Cardiopulmonary Bypass Using Centrifugal Pump

Centrifugal pumps are nonocclusive pumps that offer centrifugal blood pumping known as the constrained force vortex principle. Of 2405 open heart procedures, 370 (15.4%) were performed with a centrif ugal pump (CP). In the rest of the cases extracorporeal circulation (ECC) was established with conventional roller pump (RP). The operations performed with a CP were: 34 reoperations, 11 ascending aortic aneurysms + coronary artery bypass graft (CABG), 47 multivalve replacements, 125 CABG + left ventricu lar aneurysmectomies, and 153 CABGs alone. This prospective study consisted of two groups, each group including 50 patients undergoing CABG surgery. In group I ECC was established with a CP and in group II with a conventional RR The two selected groups of patients were considered to be similar in terms of clinical, hemodynamic, and angio graphic data. Hemodynamic study concerned flow rate, systemic vascular re sistance (SVR), and urinary output. No difference was found between the CP and RP hemodynamically. Nevertheless, the alteration of the flow rate in re sponse to variation of SVR is accepted as an advantage. As for hemolytic effect during ECC of CP compared with RP: Plasma free hemoglobin, hemotocrit, platelet count, platelet factor 4, and fibrinogen levels, were investigated. These parameters revealed that there was significant hemoly sis and trauma to the blood components in the RP group when compared with the CR (p < 0.05, p < 0.01). There was no significant difference in psychoneurologic complications, but behavioral and sensory disorders were encountered sevenfold more in the RP group. This might be due to spoliation of the tubing set or to microair embo lism. Compared with the conventional RP the CP proved to have no hemodynamic advantage. The effects on blood components showed significant trauma to the RP group. Postoperative drainage verified this observation (p < 0.05). In this study the CP appears to be superior to the RP especially in compli cated and prolonged ECC.

[1]  J. Dimsdale,et al.  Postcardiotomy delirium: conclusions after 25 years? , 1989, The American journal of psychiatry.

[2]  H. Harasaki,et al.  Hemolysis. A comparative study of four nonpulsatile pumps. , 1988, ASAIO transactions.

[3]  J. Crowley,et al.  In Vitro Comparison of the Blood Handling by the Constrained Vortex and Twin Roller Blood Pumps , 1987, The Journal of ExtraCorporeal Technology.

[4]  D. Cohn,et al.  Analysis of microembolic particles originating in extracorporeal circuits , 1987 .

[5]  J. Copeland,et al.  Thromboembolic complications of the Jarvik-7 total artificial heart: case report. , 1986, Artificial organs.

[6]  R. Wilson,et al.  Hemostasis Changes During Cardiopulmonary Bypass Surgery , 1985, Seminars in thrombosis and hemostasis.

[7]  R. Bick Hemostasis Defects Associated with Cardiac Surgery, Prosthetic Devices, and Other Extracorporeal Circuits , 1985, Seminars in thrombosis and hemostasis.

[8]  T. Komoda,et al.  [Cardiopulmonary bypass using a centrifugal pump: hemodynamic and hematological evaluation in 33 clinical cases]. , 1984, [Zasshi] [Journal]. Nihon Kyobu Geka Gakkai.

[9]  K. Sakai,et al.  [A comparison of centrifugal and roller pumps. An experimental study]. , 1984, Kyobu geka. The Japanese journal of thoracic surgery.

[10]  A. Dobell,et al.  Catastrophic hemorrhage during redo sternotomy. , 1984, The Annals of thoracic surgery.

[11]  C. Jenkins,et al.  Neuropsychological dysfunction following elective cardiac operation. I. Early assessment. , 1983, The Journal of thoracic and cardiovascular surgery.

[12]  J. Orenstein,et al.  Microemboli observed in deaths following cardiopulmonary bypass surgery: silicone antifoam agents and polyvinyl chloride tubing as sources of emboli. , 1982, Human pathology.

[13]  C. Jenkins,et al.  Neuropsychological dysfunction following elective cardiac operation: I. Early assessment , 1982 .

[14]  P. Barash Cardiopulmonary bypass and postoperative neurologic dysfunction. , 1980, American heart journal.

[15]  M F Lynch,et al.  Centrifugal blood pumping for open heart surgery. , 1978, Minnesota medicine.

[16]  H. Tufo,et al.  Neurological abnormalities following open-heart surgery. , 1969, The Journal of thoracic and cardiovascular surgery.

[17]  W. R. Schmidt,et al.  Kinetic heart pumping. , 1984, Minnesota medicine.

[18]  D. Pennington,et al.  Clinical experience with a centrifugal pump ventricular assist device. , 1982, Transactions - American Society for Artificial Internal Organs.

[19]  H. Kletschka,et al.  Artificial heart. I. Application of nonpulsatile force-vortex principle. , 1968, Minnesota medicine.