A positron tomograph with 600 BGO crystals and 2.6 mm resolution

A description is given of the imaging performance of the Donner 600-Crystal Positron Tomograph, a single 60-cm-diameter ring of 3-mm-wide bismuth germanate (BGO) crystals coupled individually to 14-mm phototubes. With a pulse height threshold of 200-keV and a slice thickness of 5 mm, the sensitivity is 7024 events/s per mu Ci/Ml in a 20-cm cylinder of water. The measured rates for 18 mu Ci/ml are 95000 trues/s plus 20000 randoms/s. A 0.3-mm-diameter /sup 22/Na line source near the center of the tomograph has a circular point-spread function (PSF) with a full-width at half-maximum (FWHM) of 2.6 mm. At 5 cm from the center the PSF is elliptical with a FWHM of 2.7 mm tangential*3.2 mm radial. At 10 cm the PSF has a FWHM of 2.8 mm tangential*4.8 mm radial. Attenuation data have been accumulated with a 20 mCi /sup 68/Ge orbiting transmission source, and 100 million coincident events have been collected in 200 s. >

[1]  M. Bergstrom,et al.  Design Studtes of Two Possible Detector Blocks for High Resolution Positron Emission Tomography of the Brain , 1987, IEEE Transactions on Nuclear Science.

[2]  R. Huesman,et al.  Scintillation Photon Detection and Event Selection in High Resolution Positron Emission Tomography , 1987, IEEE Transactions on Nuclear Science.

[3]  Stephen E. Derenzo,et al.  Mathematical Removal of Positron Range Blurring in High Resolution Tomography , 1986, IEEE Transactions on Nuclear Science.

[4]  W. Moses,et al.  Orbiting transmission source for positron tomography , 1988 .

[5]  D. Chesler,et al.  Design of a Cylindrical Shaped Scintillation Camera for Positron Tomographs , 1985, IEEE Transactions on Nuclear Science.

[6]  R. H. Huesman,et al.  Initial Results from the Donner 600 Crystal Positron Tomograph , 1987, IEEE Transactions on Nuclear Science.

[7]  S. E. Derenzo,et al.  The Electronics for the Donner 600-Crystal Positron Tomograph , 1986, IEEE Transactions on Nuclear Science.

[8]  Ronald H. Huesman,et al.  Imaging Properties of a Positron Tomograph with 280 Bgo Crystals , 1981, IEEE Transactions on Nuclear Science.

[9]  Michael V. Green,et al.  Post-injection transmission attenuation measurements for PET , 1988 .

[10]  R. Nutt,et al.  A Multicrystal Two Dimensional BGO Detector System for Positron Emission Tomography , 1986, IEEE Transactions on Nuclear Science.

[11]  S. K. Hilal,et al.  Detector Identification in a 4×4 BGO Crystal Array Coupled to Two Dual PMTS for High Resolution Positron Emission Tomography , 1987, IEEE Transactions on Nuclear Science.

[12]  H. Murayama,et al.  Development of a High Resolution Positron CT for Animal Studies , 1985, IEEE Transactions on Nuclear Science.

[13]  Stephen E. Derenzo,et al.  Initial Characterization of a Bgo-Photodiode Detector for High Resolution Positron Emission Tomography , 1984, IEEE Transactions on Nuclear Science.