Simple photon diffusion analysis of the effects of multiple scattering on pulse oximetry

Photon diffusion theory is used to derive analytical expressions that relate the AC-DC intensity ratios measured by transmission-mode and reflectance-mode pulse oximeters to arterial oxygen saturation (S/sub a/O/sub 2/). The effects of multiple scattering are examined by comparing the results of the photon diffusion analysis with those obtained using an analysis is based on the Beer-Lambert law which neglects scattering. It is shown that the difference between the average lengths of the paths traveled by red and infrared photons makes the calibration curve of oximeters sensitive to the total attenuation coefficients of the tissue in the two wavelength bands, as well as to absorption by the pulsating arterial blood. Therefore, the shape of the calibration curve is affected by tissue blood volume, source-detector placement, and other variables that change the wavelength dependence of the attenuation coefficient of the tissue. After evaluating the relationship between S/sub a/O/sub 2/ and the red/IR AC-DC ratio under a variety of physiological conditions, it is concluded that for oximeters utilizing fixed calibration curves based on measurements obtained from normal subjects, errors introduced by interfering variables should be less than a few percent when S/sub a/O/sub 2/ exceeds 70%.<<ETX>>

[1]  J. Severinghaus,et al.  Accuracy of Response of Six Pulse Oximeters to Profound Hypoxia , 1987, Anesthesiology.

[2]  A Ishimaru,et al.  Diffuse reflectance from a finite blood medium: applications to the modeling of fiber optic catheters. , 1976, Applied optics.

[3]  F. D. de Mul,et al.  Light propagation parameters for anisotropically scattering media based on a rigorous solution of the transport equation. , 1989, Applied optics.

[4]  C. Lenfant,et al.  Oxygen transport in man. , 1972, Chest.

[5]  Mathematical model of transmission pulse oximetry , 1988, Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[6]  J. C. van der Leun,et al.  FORWARD SCATTERING PROPERTIES OF HUMAN EPIDERMAL LAYERS , 1984, Photochemistry and photobiology.

[7]  J. Tulip,et al.  Optical propagation in tissue with anisotropic scattering , 1988, IEEE Transactions on Biomedical Engineering.

[8]  James D. Meindl,et al.  An Integrated Circuit-Based Optical Sensor for In Vivo Measurement of Blood Oxygenation , 1986, IEEE Transactions on Biomedical Engineering.

[9]  J M Schmitt,et al.  Multilayer model of photon diffusion in skin. , 1990, Journal of the Optical Society of America. A, Optics and image science.

[10]  A. P. Shepherd,et al.  Diffusion model of the optical absorbance of whole blood. , 1988, Journal of the Optical Society of America. A, Optics and image science.

[11]  Richard L. Longini,et al.  Diffusion dipole source , 1973 .

[12]  Ralph Nossal,et al.  Photon migration in layered media. , 1988, Applied optics.

[13]  A. Welch,et al.  A review of the optical properties of biological tissues , 1990 .

[14]  J A Pologe,et al.  Pulse Oximetry: Technical Aspects of Machine Design , 1987, International anesthesiology clinics.