In vitro evaluation of a novel autologous aortic valve (biovalve) with a pulsatile circulation circuit.

We have used in-body tissue architecture technology to develop an autologous valved conduit with intact sinuses of Valsalva (biovalve). In this study, we fabricated three different forms of biovalves and evaluated their function in vitro using a mock circulation model to determine the optimal biovalve form for aortic valve replacement. A cylindrical mold for biovalve organization was placed in a dorsal subcutaneous pouch of a goat, and the implant that was encapsulated with connective tissue was extracted 2 months later. The cylindrical mold was removed to obtain the biovalve (16 mm inside diameter) that consisted of pure connective tissue. The biovalve was connected to a pulsatile mock circulation system in the aortic valve position. The function of the three biovalves (biovalve A: normal leaflets with the sinuses of Valsalva; biovalve B: extended leaflets with the sinuses of Valsalva; biovalve C: extended leaflets without the sinuses of Valsalva) was examined under pulsatile flow conditions using saline. In addition, the mock circuit was operated continuously for 40 days to evaluate the durability of biovalve C. The regurgitation rate (expressed as a percent of the mean aortic flow rate during diastole) was 46% for biovalve A but only 3% for biovalves B and C. The durability test demonstrated that even after biovalve C pulsated more than four million times (heart rate, 70 bpm; mean flow rate, 5.0 L/min; mean aortic pressure, 92 mm Hg), stable continuous operation was possible without excessive reduction of the flow rate or bursting. The developed biovalve demonstrated good function and durability in this initial in vitro study.

[1]  V. Parravicini,et al.  Tuna cornea as biomaterial for cardiac applications. , 2012, Texas Heart Institute journal.

[2]  Tsutomu Tajikawa,et al.  Development of a Completely Autologous Valved Conduit With the Sinus of Valsalva Using In-Body Tissue Architecture Technology: A Pilot Study in Pulmonary Valve Replacement in a Beagle Model , 2010, Circulation.

[3]  T Dohi,et al.  Effect of the sinus of valsalva on the closing motion of bileaflet prosthetic heart valves. , 2000, Artificial organs.

[4]  Mitsuo Umezu,et al.  Engineering analysis of the effects of bulging sinuses in a newly designed pediatric pulmonary heart valve on hemodynamic function , 2012, Journal of Artificial Organs.

[5]  Y. Nakayama,et al.  In vivo evaluation of an in-body, tissue-engineered, completely autologous valved conduit (biovalve type VI) as an aortic valve in a goat model , 2013, Journal of Artificial Organs.

[6]  R. Tominaga,et al.  Artificial valves “up to date” in Japan , 2010, Journal of Artificial Organs.

[7]  Y. Nakayama,et al.  A completely autologous valved conduit prepared in the open form of trileaflets (type VI biovalve): mold design and valve function in vitro. , 2011, Journal of biomedical materials research. Part B, Applied biomaterials.

[8]  Alberto Redaelli,et al.  A Novel Approach to the In Vitro Hydrodynamic Study of the Aortic Valve: Mock Loop Development and Test , 2010, ASAIO journal.

[9]  Eisuke Tatsumi,et al.  Development and evaluation of endurance test system for ventricular assist devices , 2013, Journal of Artificial Organs.

[10]  K S Kunzelman,et al.  Aortic root and valve relationships. Impact on surgical repair. , 1994, The Journal of thoracic and cardiovascular surgery.

[11]  Tsutomu Tajikawa,et al.  Preparation of in-vivo tissue-engineered valved conduit with the sinus of Valsalva (type IV biovalve) , 2010, Journal of Artificial Organs.

[12]  Benyamin Rahmani,et al.  Manufacturing and hydrodynamic assessment of a novel aortic valve made of a new nanocomposite polymer. , 2012, Journal of biomechanics.

[13]  J. Ando,et al.  In vitro maturation of "biotube" vascular grafts induced by a 2-day pulsatile flow loading. , 2009, Journal of biomedical materials research. Part B, Applied biomaterials.

[14]  KAREN MENDELSON,et al.  Heart Valve Tissue Engineering: Concepts, Approaches, Progress, and Challenges , 2006, Annals of Biomedical Engineering.

[15]  Y. Nakayama,et al.  Preparation of a completely autologous trileaflet valve-shaped construct by in-body tissue architecture technology. , 2009, Journal of biomedical materials research. Part B, Applied biomaterials.