A miniature force sensor for catheter based on optical micro deformation detection

This paper presents a preliminary design of a miniature force sensor for the cardiac catheter to detect the contact force between the catheter-tip and myocardial tissue. A spring-like elastic part was specially designed to transform the distal force to the micro deformation. An optical fiber was assembled behind the elastic part to detect the deformation by analyzing the reflective light intensity. The concise structure of the force-deformation transferring part and the usage of standard material greatly promise a low-cost force sensor. The experimental results indicate the sensor works with fairly good linearity. This sensor will assist the cardiac surgeon with realtime and precise force feedback.

[1]  P. Polygerinos,et al.  Triaxial Catheter-Tip Force Sensor for MRI-Guided Cardiac Procedures , 2013, IEEE/ASME Transactions on Mechatronics.

[2]  Nicholas S. Peters,et al.  Experience of robotic catheter ablation in humans using a novel remotely steerable catheter sheath , 2008, Journal of Interventional Cardiac Electrophysiology.

[3]  Claus Schmitt,et al.  Freedom From Atrial Tachyarrhythmias After Catheter Ablation of Atrial Fibrillation: A Randomized Comparison Between 2 Current Ablation Strategies , 2005, Circulation.

[4]  L. Tenerz,et al.  A production process of silicon sensor elements for a fibre optic pressure sensor , 1997 .

[5]  Feng WeiXing,et al.  Design and Experiments of a Catheter Side Wall Tactile Sensor for Minimum Invasive Surgery , 2007, 2007 International Conference on Mechatronics and Automation.

[6]  H. Lambert,et al.  Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and Thrombus , 2008, Circulation. Arrhythmia and electrophysiology.

[7]  Kaspar Althoefer,et al.  MRI-Compatible Intensity-Modulated Force Sensor for Cardiac Catheterization Procedures , 2011, IEEE Transactions on Biomedical Engineering.

[8]  Patrick J. French,et al.  Multi-parameter sensor system with intravascular navigation for catheter/guide wire application , 2002 .

[9]  Frank Bogun,et al.  Circumferential pulmonary-vein ablation for chronic atrial fibrillation. , 2006, The New England journal of medicine.

[10]  Jan Peirs,et al.  A micro optical force sensor for force feedback during minimally invasive robotic surgery , 2003 .

[11]  Henry V. Allen,et al.  A Novel Ultra-miniature catheter tip pressure sensor fabricated using silicon and glass thinning techniques , 2001 .

[12]  Denis Donlagic,et al.  Miniature fiber optic pressure sensor for medical applications , 2002, Proceedings of IEEE Sensors.

[13]  Kaspar Althoefer,et al.  MRI-Compatible Fiber-Optic Force Sensors for Catheterization Procedures , 2010, IEEE Sensors Journal.

[14]  Kaspar Althoefer,et al.  Novel miniature MRI-compatible fiber-optic force sensor for cardiac catheterization procedures , 2010, 2010 IEEE International Conference on Robotics and Automation.