Modulation transfer function (MTF) evaluation for x-ray phase imaging system employing attenuation masks

Objective. Attenuation masks can be used in x-ray imaging systems to increase their inherent spatial resolution and/or make them sensitive to phase effects, a typical example being Edge Illumination x-ray phase contrast imaging (EI-XPCI). This work investigates the performance of a mask-based system such as EI-XPCI in terms of Modulation Transfer Function (MTF), in the absence of phase effects. Approach. Pre-sampled MTF measurements, using an edge, were performed on the same system implemented without masks, with non-skipped masks and finally with skipped masks (i.e. masks in which apertures illuminate every other pixel row/column). Results are compared to simulations and finally images of a resolution bar pattern acquired with all the above setups are presented. Main results. Compared to the detector’s inherent MTF, the non-skipped mask setup provides improved MTF results. In comparison to an ideal case where signal spill-out into neighbouring pixels is negligible, this improvement takes place only at specific frequencies of the MTF, dictated by the spatial repetition of the spill-out signal. This is limited with skipped masks, which indeed provide further MTF improvements over a larger frequency range. Experimental MTF measurements are supported through simulation and resolution bar pattern images. Significance. This work has quantified the improvement in MTF due to the use of attenuation masks and lays the foundation for how acceptance and routine quality control tests will have to be modified when systems using masks are introduced in clinical practice and how MTF results will compare to those of conventional imaging systems.

[1]  Oliver J. Larkin,et al.  Volumetric high-resolution X-ray phase-contrast virtual histology of breast specimens with a compact laboratory system , 2021, IEEE Transactions on Medical Imaging.

[2]  M. Endrizzi,et al.  Technical Note: Practical implementation strategies of cycloidal computed tomography. , 2021, Medical physics.

[3]  Oliver J. Larkin,et al.  Detection of involved margins in breast specimens with X-ray phase-contrast computed tomography , 2021, Scientific Reports.

[4]  Atsushi Momose,et al.  X-ray phase imaging reaching clinical uses. , 2020, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[5]  A. Olivo,et al.  Predicting the noise in hybrid (phase and attenuation) x-ray images acquired with the edge illumination technique. , 2020, Medical physics.

[6]  J. Jakubek,et al.  Virtual subpixel approach for single-mask phase-contrast imaging using Timepix3 , 2019, Journal of Instrumentation.

[7]  M. Endrizzi,et al.  Large field of view, fast and low dose multimodal phase-contrast imaging at high x-ray energy , 2017, Scientific Reports.

[8]  Alessandro Olivo,et al.  On the relative performance of edge illumination x‐ray phase‐contrast CT and conventional, attenuation‐based CT , 2017, Medical physics.

[9]  Feng Gao,et al.  MTF evaluation of in-line phase contrast imaging system , 2017, BiOS.

[10]  Alessandro Olivo,et al.  Concept of contrast transfer function for edge illumination x-ray phase-contrast imaging and its comparison with the free-space propagation technique. , 2016, Optics express.

[11]  Marco Endrizzi,et al.  Laboratory-based X-ray phase-contrast imaging with misaligned optical elements , 2015 .

[12]  P. C. Diemoz,et al.  Angular sensitivity and spatial resolution in edge illumination X-ray phase-contrast imaging , 2015 .

[13]  Charlotte K. Hagen,et al.  The effect of the spatial sampling rate on quantitative phase information extracted from planar and tomographic edge illumination x-ray phase contrast images , 2014 .

[14]  P. C. Diemoz,et al.  Low-dose phase contrast tomography with conventional x-ray sources. , 2014, Medical physics.

[15]  P. C. Diemoz,et al.  Spatial resolution of edge illumination X-ray phase-contrast imaging. , 2014, Optics express.

[16]  P. C. Diemoz,et al.  Sensitivity of laboratory based implementations of edge illumination X-ray phase-contrast imaging , 2013 .

[17]  Han Wen,et al.  Subnanoradian X-ray phase-contrast imaging using a far-field interferometer of nanometric phase gratings , 2013, Nature Communications.

[18]  Marco Endrizzi,et al.  Strategies for efficient and fast wave optics simulation of coded-aperture and other x-ray phase-contrast imaging methods. , 2013, Applied optics.

[19]  R D Speller,et al.  Low-dose phase contrast mammography with conventional x-ray sources. , 2013, Medical physics.

[20]  Guang-Hong Chen,et al.  Spatial resolution characterization of differential phase contrast CT systems via modulation transfer function (MTF) measurements , 2013, Physics in medicine and biology.

[21]  A Bravin,et al.  X-ray phase-contrast imaging with nanoradian angular resolution. , 2013, Physical review letters.

[22]  Paola Coan,et al.  X-ray phase-contrast imaging: from pre-clinical applications towards clinics , 2013, Physics in medicine and biology.

[23]  N. Allinson,et al.  Evaluation of a novel wafer-scale CMOS APS X-ray detector for use in mammography , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[24]  Alessandro Olivo,et al.  Phase and absorption retrieval using incoherent X-ray sources , 2012, Proceedings of the National Academy of Sciences.

[25]  Robert D. Speller,et al.  "Edge illumination" in X-ray Phase Contrast Imaging , 2012 .

[26]  R D Speller,et al.  Effects of signal diffusion on x-ray phase contrast images. , 2011, The Review of scientific instruments.

[27]  J. Jakubek,et al.  Hard x-ray phase contrast imaging using single absorption grating and hybrid semiconductor pixel detector. , 2010, The Review of scientific instruments.

[28]  P. Munro,et al.  Design and realization of a coded-aperture based X-ray phase contrast imaging for homeland security applications , 2009 .

[29]  R. Speller,et al.  A coded-aperture technique allowing x-ray phase contrast imaging with conventional sources , 2007 .

[30]  Andrew D. A. Maidment,et al.  Aliasing effects in digital images of line-pair phantoms. , 2002, Medical physics.

[31]  R. Adams Measurement of the Performance Characteristics of Diagnostic X-Ray Systems Used in Medicine: Part II, X-Ray Intensifier TelevisionSystems , 1983 .

[32]  Da Zhang,et al.  Step by step analysis toward optimal MTF algorithm using an edge test device. , 2009, Journal of X-ray science and technology.

[33]  E. Samei,et al.  A method for measuring the presampled MTF of digital radiographic systems using an edge test device. , 1998, Medical physics.