High-performance MEMS-based gas chromatography column with integrated micro heater

A high-performance MEMS-based gas chromatography (GC) device is proposed comprising a miniature serpentine column with dimensions of 3.2 m × 200 μm × 250 μm (length × width × depth) and with an integrated Pt micro heater. The column is fabricated on a Si die measuring 3.5 × 1.8 mm2 using a wet etching process and is bonded to a Pyrex cover plate incorporating the Pt micro heater via a thermal fusion process. The experimental results reveal that an applied voltage of 9.7 V is sufficient to maintain a constant temperature of 85°C for elution purposes. In addition, it is shown that the proposed device successfully detects the concentrations of both pure and mixed samples of four volatile organic compound gases, namely acetone, toluene, methanol, and benzene. Finally, the theoretical plate number obtained by the proposed MEMS-based GC device is shown to be 2–3 times higher than that obtained from a conventional capillary-based GC system under the same injection conditions.

[1]  F. Zuiderweg,et al.  Longitudinal diffusion and resistance to mass transfer as causes of nonideality in chromatography , 1995 .

[2]  Gian Carlo Cardinali,et al.  Selectivity enhancement of metal oxide gas sensors using a micromachined gas chromatographic column , 2005 .

[3]  Edward B. Overton,et al.  Trends and advances in portable analytical instrumentation , 1996 .

[4]  K. Wise,et al.  High-performance temperature-programmed microfabricated gas chromatography columns , 2005, Journal of Microelectromechanical Systems.

[5]  M. Agah,et al.  Design, Modeling, and Fabrication of MEMS-Based Multicapillary Gas Chromatographic Columns , 2009, Journal of Microelectromechanical Systems.

[6]  Jean-Baptiste Sanchez,et al.  A selective gas detection micro-device for monitoring the volatile organic compounds pollution , 2006 .

[7]  B. Eckenrode The application of an integrated multifunctional field‐portable GC/MS system , 1998 .

[8]  S. Terry,et al.  A gas chromatographic air analyzer fabricated on a silicon wafer , 1979, IEEE Transactions on Electron Devices.

[9]  Jost Goettert,et al.  Fabrication of micro-gas chromatograph columns for fast chromatography , 2006 .

[10]  E. S. Kolesar,et al.  Silicon-micromachined gas chromatography system used to separate and detect ammonia and nitrogen dioxide. I. Design, fabrication, and integration of the gas chromatography system , 1994 .

[11]  E. Zellers,et al.  Multiple-stage microfabricated preconcentrator-focuser for micro gas chromatography system , 2005, Journal of Microelectromechanical Systems.

[12]  G. Spangler Relationships for modeling the performance of rectangular gas chromatographic columns , 2001 .

[13]  N. Sakauchi [Gas chromatography]. , 2020, Horumon to rinsho. Clinical endocrinology.