Shielding requirements for constant-potential diagnostic x-ray beams determined by a Monte Carlo calculation.

A Monte Carlo calculation has been performed to determine the transmission of broad constant-potential x-ray beams through Pb, concrete, gypsum wallboard, steel and plate glass. The EGS4 code system was used with a simple broad-beam geometric model to generate exposure transmission curves for published 70, 100, 120 and 140-kVcp x-ray spectra. These curves are compared to measured three-phase generated x-ray transmission data in the literature and found to be reasonable. For calculation ease the data are fit to an equation previously shown to describe such curves quite well. These calculated transmission data are then used to create three-phase shielding tables for Pb and concrete, as well as other materials not available in Report No. 49 of the NCRP.

[1]  D. Rogers More realistic Monte Carlo calculations of photon detector response functions , 1982 .

[2]  D. Rogers Fluence to dose equivalent conversion factors calculated with EGS3 for electrons from 100 keV to 20 GeV and photons from 11 keV to 20 GeV. , 1984, Health physics.

[3]  D. Simpkin Radiation shielding of multiple x-ray sources in diagnostic radiology. , 1986, Health physics.

[4]  J. H. Hubbell,et al.  Photon cross sections, attenuation coefficients, and energy absorption coefficients from 10 keV to 100 GeV , 1969 .

[5]  F. Spiers Physics of Radiology , 1968, Nature.

[6]  F. H. Attix Introduction to Radiological Physics and Radiation Dosimetry , 1991 .

[7]  D. Simpkin Shielding requirements for mammography. , 1987, Health physics.

[8]  David W. O. Rogers,et al.  Low energy electron transport with EGS , 1984 .

[9]  D. Rogers,et al.  Calculated buildup curves for photons with energies up to 60Co. , 1985, Medical physics.

[10]  S C Bushong,et al.  Diagnostic X-ray shielding design based on an empirical model of photon attenuation. , 1983, Health physics.

[11]  Fitting parameters for medical diagnostic x-ray transmission data. , 1988, Health physics.

[12]  D. Rogers,et al.  EGS4 code system , 1985 .

[13]  C. Goodman,et al.  Food and Drug Administration Center for Devices and Radiological Health , 1988 .

[14]  J. Légaré,et al.  Blindage contre les grands champs de rayons X primaires et diffusés des appareils triphasés au moyen de panneaux de verre, de gypse et de plomb acoustique , 1978 .