001 .

[95]  K. Parodi,et al.  Systematic out‐of‐field secondary neutron spectrometry and dosimetry in pencil beam scanning proton therapy , 2017, Medical physics.

[96]  W. Reece,et al.  EFFECT OF TORSO ADIPOSE TISSUE THICKNESS ON EFFECTIVE DOSE IN A BROAD PARALLEL PHOTON BEAM , 2003, Health physics.

[97]  J. Sjöstrand,et al.  A clinical study of radiation cataract formation in adult life following γ irradiation of the lens in early childhood , 1997 .

[98]  C R Blackwell,et al.  Fetal dose from radiotherapy with photon beams: report of AAPM Radiation Therapy Committee Task Group No. 36. , 1995, Medical physics.

[99]  Chengyu Shi,et al.  A boundary-representation method for designing whole-body radiation dosimetry models: pregnant females at the ends of three gestational periods—RPI-P3, -P6 and -P9 , 2007, Physics in medicine and biology.

[100]  L. Verhey,et al.  Peripheral dose in ocular treatments with CyberKnife and Gamma Knife radiosurgery compared to proton radiotherapy , 2007, Physics in medicine and biology.

[101]  R. Howell,et al.  Secondary neutron spectrum from 250-MeV passively scattered proton therapy: measurement with an extended-range Bonner sphere system. , 2014, Medical physics.

[102]  R. Howell,et al.  Risk of second malignant neoplasm following proton versus intensity-modulated photon radiotherapies for hepatocellular carcinoma , 2010, Physics in medicine and biology.

[103]  R. Mohan,et al.  The risk of developing a second cancer after receiving craniospinal proton irradiation , 2009, Physics in medicine and biology.

[104]  D. Gomez,et al.  Radiotherapy-Induced Malfunction in Contemporary Cardiovascular Implantable Electronic Devices: Clinical Incidence and Predictors. , 2015, JAMA oncology.

[105]  D Followill,et al.  Estimates of whole-body dose equivalent produced by beam intensity modulated conformal therapy. , 1997, International journal of radiation oncology, biology, physics.

[106]  B. Bednarz,et al.  Monte Carlo modeling of a 6 and 18 MV Varian Clinac medical accelerator for in-field and out-of-field dose calculations: development and validation , 2009, Physics in medicine and biology.

[107]  R. Mohan,et al.  Monte Carlo study of neutron dose equivalent during passive scattering proton therapy , 2007, Physics in medicine and biology.

[108]  T C Chu,et al.  Monte Carlo simulation of a clinical linear accelerator. , 2001, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[109]  R. Peto,et al.  Long-term mortality from heart disease and lung cancer after radiotherapy for early breast cancer: prospective cohort study of about 300,000 women in US SEER cancer registries. , 2005, The Lancet. Oncology.

[110]  Jean Chavaudra,et al.  Frequency distribution of second solid cancer locations in relation to the irradiated volume among 115 patients treated for childhood cancer. , 2009, International journal of radiation oncology, biology, physics.

[111]  L. Robison,et al.  Radiation dose and breast cancer risk in the childhood cancer survivor study. , 2009, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[112]  J. Gómez-Ros,et al.  TWO NEW SINGLE-EXPOSURE, MULTI-DETECTOR NEUTRON SPECTROMETERS FOR RADIATION PROTECTION APPLICATIONS IN WORKPLACE MONITORING , 2017, Radiation protection dosimetry.

[113]  Torunn I Yock,et al.  Assessment of radiation-induced second cancer risks in proton therapy and IMRT for organs inside the primary radiation field , 2012, Physics in medicine and biology.

[114]  A. Ferrari,et al.  A Neutron Survey Meter with Sensitivity Extended up to 400 MeV , 1992 .

[115]  S Mutic,et al.  Whole-body dose from tomotherapy delivery. , 1998, International journal of radiation oncology, biology, physics.

[116]  E. Bezak,et al.  Out-of-Field Neutron and Leakage Photon Exposures and the Associated Risk of Second Cancers in High-Energy Photon Radiotherapy: Current Status , 2011, Radiation research.

[117]  C. Muirhead,et al.  Radiation Cataractogenesis: A Review of Recent Studies , 2009, Radiation research.

[118]  K. Leszczynski,et al.  Static and rotational step-and-shoot IMRT treatment plans for the prostate: a risk comparison study. , 2012, Medical physics.

[119]  M. Silari,et al.  Sensitivity study of CR39 track detector in an extended range Bonner sphere spectrometer. , 2007, Radiation protection dosimetry.

[120]  R. Howell,et al.  Energy Response and Angular Dependence of a Bonner Sphere Extension , 2009, IEEE Transactions on Nuclear Science.

[121]  R. Wakeford Cancer risk modelling and radiological protection , 2012, Journal of radiological protection : official journal of the Society for Radiological Protection.

[122]  F. Sánchez-Doblado,et al.  Neutron contamination in radiotherapy: estimation of second cancers based on measurements in 1377 patients. , 2013, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[123]  Gary D Fullerton,et al.  Calculation of effective dose from measurements of secondary neutron spectra and scattered photon dose from dynamic MLC IMRT for 6 MV, 15 MV, and 18 MV beam energies. , 2006, Medical physics.

[124]  D. Kuban,et al.  Uncertainty of calculated risk estimates for secondary malignancies after radiotherapy. , 2007, International journal of radiation oncology, biology, physics.

[125]  Ping Xia,et al.  Skin toxicity due to intensity-modulated radiotherapy for head-and-neck carcinoma. , 2002, International journal of radiation oncology, biology, physics.

[126]  G. Hartmann,et al.  Neutron spectrometry and determination of neutron ambient dose equivalents in different LINAC radiotherapy rooms , 2010 .

[127]  F. Lohr,et al.  Volumetric modulated arc therapy (VMAT) vs. serial tomotherapy, step-and-shoot IMRT and 3D-conformal RT for treatment of prostate cancer. , 2009, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[128]  P. Vaz,et al.  Comparison of unfolding codes for neutron spectrometry with Bonner spheres. , 2014, Radiation protection dosimetry.

[129]  R. Howell,et al.  Comparison of risk of radiogenic second cancer following photon and proton craniospinal irradiation for a pediatric medulloblastoma patient , 2013, Physics in medicine and biology.

[130]  A. Sigurdson,et al.  Radiation-related risk of basal cell carcinoma: a report from the Childhood Cancer Survivor Study. , 2012, Journal of the National Cancer Institute.

[131]  D. Lefkopoulos,et al.  Risk of a second malignant neoplasm after cancer in childhood treated with radiotherapy: correlation with the integral dose restricted to the irradiated fields. , 2008, International journal of radiation oncology, biology, physics.

[132]  R. Howell,et al.  Analytical model for out-of-field dose in photon craniospinal irradiation , 2013, Physics in medicine and biology.

[133]  M. Sontag,et al.  Management of radiation oncology patients with implanted cardiac pacemakers: report of AAPM Task Group No. 34. American Association of Physicists in Medicine. , 1994, Medical physics.

[134]  J. Boice,et al.  Dose to the contralateral breast from radiotherapy and risk of second primary breast cancer in the WECARE study. , 2008, International journal of radiation oncology, biology, physics.

[135]  U. Schneider,et al.  Field calibration of PADC track etch detectors for local neutron dosimetry in man using different radiation qualities , 2012 .

[136]  S. Kry,et al.  Effect of organ size and position on out-of-field dose distributions during radiation therapy , 2010, Physics in medicine and biology.

[137]  D. Brenner,et al.  Second malignancies in prostate carcinoma patients after radiotherapy compared with surgery , 2000, Cancer.

[138]  Henry I. Kohn,et al.  Health Effects of Exposure to Low Levels of Ionizing Radiation (BEIR V) , 1990 .

[139]  T. Zhu,et al.  Temperature dependence of commercially available diode detectors. , 2002, Medical physics.

[140]  Andrew Nisbet,et al.  Cardiac exposures in breast cancer radiotherapy: 1950s-1990s. , 2007, International journal of radiation oncology, biology, physics.

[141]  Marta Paiusco,et al.  IMRT treatment planning:- a comparative inter-system and inter-centre planning exercise of the ESTRO QUASIMODO group. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[142]  D. Kuban,et al.  The calculated risk of fatal secondary malignancies from intensity-modulated radiation therapy. , 2005, International journal of radiation oncology, biology, physics.

[143]  E. Pedroni,et al.  Secondary neutron dose during proton therapy using spot scanning. , 2002, International journal of radiation oncology, biology, physics.

[144]  G. Hartmann,et al.  Estimation of neutron-equivalent dose in organs of patients undergoing radiotherapy by the use of a novel online digital detector , 2012, Physics in medicine and biology.

[145]  L T Chylack,et al.  Radiation Cataractogenesis: Epidemiology and Biology , 2010, Radiation research.

[146]  P. Hall,et al.  Mortality from cardiovascular disease more than 10 years after radiotherapy for breast cancer: nationwide cohort study of 90 000 Swedish women , 2003, BMJ : British Medical Journal.

[147]  Ryan L. Smith,et al.  A comparison of out-of-field dose and its constituent components for intensity-modulated radiation therapy versus conformal radiation therapy: implications for carcinogenesis. , 2011, International journal of radiation oncology, biology, physics.

[148]  W. Newhauser,et al.  Monte Carlo simulations of neutron spectral fluence, radiation weighting factor and ambient dose equivalent for a passively scattered proton therapy unit , 2008, Physics in medicine and biology.

[149]  J. Gómez-Ros,et al.  Experimental test of a newly developed single-moderator, multi-detector, directional neutron spectrometer in reference monochromatic fields from 144keV to 16.5MeV , 2015 .

[150]  R. Shore,et al.  Postoperative Cataract Cases among Atomic Bomb Survivors: Radiation Dose Response and Threshold , 2007, Radiation research.

[151]  R. Collins,et al.  Effects of radiotherapy and of differences in the extent of surgery for early breast cancer on local recurrence and 15-year survival: an overview of the randomised trials , 2005, The Lancet.

[152]  Andreu Badal,et al.  Monte Carlo simulation of MOSFET detectors for high-energy photon beams using the PENELOPE code , 2007, Physics in medicine and biology.

[153]  P. Olko,et al.  Photon dosimetry methods outside the target volume in radiation therapy: Optically stimulated luminescence (OSL), thermoluminescence (TL) and radiophotoluminescence (RPL) dosimetry , 2013 .

[154]  S. Mutic,et al.  Peripheral dose distributions for a linear accelerator equipped with a secondary multileaf collimator and universal wedge , 2002, Journal of applied clinical medical physics.

[155]  Atsushi Minamoto,et al.  A REANALYSIS OF ATOMIC-BOMB CATARACT DATA, 2000–2002: A THRESHOLD ANALYSIS , 2006, Health physics.

[156]  Brian Wang,et al.  Simulation of organ-specific patient effective dose due to secondary neutrons in proton radiation treatment , 2005, Physics in medicine and biology.

[157]  B Wiegel,et al.  NEMUS—the PTB Neutron Multisphere Spectrometer: Bonner spheres and more , 2002 .

[158]  A. Pérez-Andújar,et al.  An analytic model of neutron ambient dose equivalent and equivalent dose for proton radiotherapy , 2010, Physics in medicine and biology.

[159]  Dan Isaacs,et al.  Measured Neutron Spectra and Dose Equivalents From a Mevion Single-Room, Passively Scattered Proton System Used for Craniospinal Irradiation. , 2016, International journal of radiation oncology, biology, physics.

[160]  H. R. Andrews,et al.  Response of the bubble detector to neutrons of various energies. , 2015, Radiation protection dosimetry.

[161]  S. Chiu‐Tsao,et al.  Energy Dependence of the New Gafchromic EBT3 Film: Dose Response Curves for 50 kV, 6 and 15 MV X-Ray Beams * , 2012 .

[162]  G. Fullerton,et al.  Investigation of secondary neutron dose for 18 MV dynamic MLC IMRT delivery. , 2005, Medical physics.

[163]  W. Newhauser,et al.  A simple and fast physics-based analytical method to calculate therapeutic and stray doses from external beam, megavoltage x-ray therapy , 2015, Physics in medicine and biology.

[164]  J. Chow,et al.  Peripheral dose outside applicators in electron beams , 2006, Physics in medicine and biology.

[165]  H. Paganetti,et al.  Risk of developing second cancer from neutron dose in proton therapy as function of field characteristics, organ, and patient age. , 2008, International journal of radiation oncology, biology, physics.

[166]  Timothy C Zhu,et al.  Dose rate and SDD dependence of commercially available diode detectors. , 2004, Medical physics.

[167]  E. Klein,et al.  Surface and peripheral doses of dynamic and physical wedges. , 1997, International journal of radiation oncology, biology, physics.

[168]  Martin J Butson,et al.  Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range. , 2006, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[169]  Firas Mourtada,et al.  The use of LiF (TLD‐100) as an out‐of‐field dosimeter , 2007, Journal of applied clinical medical physics.

[170]  Oliver Jäkel,et al.  Radiation therapy with charged particles. , 2006, Seminars in radiation oncology.

[171]  K. Eckerman,et al.  Handbook of Anatomical Models for Radiation Dosimetry , 2009 .

[172]  H. G. Menzel,et al.  Calibration of Tissue-Equivalent Proportional Counters Used as Radiation Protection Dosemeters , 1984 .

[173]  P. Johnston,et al.  Stereotactic fields shaped with a micro-multileaf collimator: systematic characterization of peripheral dose. , 2010, Physics in medicine and biology.

[174]  Madan M Rehani,et al.  Risk for radiation‐induced cataract for staff in interventional cardiology: Is there reason for concern? , 2010, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[175]  R. Howell,et al.  Impact of margin size on the predicted risk of radiogenic second cancers following proton arc therapy and volumetric modulated arc therapy for prostate cancer , 2012, Physics in medicine and biology.

[176]  Jack Valentin,et al.  Relative biological effectiveness (RBE), quality factor (Q), and radiation weighting factor (wR) , 2003 .

[177]  R. Howell,et al.  Calibration of a Bonner sphere extension (BSE) for high-energy neutron spectrometry. , 2010, Radiation measurements.

[178]  Patrick Cheung,et al.  Assessing the role of volumetric modulated arc therapy (VMAT) relative to IMRT and helical tomotherapy in the management of localized, locally advanced, and post-operative prostate cancer. , 2011, International journal of radiation oncology, biology, physics.

[179]  R. Kudchadker,et al.  An evaluation of radiation exposure from portal films taken during definitive course of pediatric radiotherapy. , 2003, International journal of radiation oncology, biology, physics.

[180]  M. Butson,et al.  Dose response of various radiation detectors to synchrotron radiation. , 1998, Physics in medicine and biology.

[181]  Uwe Oelfke,et al.  Proton versus photon radiotherapy for common pediatric brain tumors: Comparison of models of dose characteristics and their relationship to cognitive function , 2008, Pediatric blood & cancer.

[182]  S. Kry,et al.  Out‐of‐field doses and neutron dose equivalents for electron beams from modern Varian and Elekta linear accelerators , 2016, Journal of applied clinical medical physics.

[183]  S. Kry,et al.  Methodology for determining doses to in-field, out-of-field and partially in-field organs for late effects studies in photon radiotherapy , 2010, Physics in medicine and biology.

[184]  K. Okudaira,et al.  Monte Carlo simulation and measurement of radiation leakage from applicators used in external electron radiotherapy. , 2013, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[185]  H. R. Andrews,et al.  Review of bubble detector response characteristics and results from space. , 2012, Radiation protection dosimetry.

[186]  J. Lehmann,et al.  Film dosimetry in the peripheral region using multiple sensitometric curves. , 2004, Medical Physics (Lancaster).

[187]  H. Ing,et al.  Bubble detectors—A maturing technology , 1997 .

[188]  R. Bogaerts,et al.  Out-of-field contributions for IMRT and volumetric modulated arc therapy measured using gafchromic films and compared to calculations using a superposition/convolution based treatment planning system. , 2012, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[189]  Uwe Schneider,et al.  Potential reduction of the incidence of radiation-induced second cancers by using proton beams in the treatment of pediatric tumors. , 2002, International journal of radiation oncology, biology, physics.

[190]  B. Bednarz,et al.  A feasibility study to calculate unshielded fetal doses to pregnant patients in 6-MV photon treatments using Monte Carlo methods and anatomically realistic phantoms. , 2008, Medical physics.

[191]  S. Mutic,et al.  A reduction in the AAPM TG-36 reported peripheral dose distributions with tertiary multileaf collimation. American Association of Physicists in Medicine Task Group 36. , 1999, International journal of radiation oncology, biology, physics.

[192]  J. Pena,et al.  Monte Carlo study of Siemens PRIMUS photoneutron production , 2005, Physics in medicine and biology.

[193]  Harald Paganetti,et al.  Reduction of the secondary neutron dose in passively scattered proton radiotherapy, using an optimized pre-collimator/collimator , 2009, Physics in medicine and biology.

[194]  John M. Boone,et al.  Comprehensive Methodology for the Evaluation of Radiation Dose in X-Ray Computed Tomography , 2010 .

[195]  Icrp Recommendations of the International Commission on Radiological Protection Publication 60 , 1991 .

[196]  R. Mohan,et al.  Monte Carlo simulations of stray neutron radiation exposures in proton therapy , 2007 .

[197]  G. Starkschall,et al.  Evaluation of the accuracy of fetal dose estimates using TG-36 data. , 2007, Medical Physics (Lancaster).

[198]  R. Stewart,et al.  Neutron scattered dose equivalent to a fetus from proton radiotherapy of the mother. , 2006, Medical physics.

[199]  W. Dries,et al.  Measurement and calculation of the dose at large distances from brachytherapy sources: Cs-137, Ir-192, and Co-60. , 1996, Medical physics.

[200]  Radhe Mohan,et al.  Stray radiation dose and second cancer risk for a pediatric patient receiving craniospinal irradiation with proton beams , 2009, Physics in medicine and biology.

[201]  D. Followill,et al.  Monte Carlo study shows no significant difference in second cancer risk between 6- and 18-MV intensity-modulated radiation therapy. , 2009, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[202]  E. Burgett A broad spectrum neutron spectrometer utilizing a high energy Bonner sphere extension , 2008 .

[203]  Chester R. Ramsey,et al.  Out‐of‐field dosimetry measurements for a helical tomotherapy system† , 2006, Journal of applied clinical medical physics.

[204]  I. Kawrakow Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version. , 2000, Medical physics.

[205]  D. Followill,et al.  Neutron spectra and dose equivalents calculated in tissue for high-energy radiation therapy. , 2009, Medical physics.

[206]  M. Kreuzer,et al.  Mortality from cardiovascular diseases in the German uranium miners cohort study, 1946–1998 , 2006, Radiation and Environmental Biophysics.

[207]  V. Smith,et al.  Peripheral dose measurement for CyberKnife radiosurgery with upgraded linac shielding. , 2008, Medical physics.

[208]  V. Smith,et al.  Peripheral doses in CyberKnife radiosurgery. , 2006, Medical physics.

[209]  David J. Thomas,et al.  Bonner sphere spectrometers—a critical review , 2002 .

[210]  Uwe Titt,et al.  Monte Carlo investigation of collimator scatter of proton-therapy beams produced using the passive scattering method. , 2008, Physics in medicine and biology.

[211]  Andrew K. Lee,et al.  Reducing stray radiation dose to patients receiving passively scattered proton radiotherapy for prostate cancer , 2008, Physics in medicine and biology.

[212]  Yan Liu,et al.  New primary neoplasms of the central nervous system in survivors of childhood cancer: a report from the Childhood Cancer Survivor Study. , 2006, Journal of the National Cancer Institute.

[213]  Radhe Mohan,et al.  Energy spectra, sources, and shielding considerations for neutrons generated by a flattening filter-free Clinac. , 2008, Medical physics.

[214]  R. Nath,et al.  Superheated drop detector for determination of neutron dose equivalent to patients undergoing high-energy x-ray and electron radiotherapy. , 1993, Medical physics.

[215]  D. Hunt,et al.  Quantification and reduction of peripheral dose from leakage radiation on Siemens primus accelerators in electron therapy mode , 2010, Journal of applied clinical medical physics.

[216]  Shuang Luan,et al.  Comparing radiation treatments using intensity-modulated beams, multiple arcs, and single arcs. , 2010, International journal of radiation oncology, biology, physics.

[217]  M. Griem Neutron Contamination From Medical Electron Accelerators , 1985 .

[218]  D. Followill,et al.  Accuracy and sources of error of out‐of field dose calculations by a commercial treatment planning system for intensity‐modulated radiation therapy treatments , 2013, Journal of applied clinical medical physics.

[219]  J. Galvin,et al.  Comparative dose evaluations between XVI and OBI cone beam CT systems using Gafchromic XRQA2 film and nanoDot optical stimulated luminescence dosimeters. , 2013, Medical physics.

[220]  P. Cappellaro,et al.  Development of an Extended Range Bonner Sphere Spectrometer , 2001 .

[221]  Fang-Fang Yin,et al.  Radiotherapy treatment plans with RapidArc for prostate cancer involving seminal vesicles and lymph nodes. , 2010, International journal of radiation oncology, biology, physics.

[222]  P. Munro,et al.  Reducing radiation exposure to patients from kV-CBCT imaging. , 2010, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[223]  H. Prentice,et al.  Incidence of cataracts after single fraction total body irradiation: the role of steroids and graft versus host disease. , 1993, Bone marrow transplantation.

[224]  J. Mcdonald,et al.  Kerma factor of carbon for 14.1-MeV neutrons. , 1984, Radiation research.

[225]  R. Stern Peripheral dose from a linear accelerator equipped with multileaf collimation. , 1999, Medical physics.

[226]  G. Merriam,et al.  The Effects of Ionizing Radiations on the Eye1 , 1972 .

[227]  Bingshu E. Chen,et al.  Second cancers among 104,760 survivors of cervical cancer: evaluation of long-term risk. , 2007, Journal of the National Cancer Institute.

[228]  A. Pérez-Andújar,et al.  Microdosimetric measurements for neutron-absorbed dose determination during proton therapy. , 2012, Radiation protection dosimetry.

[229]  Anatoly Rosenfeld,et al.  Assessment of out-of-field absorbed dose and equivalent dose in proton fields. , 2009, Medical physics.

[230]  N Shandala,et al.  Scope of radiological protection control measures. , 2007, Annals of the ICRP.

[231]  X. Xu,et al.  Measurements of non-target organ doses using MOSFET dosemeters for selected IMRT and 3D CRT radiation treatment procedures. , 2007, Radiation protection dosimetry.

[232]  G. Ozyigit,et al.  Investigating in-field and out-of-field neutron contamination in high-energy medical linear accelerators based on the treatment factors of field size, depth, beam modifiers, and beam type. , 2015, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[233]  A J Lomax,et al.  A treatment planning inter-comparison of proton and intensity modulated photon radiotherapy. , 1999, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[234]  P F O'Brien,et al.  Radiosurgery with unflattened 6-MV photon beams. , 1991, Medical physics.

[235]  K. Otto,et al.  Volumetric modulated arc therapy for delivery of prostate radiotherapy: comparison with intensity-modulated radiotherapy and three-dimensional conformal radiotherapy. , 2008, International journal of radiation oncology, biology, physics.

[236]  Stephen F. Kry,et al.  Skin dose during radiotherapy: a summary and general estimation technique , 2012, Journal of applied clinical medical physics.

[237]  K. Kase,et al.  Measurements of dose from secondary radiation outside a treatment field: effects of wedges and blocks. , 1985, International journal of radiation oncology, biology, physics.

[238]  Uwe Titt,et al.  Monte Carlo simulations of a nozzle for the treatment of ocular tumours with high-energy proton beams , 2005, Physics in medicine and biology.

[239]  Å. Palm,et al.  A review of the impact of photon and proton external beam radiotherapy treatment modalities on the dose distribution in field and out-of-field; implications for the long-term morbidity of cancer survivors , 2007, Acta oncologica.

[240]  R. Mohan,et al.  A Monte Carlo model for calculating out-of-field dose from a Varian 6MV beam. , 2006, Medical physics.

[241]  E. Hall,et al.  Intensity-modulated radiation therapy, protons, and the risk of second cancers. , 2006, International journal of radiation oncology, biology, physics.

[242]  X. Xu,et al.  RPI-AM and RPI-AF, a pair of mesh-based, size-adjustable adult male and female computational phantoms using ICRP-89 parameters and their calculations for organ doses from monoenergetic photon beams , 2009, Physics in medicine and biology.

[243]  K. Manton,et al.  THE RISK OF RADIATION-INDUCED CEREBROVASCULAR DISEASE IN CHERNOBYL EMERGENCY WORKERS , 2006, Health physics.

[244]  L. Robison,et al.  Endocrine and cardiovascular late effects among adult survivors of childhood brain tumors , 2003, Cancer.

[245]  F. Bochud,et al.  Evaluation of organ-specific peripheral doses after 2-dimensional, 3-dimensional and hybrid intensity modulated radiation therapy for breast cancer based on Monte Carlo and convolution/superposition algorithms: implications for secondary cancer risk assessment. , 2013, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[246]  F. Ballester,et al.  Calculated organ doses using Monte Carlo simulations in a reference male phantom undergoing HDR brachytherapy applied to localized prostate carcinoma. , 2013, Medical physics.

[247]  Whole body doses from linear accelerator-based stereotactic radiotherapy. , 1997, International journal of radiation oncology, biology, physics.

[248]  Heng Li,et al.  Dosimetric effects caused by couch tops and immobilization devices: report of AAPM Task Group 176. , 2014, Medical physics.

[249]  A. Waker Experimental uncertainties in microdosimetric measurements and an examination of the performance of three commercially produced proportional counters , 1985 .

[250]  R. Olsher,et al.  High-energy response of the PRESCILA and WENDI-II neutron rem meters. , 2008, Radiation protection dosimetry.

[251]  D. Georg,et al.  Rotational IMRT techniques compared to fixed gantry IMRT and Tomotherapy: multi-institutional planning study for head-and-neck cases , 2011, Radiation oncology.

[252]  M. Angelone,et al.  Performance of the UAB and the INFN-LNF Bonner sphere spectrometers in quasi monoenergetic neutron fields. , 2007, Radiation protection dosimetry.

[253]  A method to improve the effectiveness of diode in vivo dosimetry. , 1998, Medical physics.

[254]  Sarah C. Darby,et al.  Low Doses of Ionizing Radiation and Circulatory Diseases: A Systematic Review of the Published Epidemiological Evidence , 2005, Radiation research.

[255]  P. Hall,et al.  Risk of ischemic heart disease in women after radiotherapy for breast cancer. , 2013, The New England journal of medicine.

[256]  Anatoly Rosenfeld,et al.  Out-of-field dose equivalents delivered by proton therapy of prostate cancer. , 2007, Medical physics.

[257]  H. Paganetti,et al.  Comparison of second cancer risk due to out-of-field doses from 6-MV IMRT and proton therapy based on 6 pediatric patient treatment plans. , 2011, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[258]  Marilyn Stovall,et al.  Second neoplasms in survivors of childhood cancer: findings from the Childhood Cancer Survivor Study cohort. , 2009, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[259]  S. Marcié,et al.  Protection of testes during radiation treatment by irregular and focused fields of 25 MV x-rays: in vivo evaluation of the absorbed dose. , 1995, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[260]  G. Defraene,et al.  Companding technique for high dynamic range measurements using Gafchromic films. , 2011, Medical physics.

[261]  A. Koehler,et al.  Measurement of neutron dose equivalent to proton therapy patients outside of the proton radiation field , 2002 .

[262]  D Huyskens,et al.  Build-up modification of commercial diodes for entrance dose measurements in 'higher energy' photon beams. , 1999, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[263]  Hans Garmo,et al.  Distribution of coronary artery stenosis after radiation for breast cancer. , 2012, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[264]  Yukiko Shimizu,et al.  Studies of Mortality of Atomic Bomb Survivors. Report 13: Solid Cancer and Noncancer Disease Mortality: 1950–1997 , 2003, Radiation research.

[265]  Marcel Reginatto,et al.  Spectrum unfolding, sensitivity analysis and propagation of uncertainties with the maximum entropy deconvolution code MAXED , 2002 .

[266]  B. Bednarz,et al.  Calculated organ doses from selected prostate treatment plans using Monte Carlo simulations and an anatomically realistic computational phantom , 2009, Physics in medicine and biology.

[267]  John C. Roeske,et al.  Linac-Based Intensity Modulated Total Marrow Irradiation (IM-TMI) , 2006 .

[268]  Charles W. Coffey,et al.  Radiation dose from kilovoltage cone beam computed tomography in an image-guided radiotherapy procedure. , 2009, International journal of radiation oncology, biology, physics.

[269]  J. Dennis Relative biological effectiveness of neutron radiation and its implications for quality factor and dose limitation , 1987 .

[270]  W. Satariano,et al.  Risk factors for age-related cataracts. , 1995, Epidemiologic reviews.

[271]  R. Walstam,et al.  Radiation Induced Cataracts after Radium Therapy in Children , 1966 .

[272]  Lawrence B Marks,et al.  The incidence and functional consequences of RT-associated cardiac perfusion defects. , 2005, International journal of radiation oncology, biology, physics.

[273]  Tommy Knöös,et al.  Flattening filter free beams in SBRT and IMRT: dosimetric assessment of peripheral doses. , 2011, Zeitschrift fur medizinische Physik.

[274]  Wendy Leisenring,et al.  Cause-specific late mortality among 5-year survivors of childhood cancer: the Childhood Cancer Survivor Study. , 2008, Journal of the National Cancer Institute.

[275]  Stephanie Lamart,et al.  Proportion of second cancers attributable to radiotherapy treatment in adults: a cohort study in the US SEER cancer registries. , 2011, The Lancet. Oncology.

[276]  A. Niemierko,et al.  Secondary Carcinogenesis in Patients Treated with Radiation: A Review of Data on Radiation-Induced Cancers in Human, Non-human Primate, Canine and Rodent Subjects , 2007, Radiation research.

[277]  D. Georg,et al.  Multicenter evaluation of different target volume delineation concepts in pediatric Hodgkin’s lymphoma , 2012, Strahlentherapie und Onkologie.

[278]  G. Armstrong,et al.  Late mortality among 5-year survivors of childhood cancer: a summary from the Childhood Cancer Survivor Study. , 2009, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[279]  I. Toma-Dasu,et al.  The use of risk estimation models for the induction of secondary cancers following radiotherapy , 2005, Acta oncologica.

[280]  P. Olko,et al.  A comprehensive spectrometry study of a stray neutron radiation field in scanning proton therapy , 2016, Physics in medicine and biology.

[281]  Steve Webb,et al.  Clinical implementation of dynamic and step-and-shoot IMRT to treat prostate cancer with high risk of pelvic lymph node involvement. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[282]  G. Fullerton,et al.  Measurements of secondary neutron dose from 15 MV and 18 MV IMRT. , 2005, Radiation protection dosimetry.

[283]  M. Butson,et al.  Energy response of the new EBT2 radiochromic film to x-ray radiation , 2010 .

[284]  Radhe Mohan,et al.  A procedure to determine the planar integral spot dose values of proton pencil beam spots. , 2012, Medical physics.

[285]  V. Smyth,et al.  Comments on "Calculated response and wall correction factors for ionization chambers exposed to 60Co gamma-rays". , 1984, Medical physics.

[286]  Andrew K. Lee,et al.  Risk of secondary malignant neoplasms from proton therapy and intensity-modulated x-ray therapy for early-stage prostate cancer. , 2009, International journal of radiation oncology, biology, physics.

[287]  D. Followill,et al.  Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators , 2003, Journal of applied clinical medical physics.

[288]  M. Sontag,et al.  Management of Radiation Oncology Patients with Implanted Cardiac Pacemakers , 1994 .

[289]  Timothy C Zhu,et al.  Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM. , 2008, Medical physics.

[290]  U. Schneider,et al.  Systematic measurements of whole-body imaging dose distributions in image-guided radiation therapy. , 2012, Medical physics.

[291]  P. Olko,et al.  Measurement of stray radiation within a scanning proton therapy facility: EURADOS WG9 intercomparison exercise of active dosimetry systems. , 2014, Medical physics.

[292]  U. Schneider,et al.  A general model for stray dose calculation of static and intensity-modulated photon radiation. , 2016, Medical physics.

[293]  Influence of beam efficiency through the patient-specific collimator on secondary neutron dose equivalent in double scattering and uniform scanning modes of proton therapy. , 2010, Medical physics.

[294]  E. Benton,et al.  Leakage and scatter radiation from a double scattering based proton beamline. , 2007, Medical physics.

[295]  V. Mareš,et al.  Response functions of the Andersson–Braun and extended range rem counters for neutron energies from thermal to 10 GeV , 2002 .

[296]  D. Mulrooney,et al.  Cardiac outcomes in a cohort of adult survivors of childhood and adolescent cancer: retrospective analysis of the Childhood Cancer Survivor Study cohort , 2009, BMJ : British Medical Journal.

[297]  S. Woo,et al.  Reducing Stray Radiation Dose for a Pediatric Patient Receiving Proton Craniospinal Irradiation , 2008, Nuclear technology.

[298]  D P Dearnaley,et al.  Cardiovascular disease as a long-term complication of treatment for testicular cancer. , 2003, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[299]  S L Hancock,et al.  Factors affecting late mortality from heart disease after treatment of Hodgkin's disease. , 1993, JAMA.

[300]  A Joosten,et al.  Variability of a peripheral dose among various linac geometries for second cancer risk assessment , 2011, Physics in medicine and biology.

[301]  Jason Cashmore,et al.  The characterization of unflattened photon beams from a 6 MV linear accelerator , 2008, Physics in medicine and biology.

[302]  J. Ruben,et al.  The effect of intensity-modulated radiotherapy on radiation-induced second malignancies. , 2008, International journal of radiation oncology, biology, physics.

[303]  P. Olko,et al.  Performance tests and comparison of microdosimetric measurements with four tissue-equivalent proportional counters in scanning proton therapy , 2017 .

[304]  J. Manens,et al.  Evaluation And Analyze Of Out-of-field Doses In Head and Neck Radiation Therapy For Different Delivery Techniques: From 3DCRT To VMAT , 2011 .

[305]  A. Dell'Acqua,et al.  Geant4 - A simulation toolkit , 2003 .

[306]  Steven H. Lin,et al.  Dosimetric effects of jaw tracking in step‐and‐shoot intensity‐modulated radiation therapy , 2012, Journal of Applied Clinical Medical Physics.

[307]  J. Smathers,et al.  Book Review , 2006, Journal of Applied Clinical Medical Physics.

[308]  G. Mandarano,et al.  Biomedical Imaging and Intervention Journal the Sonographer's Role in Rfa Therapy of Liver Lesions , 2022 .

[309]  James A Purdy,et al.  DOSE TO NORMAL TISSUES OUTSIDE THE RADIATION THERAPY PATIENT’S TREATED VOLUME: A REVIEW OF DIFFERENT RADIATION THERAPY TECHNIQUES , 2008, Health physics.

[310]  J. Gómez-Ros,et al.  First test of SP2: A novel active neutron spectrometer condensing the functionality of Bonner spheres in a single moderator , 2014 .

[311]  D. Followill,et al.  Secondary neutron spectra from modern Varian, Siemens, and Elekta linacs with multileaf collimators. , 2009, Medical physics.

[312]  Anatoly Rosenfeld,et al.  RBE estimation of proton radiation fields using a ΔE-E telescope. , 2009, Medical physics.

[313]  Marcel Reginatto,et al.  Overview of spectral unfolding techniques and uncertainty estimation , 2010 .

[314]  M. Tubiana La prévention du cancer et la relation dose–effet : l’effet cancérogène des rayonnements ionisants ☆ , 2009 .

[315]  A. Mesbahi,et al.  Effect of wedge filter and field size on photoneutron dose equivalent for an 18MV photon beam of a medical linear accelerator. , 2010, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[316]  A. Pica,et al.  Modern cataract surgery for radiation-induced cataracts in retinoblastoma , 2010, British Journal of Ophthalmology.

[317]  John D Boice,et al.  Genetic effects of radiotherapy for childhood cancer: gonadal dose reconstruction. , 2004, International journal of radiation oncology, biology, physics.

[318]  C. Pui,et al.  Second malignant neoplasms and cardiovascular disease following radiotherapy. , 2012, Journal of the National Cancer Institute.

[319]  D. Followill,et al.  Estimate of the uncertainties in the relative risk of secondary malignant neoplasms following proton therapy and intensity-modulated photon therapy , 2010, Physics in medicine and biology.

[320]  K. Kase,et al.  Measurements of dose from secondary radiation outside a treatment field. , 1983, International journal of radiation oncology, biology, physics.

[321]  Susanta K Hui,et al.  Feasibility study of helical tomotherapy for total body or total marrow irradiationa). , 2005, Medical physics.

[322]  M. Ay,et al.  Monte Carlo estimation of photoneutrons contamination from high-energy X-ray medical accelerators in treatment room and maze: a simplified model. , 2009, Radiation protection dosimetry.

[323]  R. Mohan,et al.  Reduced neutron production through use of a flattening-filter-free accelerator. , 2007, International journal of radiation oncology, biology, physics.

[324]  Daniel Lodwick,et al.  The UF family of reference hybrid phantoms for computational radiation dosimetry , 2010, Physics in medicine and biology.

[325]  F. Sánchez-Doblado,et al.  Active on-line detector for in-room radiotherapy neutron measurements , 2010 .

[326]  J. Buatti,et al.  In vivo determination of extra-target doses received from serial tomotherapy. , 2002, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[327]  Marilyn Stovall,et al.  Dose Reconstruction for Therapeutic and Diagnostic Radiation Exposures: Use in Epidemiological Studies , 2006, Radiation research.

[328]  T Lindmo,et al.  Monte Carlo study of in-field and out-of-field dose distributions from a linear accelerator operating with and without a flattening-filter. , 2012, Medical physics.