A new laser Doppler flowmeter prototype for depth dependent monitoring of skin microcirculation.

Laser Doppler flowmetry (LDF) is now commonly used in clinical research to monitor microvascular blood flow. However, the dependence of the LDF signal on the microvascular architecture is still unknown. That is why we propose a new laser Doppler flowmeter for depth dependent monitoring of skin microvascular perfusion. This new laser Doppler flowmeter combines for the first time, in a device, several wavelengths and different spaced detection optical fibres. The calibration of the new apparatus is herein presented together with in vivo validation. Two in vivo validation tests are performed. In the first test, signals collected in the ventral side of the forearm are analyzed; in the second test, signals collected in the ventral side of the forearm are compared with signals collected in the hand palm. There are good indicators that show that different wavelengths and fibre distances probe different skin perfusion layers. However, multiple scattering may affect the results, namely the ones obtained with the larger fibre distance. To clearly understand the wavelength effect in LDF measurements, other tests have to be performed.

[1]  P. Öberg,et al.  Single-fibre laser Doppler flowmetry , 1987, Medical and Biological Engineering and Computing.

[2]  R. Anderson,et al.  The optics of human skin. , 1981, The Journal of investigative dermatology.

[3]  R. Maniewski,et al.  Multichannel laser-Doppler probe for blood perfusion measurements with depth discrimination , 1998, Medical and Biological Engineering and Computing.

[4]  R. Nossal,et al.  Model for laser Doppler measurements of blood flow in tissue. , 1981, Applied optics.

[5]  Rodney Gush,et al.  Evaluation of a new high power, wide separation laser Doppler probe: potential measurement of deeper tissue blood flow. , 2009, Microvascular research.

[6]  J. Lévêque,et al.  EEMCO Guidance for the Measurement of Skin Microcirculation , 2002, Skin Pharmacology and Physiology.

[7]  Ingemar Fredriksson,et al.  Optical microcirculatory skin model: assessed by Monte Carlo simulations paired with in vivo laser Doppler flowmetry. , 2008, Journal of biomedical optics.

[8]  Anne Humeau-Heurtier,et al.  Laser Doppler Flowmeters Prototypes Validation using Monte Carlo Simulations , 2012, BIOINFORMATICS.

[9]  T. King,et al.  Comparison of red and green laser doppler imaging of blood flow , 2004, Lasers in surgery and medicine.

[10]  T. King,et al.  Aspects of laser light scattering from skin tissue with application to laser Doppler blood flow measurement. , 1984, Physics in medicine and biology.

[11]  Edite Figueiras,et al.  Some Reasons to Build a New Laser Doppler Flowmeter to Monitor Microvascular Blood Flow , 2009 .

[12]  Karin Wårdell,et al.  Laser Doppler perfusion,monitoring and imaging , 2003 .

[13]  Anne Humeau-Heurtier,et al.  Comparison between Laser Doppler Flowmetry Signals Recorded in Glabrous and Non Glabrous Skin - Time and Frequency Analyses , 2010, BIOSIGNALS.

[14]  Henrik Nilsson,et al.  Laser Doppler perfusion monitoring and imaging: novel approaches , 2007, Medical & Biological Engineering & Computing.

[15]  T. King,et al.  Laser Doppler imaging: a developing technique for application in the rheumatic diseases. , 2004, Rheumatology.

[16]  Anne Humeau-Heurtier,et al.  Laser Doppler Flowmeters Prototypes: Monte Carlo Simulations Validation Paired with Measurements , 2012, BIOSTEC.

[17]  Ingemar Fredriksson,et al.  Measurement depth and volume in laser Doppler flowmetry. , 2009, Microvascular research.

[18]  Per Wollmer,et al.  The influence of wavelength and probe configuration on findings of a skin vasoconstriction test when using laser Doppler perfusion devices. , 2006, Microvascular research.