Transabdominal fetal pulse oximetry: The case of fetal signal optimization

Current technology used for monitoring fetal well-being has been ineffective at reducing rates of harm to the fetus during the intrapartum period, yet its adoption has significantly increased the number of emergency C-sections performed. Transabdominal fetal pulse oximetry (TFO) aims to reduce the number of surgical interventions through non-invasive measurements of fetal oxygen saturation. When developing an optode for TFO, it is important to select design parameters that will maximize the measurement of the fetal signal. In this paper, we optimize the source-detector distance and wavelengths through Monte Carlo simulations using a multi-layered tissue model for various fetal depths. The results were validated by developing an optical probe with two wavelengths of light to observe pulsating arterial tissue through an optical phantom that mimics the maternal abdomen as a step towards oximetry. Our results show that 735nm and 850nm seem to be the optimal selection of peak wavelengths of light sources to obtain a stronger fetal signal for the fetal depths between 2–5 cm. Improving the signal sensitivity is approached by increasing the spacing between the source and detector, and is limited by the noise-equivalent power of the detector.

[1]  D B Swedlow,et al.  A multicenter controlled trial of fetal pulse oximetry in the intrapartum management of nonreassuring fetal heart rate patterns. , 2000, American journal of obstetrics and gynecology.

[2]  P.D. Mannheimer,et al.  Wavelength selection for low-saturation pulse oximetry , 1997, IEEE Transactions on Biomedical Engineering.

[3]  Peter Husar,et al.  Simulation of Photon Propagation in Multi-layered Tissue for Non-invasive Fetal Pulse Oximetry , 2014 .

[4]  W. Zijlstra,et al.  Absorption spectra of human fetal and adult oxyhemoglobin, de-oxyhemoglobin, carboxyhemoglobin, and methemoglobin. , 1991, Clinical chemistry.

[5]  H W Jongsma,et al.  Intrapartum fetal pulse oximetry: fetal oxygen saturation trends during labor and relation to delivery outcome. , 1994, American journal of obstetrics and gynecology.

[6]  P. Marquet,et al.  In vivo local determination of tissue optical properties: applications to human brain. , 1999, Applied optics.

[7]  Y L Narasinga Rao,et al.  Amniotic Fluid Optical Density (AFOD) correlates with the lung maturity as well as complete maturity of the fetus observational study , 2015 .

[8]  Erica Hyman Kates,et al.  Anemia and polycythemia in the newborn. , 2007, Pediatrics in review.

[9]  Z. Alfirevic,et al.  Continuous cardiotocography (CTG) as a form of electronic fetal monitoring (EFM) for fetal assessment during labour. , 2017, The Cochrane database of systematic reviews.

[10]  Shoko Nioka,et al.  Transabdominal fetal pulse oximetry with near-infrared spectroscopy. , 2005, American journal of obstetrics and gynecology.

[11]  S. Jacques Optical properties of biological tissues: a review , 2013, Physics in medicine and biology.

[12]  K. Belanger,et al.  Indications Contributing to the Increasing Cesarean Delivery Rate , 2011, Obstetrics and gynecology.

[13]  M. Kohl,et al.  Near-infrared optical properties of ex vivo human skin and subcutaneous tissues measured using the Monte Carlo inversion technique. , 1998, Physics in medicine and biology.

[14]  N Ramanujam,et al.  Photon migration through fetal head in utero using continuous wave, near-infrared spectroscopy: development and evaluation of experimental and numerical models. , 2000, Journal of biomedical optics.

[15]  A D Gordon,et al.  Near-infrared optical properties of ex vivo human uterus determined by the Monte Carlo inversion technique. , 1999, Physics in medicine and biology.

[16]  D. Boas,et al.  Trans-abdominal monitoring of fetal arterial blood oxygenation using pulse oximetry. , 2000, Journal of biomedical optics.

[17]  L Wang,et al.  MCML--Monte Carlo modeling of light transport in multi-layered tissues. , 1995, Computer methods and programs in biomedicine.

[18]  園子 伊地知,et al.  Developmental changes of optical properties in neonates determined by near-infrared time-resolved spectroscopy , 2005 .

[19]  I Shapiro,et al.  Myometrial thickness in pregnancy: longitudinal sonographic study. , 1998, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[20]  S. Fieni,et al.  Amniotic fluid dynamics. , 2004, Acta bio-medica : Atenei Parmensis.