A Novel Method (CiMON) for Continuous Intra-Abdominal Pressure Monitoring: Pilot Test in a Pig Model

Background. Intravesical pressure (IAPivp) measurement is considered to be the gold standard for assessment of intra-abdominal pressure (IAP). This study evaluated a new minimally invasive IAP monitoring device (CiMON) against three other devices in a wide range of clinically relevant IAP and in different body positions in healthy pigs. Methods. The CiMON catheter (IAPCiM) and another balloon-tipped catheter (IAPspie) were positioned into the stomach. Fluid-filled catheters were used for direct intraperitoneal (IAPdir) and IAPivp measurement. Both in supine and 25° head-of-bed positions, IAP was increased from baseline to 30 mmHg. At every IAP level, 4 IAP measurements were recorded simultaneously. Mean differences and the limits of agreement were calculated. Results. Bias between IAPCiM and IAPspie was nearly zero with very good agreement, both in supine and 25° position. In supine position, IAPCiM slightly overestimated IAPivp and IAPdir by 1.5 and 2.1 mmHg with reasonable agreement. In 25° position, IAPCiM underestimated IAPivp and IAPdir by 1.0 and 0.5 mmHg, again with reasonable agreement. Conclusions. Agreement between IAPCiM and IAPspie was very good, while good-to-moderate agreement exists between IAPCiM and IAPdir or IAPivp. Simplicity, continuous monitoring, and the combination with a feeding tube should lead to further clinical studies, evaluating this new CiMON device.

[1]  W. Marsden I and J , 2012 .

[2]  A. Kirkpatrick,et al.  RECOMMENDATIONS FOR RESEARCH FROM THE INTERNATIONAL CONFERENCE OF EXPERTS ON INTRA-ABDOMINAL HYPERTENSION AND ABDOMINAL COMPARTMENT SYNDROME , 2009, Acta clinica Belgica.

[3]  E. Hoste,et al.  Semicontinuous intra-abdominal pressure measurement using an intragastric Compliance catheter , 2007, Intensive Care Medicine.

[4]  G. Mills,et al.  Intra-abdominal pressure measurement: validation of intragastric pressure as a measure of intra-abdominal pressure. , 2007, British journal of anaesthesia.

[5]  K. Hillman,et al.  Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. II. Recommendations , 2006, Intensive Care Medicine.

[6]  M. Malbrain,et al.  Continuous intra-abdominal pressure monitoring. , 2007, Acta clinica Belgica.

[7]  M. Malbrain,et al.  Rational intraabdominal pressure monitoring: how to do it? , 2007, Acta clinica Belgica.

[8]  M. Malbrain,et al.  ICU management of the patient with intra-abdominal hypertension: what to do, when and to whom? , 2007, Acta clinica Belgica.

[9]  B. Seifert,et al.  Measurement of compartment pressure of the rectus sheath during intra-abdominal hypertension in rats , 2006, Intensive Care Medicine.

[10]  A. Schachtrupp,et al.  Evaluation of a modified piezoresistive technique and a water-capsule technique for direct and continuous measurement of intra-abdominal pressure in a porcine model* , 2006, Critical care medicine.

[11]  K. Hillman,et al.  Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. I. Definitions , 2006, Intensive Care Medicine.

[12]  M. Malbrain,et al.  Intra- and interobserver variability during in vitro validation of two novel methods for intra-abdominal pressure monitoring , 2005, Intensive Care Medicine.

[13]  K. Hillman,et al.  Intra-abdominal pressure measurement using a modified nasogastric tube: Description and validation of a new technique , 1994, Intensive Care Medicine.

[14]  W. Butt,et al.  Comparison of indirect methods of measuring intra-abdominal pressure in children , 2005, Intensive Care Medicine.

[15]  M. Reinges,et al.  Evaluation of two novel methods for the direct and continuous measurement of the intra-abdominal pressure in a porcine model , 2003, Intensive Care Medicine.

[16]  C. Cocanour,et al.  Supranormal trauma resuscitation causes more cases of abdominal compartment syndrome. , 2003, Archives of surgery.

[17]  A. Viste,et al.  Comparison of different methods for measuring intra-abdominal pressure , 2002, Intensive Care Medicine.

[18]  E. Moore,et al.  The abdominal compartment syndrome is a morbid complication of postinjury damage control surgery. , 2001, American journal of surgery.

[19]  J. Tack,et al.  Perception of changes in wall tension of the proximal stomach in humans , 2001, Gut.

[20]  Jeffrey L. Johnson,et al.  Abdominal perfusion pressure: a superior parameter in the assessment of intra-abdominal hypertension. , 2000, The Journal of trauma.

[21]  R. Ivatury,et al.  Intra-abdominal hypertension after life-threatening penetrating abdominal trauma: prophylaxis, incidence, and clinical relevance to gastric mucosal pH and abdominal compartment syndrome. , 1998, The Journal of trauma.

[22]  B. Guslits,et al.  Increases in intra-abdominal pressure affect pulmonary compliance. , 1995, Archives of surgery.

[23]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[24]  Laurent Degos,et al.  HISTOCOMPATIBILITY DETERMINANTS IN MULTIPLE SCLEROSIS , 1974 .

[25]  R. Stephenson A and V , 1962, The British journal of ophthalmology.